Kristina Andelovic1,2, Patrick Winter2, Thomas Kampf2,3, Julius Heidenreich3, Anton Xu2, Peter M. Jakob2, Wolfgang R. Bauer1, and Volker Herold2
1Medicine I, Cardiology, University Hospital Wuerzburg, Wuerzburg, Germany, 2Experimental Physics V, University of Wuerzburg, Wuerzburg, Germany, 3Interventional and Diagnostic Radiology, University Hospital Wuerzburg, Wuerzburg, Germany
Synopsis
4D phase
contrast (PC)-MRI is a non-invasive tool for the assessment of cardiovascular
hemodynamics or the Wall Shear Stress (WSS) to study atherosclerotic risks in
vivo. Major limitations of conventional triggered methods are the long
measurement times needed for high-resolution data sets and the requirement of
stable ECG triggering, which is diffcult at high magnetic field strengths. In
this work, an ECG-free, retrospectively synchronized method is presented that
enables fast high-resolution measurements of 4D flow and wall shear stress in
the murine aortic arch.
Purpose
4D phase contrast (PC)-MRI became an important tool for preclinical studies
of atherosclerosis, which enables non-invasive measurements of cardiovascular
hemodynamics as well as the assessment of functional parameters such as the
wall shear stress (WSS). Major limitations of conventional triggered methods
are the long measurement times needed for high-resolution data sets and the
necessity of stable ECG triggering. In
this abstract, we
present an advanced
method based on
a self-navigated 3D
radial PC-cine FLASH acquisition,
that enables high-resolution (100 μm)3 4D flow measurements
of the complete aortic arch in approximately 30 minutes. 4D-flow
dynamics, aortic cross
sectional areas, longitudinal,
circumferential and radial
wall shear stress
components as well as the oscillatory shear
index (OSI) can be determined from
the aquired data. The use of
retrospective reconstruction and self-gating enables a very flexible data
analysis and increases the robustness since prospective ECG triggering is no
longer needed.Methods
MR-Measurements
All data were
acquired on a vertical 17.6T small animal MRI scanner with a 1 T/m gradient
system and a 24 mm birdcage coil. 4D flow and WSS was measured in the aortic
arch of 12-weeks-old wild-type C57/BL6J mice (n=7) with a radial 4D PC-MRI
sequence. For slice excitation, a sinc-shaped pulse with a flip angle of 15°
was used. The acquisition parameters were: VENC=125 cm/s,
TR/TE=3.0/1.1 ms, echo asymmetry: 10%, 160000 spokes, FOV: 25x25x4 mm3,
spatial resolution: 100 µm (isotropic). 4D flow measurements were conducted
without triggering during free breathing and required a total scan time of 32
minutes.
Reconstruction and Data Acquisition
Cardiac and
respiratory motion signals extracted from the radial MR signal were used for
the retrospective reconstruction1 of 4D-flow cines. Rigid motion
correction and a first order B0 correction was used to improve the
robustness of magnitude and velocity data. For each velocity encoding step, 3D
images with an isotropic spatial resolution of 100μm were
reconstructed with convolution gridding at 30 different cardiac phases. After
segmenting the aortic lumen semi-automatically2, time-resolved WSS-components and OSI3 were calculated from the spatial velocity gradients at the lumen
surface at 14 locations along the aorta (Fig.1C+D). All reconstructions were
done with Matlab (The Mathworks, Inc., Natick, USA), WSS and OSI were
calculated and visualized with Ensight (CEI Software, USA)3.
Results
Self-navigation signals were successfully extracted from the radial DC signal. The signal modulations due to cardiac and respiratory
motions are clearly recognizable and were used to determine cardiac phases and
breath-gating windows (Fig.1A). As shown in Fig.1B, no significant variations
between the 4 flow-encoding measurements were observed. Volume flow, cross-sectional areas (Fig.3C),
WSS (Fig.3+4) and OSI (Fig.3C) were determined in a measurement time of only 32
minutes. Mean and median of all 3 WSS components (Fig.3A+B) as well as time-resolved WSS over the whole heart cycle (Fig.4) was assessed at 14
analysis planes along the aortic arch. Interestingly, an opposing behavior was
observed between maximum differences of cross sectional areas and temporally averaged radial WSS
(Fig.3C). Average longitudinal,
circumferential and radial WSS values were 1.52 ± 0.29 N/m2 , 0.28 ±
0.24 N/m2 and −0.21 ± 0.19 N/m2, which corresponds well
with literature3,4,5.Discussion
In this abstract we demonstrate the feasibility of fast self-navigated 4D flow and WSS measurements
in the murine aortic arch with very high temporal (30 frames / cardiac cycle)
and spatial resolutions (100 μm)3. The proposed method does not
require ECG signals for motion synchronization and hence leads to higher
robustness and an improved animal handling. The retrospective approach allows a
very flexible data analysis that can be adapted to either high temporal or high
spatial resolution. Possible applications of this new method are studies of
WSS and hemodynamics in atherosclerotic mouse models as well as 3D measurements
of the aortic pulse-wave-velocity.Acknowledgements
This work was supported by
grants from the Deutsche Forschungsgemeinschaft (SFB 688 B5, Z2) and the
Bundesministerium für Bildung und Forschung (BMBF01 EO1004).References
1. Winter et al., JCMR [2013]; 15:88-98.
2. Herold et al., JCMR
[2017]; 19(1):77
3. Stalder et al., MRM
[2008]; 60(5): 1218-1231
4 Frydrychowitz et
al., JMRI [2009]; 30(1), 77-84
5. Janiczek et al.,
MRM[2011]; 66(5), 1382-1390