Yu Chun-Yang1, Huang Teng-Yi2, and Chung Hsiao-Wen1
1National Taiwan University, Taipei, Taiwan, 2National Taiwan University of Science and Technology, Taipei, Taiwan
Synopsis
A hybrid MOLLI method that integrated saturation recovery with the classical
inversion recovery sequence was proposed for quantitative T1 mapping in the
myocardium within one single breath-hold. By replacing the second inversion
pulse of the original MOLLI53 technique with a saturation pulse, the long
recovery time could be alleviated in hybrid MOLLI, thereby allowing more images
to be sampled from the T1 relaxation curve. Phantom and healthy subject experiments
conducted in comparison with the classical MOLLI53 demonstrated that the
proposed method was able to provide comparable image quality as well as precise
T1 quantification in the myocardium.
Introduction
Alternations of myocardial T1 values have been shown to be related to
disease states. Therefore, characterization of the native T1 values may provide
useful information to detect and assess cardiomyopathies. In terms of T1
measurements, inversion recovery methods such as MOLLI are widely used
clinically but are subject to magnetization transfer effects (1), whereas
saturation recovery methods including SASHA are potentially more accurate but
are prone to noise influences (2). The purpose of this study is therefore to
propose a new hybrid method combining inversion and saturation recovery and to investigate
the precision of T1 measurements on both phantoms and healthy subjects.Materials and Methods
The
hybrid MOLLI method proposed in this study is composed of an inversion pulse after
which five images are acquired, followed by a saturation pulse after which six
images are acquired (Fig.1). Compared with the original MOLLI5(3s)3 design (two
inversion pulses separated by 3 seconds, with five and three images acquired
after each inversion pulse, respectively), there is no need to wait for full
recovery of the magnetization, hence allowing more images to be sampled for the
T1 relaxation curve (3).
A phantom consisting
of eight bottles of MnCl2 solution was imaged at 3T using the
original MOLLI5(3s)3 and the hybrid MOLLI, both repeated twice to assess
reproducibility. The T1 values estimated were compared against inversion
recovery turbo spin-echo as the gold standard. Simulated ECG was varied from 50
bpm to 90 bpm to assess the effect of heart rate variability. In addition,
twelve healthy subjects were imaged using both MOLLI5(3s)3 and hybrid MOLLI in
a breath-hold. The measurements were again repeated twice to assess
reproducibility. T1 was estimated on a pixel-by-pixel basis using nonlinear
least square error fitting in MATLAB. Myocardial T1 values measured from
selected regions-of-interest were compared.Results
Nonlinear fitting of the data to the signal model was successful
(Fig.1). Both MOLLI53 and hybrid MOLLI showed T1 estimation with good
precision, although underestimation was found when compared with inversion
recovery turbo spin-echo measurements in the phantom (Fig.2). Figure 3 shows
one representative slice of T1 maps acquired using MOLLI53 and hybrid MOLLI,
respectively. It is seen that both techniques were able to accomplish T1
mapping in a single breath-hold, and yielded quantitative values very similar
to each other. Regions-of-interest measurements from the myocardium also showed
comparable precision at the level of about 3% errors for MOLLI53 and hybrid
MOLLI (Fig.4).Discussion
The results from our healthy subject study suggest that the hybrid MOLLI
technique combining saturation recovery and inversion recovery is an effective
alternative to the widely used MOLLI5(3s)3 sequence for quantitative T1 mapping
in the myocardium. The use of saturation pulse instead of a second inversion
pulse alleviated the need of long recovery periods as in the conventional
MOLLI53 sequence (3), allowing more images to be acquired than conventional
MOLLI5(3s)3. In the case of patients with arrhythmia, it is expected that the
hybrid MOLLI method may provide advantages by sufficiently sampling the T1 relaxation
curve.Conclusion
The hybrid MOLLI technique proposed in this study has been shown to
yield precise T1 mapping in a single breath-hold, with quality comparable to
that shown in the conventional MOLLI53 method on phantoms and healthy subjects.Acknowledgements
No acknowledgement found.References
- Kellman P, Hansen MS. T1-mapping in the heart: accuracy and precision. J
Cardiovasc Magn Reson 2014;16:2.
- Kim PK, Hong YJ, Im DJ, et al. Myocardial T1 and T2 Mapping: Techniques
and Clinical Applications. Korean J Radiol 2017;18:113-131.
- Roujol S, Weingärtner S, Foppa M, et al. Accuracy, Precision, and
Reproducibility of Four T1 Mapping Sequences: A Head-to-Head Comparison of
MOLLI, ShMOLLI, SASHA, and SAPPHIRE. Radiology 2014;272:683-689.