Wanyong Shin1 and Mark J Lowe1
1Radilogy, Cleveland Clinic, Cleveland, OH, United States
Synopsis
Recently, a physiological log file timing error
with fMRI acquisition was fixed and physiologic data with corrected timing was uploaded
to the WU-MINN human connectome project
(HCP) cloud. While HCP preprocessing pipeline for resting state (rs-) and fMRI has
been proposed, the physiologic noise correction sub-pipeline has not been established
yet and the physiologic noise data has had less attention in HCP community. In
this study, we investigate the quality of HCP physiologic data and propose a
standard physiological noise quality assurance.
Introduction
The purpose of this study
is to investigate which properties of physiologic data could predict the
performance of the physiological noise correction in HCP rs-fMRI data, and
which level of factors could be used as an outlier threshold for bad
physiologic data.
Method
Two different phase
encoding direction (RL or LR) rsfMRI data and corresponding physiological data
were used from randomly selected 143 subjects in WU-MINN data set. Total 286 scans
data with physiologic files were used.
We monitored A) the number
of nulled and B) saturated signal measures from the physiologic data. Then the
physiologic data was processed using modified RetroTS.m (https://afni.nimh.nih.gov)
to define each physiologic cycle by the detection of adjacent peaks. The
periodicity time was calculated and C) its standard deviation divided by its
mean (COV) was calculated, which represents the consistency of the physiologic
periodicity. The repeated, normalized signals were resampled with 2π/50 window over
the phase cycles, and D) the standard deviation (SD) of signals in each phase window
was calculated and averaged, which represents the consistency of signals over
phases.
We calculated vowelwise F
values of the second order Fourier series model (M=2, 4 each regressors) for
cardiac and respiratory signals using RETROICOR [1]. The mean of the top
highest 1% of F values was used for the output of the physiological noise
correction performance. Result
Table1 shows the list of
variables and the factor analysis of them. As shown, the number of nulled
signals are highly correlated to the number of saturated signals with 0.74 and
0.68 in cardiac and respiratory signals, respectively. Also, COV of period is
highly correlated to SD of the repeated signals over the phase with 0.85 and
0.71 in cardiac and respiratory signals, respectively.
Figure 1 and 2 show the
correlation coefficient between mean value of top 1% F value from cardiac and
respiratory RETROICOR results and each predictor. It is observed that COV of
the period and SD of normalized signals are the significant predictors of the goodness
of physio correction. Figure 3 plots the relationship between COV of the period
and SD of the repeated signals. Note that low numbers of COV and SD indicate
the consistent patterns of physiologic data. We set the outlier threshold as
0.6 and 0.2 for the cardiac signals, and 0.6 and 0.3 for the respiratory
signals.
Figure 4 shows the example
of quality assurance result of good and bad cases. Discusion
The number of nulled and
the saturated physiologic signals is not highly correlated to the quality of
physiologic correction. However, it could represent the quality of the
physiologic data measures. We excluded physiologic data with more than 500 and
3000 points of null and saturated signals for the cardiac and respiratory,
which results in the exclusion of 6% of the physiologic data.
The proposed outlier
threshold using COV of the period and SD of the repeated signals leave 197
cardiac and 186 respiratory data among 282, which is 70% and 67 % of cardiac
and respiratory data. A more comprehensive investigation is needed to determine
the threshold more accurately.
It was found that the physiological correction could
be conducted more efficiently using alternative physiologic estimator methods, such
as PESTICA [2], in case that physiologic
data is corrupted [3]. The proposed QA standard could be considered as a
matrix to determine which physiologic noise data should be employed.Conclusion
We investigate the quality of HCP physiologic
data and propose the quality assurance standard of them. The script to generate
physiologic QA, shown in Fig4 will be provided in public in NITRC site soon (www.nitrc.org/projects/pestica)Acknowledgements
This work was supported by
Cleveland Clinic. Authors appreciate HCP community for the helpful discussion.References
1. Glover GH, Li TQ, Ress D.
Image-based method for retrospective correction of physiological motion effects
in fMRI: RETROICOR. Magn Reson Med. 2000;44(1):162-7. Epub 2000/07/14. doi:
10.1002/1522-2594(200007)44:1<162::AID-MRM23>3.0.CO;2-E [pii]. PubMed
PMID: 10893535.
2. Beall EB,
Lowe MJ. Isolating physiologic noise sources with independently determined
spatial measures. Neuroimage. 2007;37(4):1286-300. Epub 2007/08/11. doi:
S1053-8119(07)00585-X [pii]