2891

Matching Error Evaluation in Magnetic Resonance Fingerprinting with a Fast Imaging with Steady Precession sequence using Bloch Equation Simulations with a Diffusion Propagator
Shota Hodono1, Yun Jiang2, Gregor Körzdörfer3,4, Naren Nallapareddy5, Vikas Gulani2, and Mark Griswold2

1Physics and Astronomy, Ohio Northern University, Ada, OH, United States, 2Radiology, Case Western Reserve University, Cleveland, OH, United States, 3Siemens Healthcare, Erlangen, Germany, 4Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany, 5Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States

Synopsis

The robustness of T1, T2 values derived from Magnetic Resonance Fingerprinting (MRF) is limited in certain situations because MRF dictionaries have in general not included apparent diffusion coefficients (ADC). In this study, the potential estimated T1, T2 errors due to the omission of diffusion were evaluated for the MRF-fast imaging with steady precession sequence. Dictionaries with ADC values were generated by using Bloch equations with a diffusion propagator. The generated signal evolutions with ADC were matched to those generated by Bloch equation simulations without ADC by employing a template-matching algorithm.

Introduction

Magnetic Resonance Fingerprinting1 (MRF) can provide fast and quantitative measurements of several clinically-relevant tissue properties using a pulse sequence with pseudo-random acquisition parameters. However, T1, T2 values estimated from MRF rely on an accurate model of the signal dynamics, in order to derive tissue properties by matching the acquired signal with a “dictionary” of pre-calculated signal evolutions. The reported T1, T2 values derived from MRF are not yet perfectly robust in some situations. A likely cause for deviation of the estimated relaxation values from their true values is that MRF dictionaries have in general not included apparent diffusion coefficients (ADC), primarily due to the difficulties associated with simulating the MRF signal evolutions. Hence, the purpose of this work was to analyze the errors induced by omitting diffusion in estimates of relaxation parameters derived from the MRF- Fast Imaging with Steady Precession (FISP) sequence2, by including a diffusion propagator in Bloch equation simulations.

Methods

Bloch simulations with a diffusion propagator were applied to the MRF-FISP sequence (Fig.1). In this study, three sizes of the gradient-induced dephasing (8, 12, 16$$$\pi$$$) across the slice thickness (5mm) were employed during a 1ms magnetic field gradient in each repetition time (TR). TR and flip angles (FA) were varied throughout the acquisition as shown in Fig.1. Firstly, “no-ADC-dictionary” MRF signal evolutions were generated over a range of T1 (100-3000ms, in 1ms increments), T2 (20-300ms, in 1ms increments) using the conventional Bloch simulations. For the purpose of analyzing the matching errors between signal evolutions with and without diffusion, MRF signal evolutions corresponding to three “ADC-dictionaries”, including two ADC values (1.0$$$\times$$$10-3 and 3.0$$$\times$$$10-3mm2/s) were generated for the three dephasing sizes. For each of the ADC-dictionaries, the following T1 and T2 values were employed: fixed T1=800ms with a range of T2=50-300ms, and T1=500-3000ms with fixed T2=80ms. In addition, to evaluate the matching errors in tissue properties, signal evolutions for grey matter (GM) and white matter (WM) were also generated: T1/T2=1200/90ms, ADC=0.64$$$\times$$$10-3mm2/s, and T1/T2=700/70ms, ADC=0.83$$$\times$$$10-3mm2/s for GM and WM, respectively3. Off-resonance frequencies were set to be equal to zero in all cases. Each ADC-dictionary entry was matched with an entry of the no-ADC-dictionary by employing a template-matching algorithm1. T1 and T2 matching percentage errors3 between the no-ADC-dictionary and the ADC-dictionary entries were calculated for each of the three different dephasing sizes: $$$Error=\frac{T_{1,2}(no-ADC-dictionary)-T_{1,2}(ADC-dictionary)}{T_{1,2}(ADC-dictionary)}\times100$$$[%].

Results

Figs.2 and 3 show matching errors when T2 was varied (T1 fixed) and T1 was varied (T2 fixed), respectively. These results imply that for all conditions, both of the relaxation parameters were underestimated in the matching process for the MRF-FISP sequence if diffusion was not accounted for. Fig.2 shows that the size of the matching error (degree of underestimation) of T2 increased with larger T2 values and dephasing sizes. Conversely, Fig. 3 shows that the matching error of T1 values was relatively smaller than T2 matching errors, with some evidence of a trend to higher matching error for lower T1. Over all simulations, the T1 matching errors were always lower in magnitude than -1%, even in the case of 16$$$\pi$$$ dephasing with ADC=3.0$$$\times$$$10-3mm2/s, whilst T2 matching errors reached ~-10% for 16$$$\pi$$$ and long T2. Fig. 4 shows the T2 matching errors for GM and WM. As the dephasing size increased, the degree of T2 underestimation increased in magnitude, reaching an error of -2.2% for GM in the case of 16$$$\pi$$$ dephasing.

Discussion

Here we presented an error analysis of T1 and T2 estimation of MRF-FISP sequence with different dephasing gradient moment in each repetition time. While Extended Phase Graphs (EPG)4 can also be employed to generate MRF dictionaries, diffusion propagator, while taking longer time in calculation, could be more flexible to take account other parameters, such as off-resonance for extending the current work. This study indicates that the previously reported MRF T1, T2 values are likely underestimated and should be corrected. However, the error introduced by the diffusion sensitivity is only 2.2% in normal gray matter with 16$$$\pi$$$.

Conclusion

Potential T1, T2 matching errors due to the omission of diffusion in the signal evolution during the MRF-FISP sequence have been evaluated by implementing a diffusion propagator in Bloch simulations. In the near future, inclusion of ADC values in MRF dictionaries should permit the use of MRF pulse sequences with magnetic field gradients of large magnitude.

Acknowledgements

This work was supported by Siemens Healthcare and NIH grants 1R01EB016728, 1R01DK098503 and 1R01BB017219.

References

1. Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature. 2013;495(7440):187-192.

2. Jiang Y, et al. MR Fingerprinting Using Fast Imaging with Steady State Precession (FISP) with Spiral Readout, Magn. Reson. Med. 2015;74(6):1621-1631.

3. Terada Y, et al. Effect of diffusion weighting due to spoiler gradients in MR Fingerprinting. Proc. Intl. Soc. Mag. Reson. Med. 2017:25:0940.

4. Weigel. M. Extended phase graphs: Dephasing, RF pulses, and echoes - pure and simple. J Magn Reson Imaging. 2015;41(2):266-295

Figures

Figure 1. (a) Diagram of the MRF-FISP sequence. TE was 2ms which was fixed for all frames. The gradient magnitude was varied corresponding to the dephasing sizes. (b) variation in TR (11.7-14.3ms) used in the sequence. (c) variation in FA (0-67 degree) used in the sequence.

Figure 2. T2 matching percentage errors when T2 was varied from 500 to 3000ms, with fixed T1 = 800ms.

Figure 3. T1 matching percentage errors when T1 was varied from 500 to 3000ms, with fixed T2 = 80ms.

Figure 4. T2 matching errors in grey matter and white matter. In this study, T2 matching errors were not observed for white matter when the dephasing sizes were 8$$$\pi$$$ and 12$$$\pi$$$, likely due to insufficient resolution of the no-ADC-dictionary for detection of matching errors.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
2891