Shota Hodono1, Yun Jiang2, Gregor Körzdörfer3,4, Naren Nallapareddy5, Vikas Gulani2, and Mark Griswold2
1Physics and Astronomy, Ohio Northern University, Ada, OH, United States, 2Radiology, Case Western Reserve University, Cleveland, OH, United States, 3Siemens Healthcare, Erlangen, Germany, 4Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany, 5Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States
Synopsis
The robustness of T1, T2 values derived from Magnetic Resonance Fingerprinting
(MRF) is limited in certain situations because MRF dictionaries have in general
not included apparent diffusion coefficients (ADC). In this study, the
potential estimated T1, T2 errors due to the omission of diffusion were
evaluated for the MRF-fast imaging with steady precession sequence. Dictionaries
with ADC values were generated by using Bloch equations with a diffusion
propagator. The generated signal evolutions with ADC were matched to those generated
by Bloch equation simulations without ADC by employing a template-matching
algorithm.
Introduction
Magnetic Resonance
Fingerprinting1 (MRF) can provide fast and quantitative measurements
of several clinically-relevant tissue properties using a pulse sequence with pseudo-random
acquisition parameters. However, T1, T2 values estimated from MRF rely on an accurate model
of the signal dynamics, in order to derive tissue properties by matching the
acquired signal with a “dictionary” of pre-calculated signal evolutions. The
reported T1, T2 values derived from MRF
are not yet perfectly robust in some situations. A likely cause for deviation
of the estimated relaxation values from their true values is that MRF
dictionaries have in general not included apparent diffusion coefficients
(ADC), primarily due to the difficulties associated with simulating the MRF
signal evolutions. Hence, the purpose of this work was to analyze the errors
induced by omitting diffusion in estimates of relaxation parameters derived
from the MRF- Fast Imaging with Steady Precession (FISP) sequence2, by
including a diffusion propagator in Bloch equation simulations. Methods
Bloch simulations
with a diffusion propagator were applied to the MRF-FISP sequence (Fig.1). In this
study, three sizes of the gradient-induced dephasing (8, 12, 16$$$\pi$$$) across the slice thickness (5mm) were
employed during a 1ms magnetic field gradient in each repetition time (TR). TR
and flip angles (FA) were varied throughout the acquisition as shown in Fig.1. Firstly,
“no-ADC-dictionary” MRF signal evolutions were generated over a range of T1 (100-3000ms,
in 1ms increments), T2 (20-300ms, in 1ms increments) using the conventional
Bloch simulations. For the purpose of analyzing the matching errors between signal
evolutions with and without diffusion, MRF signal evolutions corresponding to three
“ADC-dictionaries”, including two ADC values (1.0$$$\times$$$10-3 and 3.0$$$\times$$$10-3mm2/s) were generated for the three dephasing sizes.
For each of the ADC-dictionaries, the following T1 and T2 values
were employed: fixed T1=800ms with a range of T2=50-300ms, and T1=500-3000ms with
fixed T2=80ms. In addition, to evaluate the matching errors in tissue properties,
signal evolutions for grey matter (GM) and white matter (WM) were also
generated: T1/T2=1200/90ms, ADC=0.64$$$\times$$$10-3mm2/s, and T1/T2=700/70ms, ADC=0.83$$$\times$$$10-3mm2/s for GM
and WM, respectively3. Off-resonance frequencies were set to be equal
to zero in all cases. Each ADC-dictionary entry was matched with an entry of the
no-ADC-dictionary by employing a template-matching algorithm1. T1 and
T2 matching percentage errors3 between the no-ADC-dictionary and the
ADC-dictionary entries were calculated for each of the three different
dephasing sizes: $$$Error=\frac{T_{1,2}(no-ADC-dictionary)-T_{1,2}(ADC-dictionary)}{T_{1,2}(ADC-dictionary)}\times100$$$[%].Results
Figs.2 and 3 show
matching errors when T2 was varied (T1 fixed) and T1 was varied (T2 fixed),
respectively. These results imply that for all conditions, both of the
relaxation parameters were underestimated in the matching process for the MRF-FISP
sequence if diffusion was not accounted for. Fig.2 shows that the size of the
matching error (degree of underestimation) of T2 increased with larger T2 values
and dephasing sizes. Conversely, Fig. 3 shows that the matching error of T1 values
was relatively smaller than T2 matching errors, with some evidence of a trend
to higher matching error for lower T1. Over all simulations, the T1 matching
errors were always lower in magnitude than -1%, even in the case of 16$$$\pi$$$ dephasing with ADC=3.0$$$\times$$$10-3mm2/s,
whilst T2 matching errors reached ~-10% for 16$$$\pi$$$ and long T2. Fig. 4 shows the
T2 matching errors for GM and WM. As the dephasing size increased, the degree
of T2 underestimation increased in magnitude, reaching an error of -2.2% for GM
in the case of 16$$$\pi$$$ dephasing. Discussion
Here we presented
an error analysis of T1 and T2 estimation of MRF-FISP sequence with different
dephasing gradient moment in each repetition time. While Extended Phase Graphs
(EPG)4 can also be employed to generate MRF dictionaries, diffusion
propagator, while taking longer time in calculation, could be more flexible to
take account other parameters, such as off-resonance for extending the current
work.
This study indicates
that the previously reported MRF T1, T2 values are likely underestimated and
should be corrected. However, the error introduced by the diffusion sensitivity
is only 2.2% in normal gray matter with 16$$$\pi$$$.Conclusion
Potential T1, T2 matching errors
due to the omission of diffusion in the signal evolution during the MRF-FISP
sequence have been evaluated by implementing a diffusion propagator in Bloch
simulations. In the near future, inclusion of ADC values in MRF dictionaries should
permit the use of MRF pulse sequences with magnetic field gradients of large
magnitude. Acknowledgements
This work was supported by Siemens Healthcare and NIH grants
1R01EB016728, 1R01DK098503 and 1R01BB017219.References
1. Ma D, Gulani V,
Seiberlich N, et al. Magnetic resonance fingerprinting. Nature.
2013;495(7440):187-192.
2. Jiang Y, et al.
MR Fingerprinting Using Fast Imaging with Steady State Precession (FISP) with
Spiral Readout, Magn. Reson. Med. 2015;74(6):1621-1631.
3. Terada Y, et
al. Effect of diffusion weighting due to spoiler gradients in MR Fingerprinting.
Proc. Intl. Soc. Mag. Reson. Med. 2017:25:0940.
4. Weigel. M. Extended
phase graphs: Dephasing, RF pulses, and echoes - pure and simple. J Magn Reson
Imaging. 2015;41(2):266-295