Evan McNabb1, Raimond Wong2, and Michael D Noseworthy1,3
1School of Biomedical Engineering, McMaster University, Hamilton, ON, Canada, 2Juravinski Cancer Centre, Hamilton, ON, Canada, 3Electrical and Computer Engineering, McMaster University, Hamilton, ON, Canada
Synopsis
The dual-plane co-RASOR sequence is able to differentiate between a
LDR brachytherapy seeds and gold fiducial markes used in prostatic imaging, by exploiting signal pileups and
rewinding them radially inwards using different off-resonant
frequency offsets and has the potential to avoid a radiological CT
scan that’s clinically used to differentiate the two in boost
therapy.
Introduction
Magnetic Resonance Imaging is increasingly used in the integration of
image guided radiation therapy planning [1]. Geometric and dosimetric
accuracy of MR-only workflows have also been recently reported [2].
Two current issues in MR-only workflows are identifying implanted
gold fiducial markers (FMs) with positive contrast, and
differentiating the FMs from brachytherapy seeds in patients
undergoing external beam boost [3]. An MR pulse sequence, developed
for visualizing low-dose rate brachytherapy seeds with positive
contrast, centre-out radial sampling with off-resonance reception
(co-RASOR) [4], is proposed here to differentiate the two seeds as
they have different magnetic susceptibilities, with gold being
diamagnetic (χm < 0) and the platinum layer of the brachytherapy seed being
paramagnetic (χm
> 0). This will alter the B0-magnetic field inhomogeneity patterns
observed in the co-RASOR signal model as the signs of ΔBz
are opposite for
each seed. The diamagnetic FMs can reconstructed with positive
contrast using positive off-resonant frequency offsets to rewind
their signal pileups to their geometric centre, whereas the
paramagnetic brachytherapy seed can similarly use a negative
frequency offset.Methods
A gel phantom (3% agarose) was constructed in two layers where two
seeds were placed between layers parallel to the B0-magnetic field
approximately 5 cm apart (Figure 1a). The left seed is a 125I
LDR brachytherapy seed (DraxImage Model LS-1, Quebec, Canada) of
diameter 0.8 mm and length 4.5 mm containing a 10% platinum outer
layer. The right seed is a gold fiducial marker (IZI Medical
Products, MD, USA) of diameter 1.2 mm and length 3.0 mm (Figure 1b).
The MR acquisition protocol used a GE 3T Discovery 750 and a
32-channel head coil and consisted of the following scans: a negative
contrast 3D axial bSSFP (TE/TR=1.3/3.5ms, bandwidth=488Hz/pixel,
FOV=16.0cm2x11cm, 256x256, thickness=1.0mm, NSA=2,
flip=20, scan time=2min 11s); 2D dual-plane co-RASOR (TE/TR=2.8/25ms;
FOV=16.0cm; thickness=3.0mm; matrix=256x804; bandwidth=250 Hz/pixel,
flip=30, scan time=40s).
Each co-RASOR acquisition was reconstructed with off-resonant
frequencies from +/- 1.25 kHz in 50 Hz steps resulting in 51 offsets.
A projection using all frequency offsets was used to view both
seeds in axial and coronal planes. A smaller positive frequency
range, centred about 500 Hz +/ 50 Hz, was used to rewind signal
pileups from the gold fiducial marker, whereas a negative frequency
range, centred at -600 Hz, was used for the brachytherapy seed.
Results
The clinical MR scan, used for negative contrast scan identification
of signal voids due to the FMs is shown as a reference (Figure 2).
The increased signal void of the brachytherapy seed is due to its
stronger magnitude of susceptibility, however specificity of the true
seed location is reduced. A projection through all frequency offsets
followed by a maximum intensity projection through the slices, and
80% thresholding yields both seeds with positive contrast (Figure 3).
Finally, when applying a smaller range of frequency offsets centred
about -600 and +500 Hz, each seed can be individually identified
(Figure 4a,b). The MR signal vs frequency offset of a small ROI placed
in the centre of the brachytherapy seed and gold FM shows the signal
maxima for each seed (Figure 4c).Discussion
The dual-plane co-RASOR sequence is able to differentiate the two
seeds based on their susceptibility difference, as this causes the
signal pileups to have unique patterns around the seeds, seen in the
on-resonant images. Different frequency offsets are able to bring
into focus the hyperintensities at the geometric centre for a single
marker, as shown by the red arrows in Figure 4. When reconstructed at
-600 Hz, the brachytherapy seed has a much higher local signal
intensity (red arrow Fig 4a) and can be isolated by thresholding,
whereas at 500 Hz, the gold FM has the highest local signal and is
also isolated by thresholding. The signal vs frequency plot reveals
that each seed has a signal maxima which is used to tune the
off-resonant frequency offsets. One limitation is that the
hyperintensities must have distinct patterns. In the axial plane
(Fig. 3a) the hyperintensities show a ring pattern around the signal
voids and rewind approximately equally with similar frequency offsets.
The coronal plane (Fig 3b) has signal pileups in a characteristic
dipole pattern with opposite positive and negative lobes which was
exploited to rewind signal pileup at its centre for different
frequencies. Conclusion
The dual-plane co-RASOR sequence is able to differentiate between the
LDR brachytherapy seed and gold FM by exploiting signal pileups and
rewinding them radially inwards using different off-resonant
frequency offsets. This has the potential to avoid a radiological CT
scan that’s clinically used to differentiate these seeds in boosting
therapy.Acknowledgements
This project is based upon work supported by the Ontario Graduate Scholarship (OGS). The authors also thank the financial contributions by Mr. Robert Dunsmore.References
[1]: Baumann et al. (2016) Nature Reviews Cancer 16:234-249
[2]: Tyagi et al. (2017) Phys. Med. Biol. 62:2961-2975
[3]: Tyagi et al. (2017) Radiation Oncology 12:119
[4]: Seevinck et al. (2011) MRM; 65:146