Rolf F Schulte1 and Ana Beatriz Solana1
1GE Healthcare, Munich, Germany
Synopsis
ZTE
acquisition is combined with spin-echo burst encoding for quiet T2 mapping. An
initial ZTE excitation train encodes multiple 3D radial spokes, which get
refocused by reversing the gradients. A double spin-echo leads to T2 decay,
from which T2 maps are extracted by exponential fitting. Accuracy is validated
in the Eurospin TO5 relaxation phantom, while in vivo feasibility is demonstrated
by T2 mapping in healthy brains.
Introduction
Burst
imaging enables relatively rapid and silent encoding in MRI via encoding
multiple k-space lines with a train of short and evenly-spaced RF block pulses [1,2].
Another silent acquisition method is ZTE [3,4,5], where the object is encoded
with 3D radial trajectories, ramping the gradients already before the RF pulse and
updating them with slow, hence silent switching from one spoke to the next one.
In this work, ZTE is combined with spin-echo burst encoding for quiet T2
mapping.Methods
Burst
encoding was implemented into a ZTE sequence by reversing the gradients,
switching off the RF pulses during the burst acquisition part and repeating the
same gradient trajectories. A double spin-echo module was included for T2 mapping
using optimal adiabatic pulses [6]. The pulse sequence, named T2-BURZTE
(Fig.1), was implemented on a GE MR750w scanner. Data is reconstructed
automatically on the scanner using standard 3D gridding and FFT. T2 maps were extracted
by fitting the images of three echo times to an exponential decay. Accuracy was
investigated in a reference relaxation-phantom, consisting of 12 different
vials with different T1 and T2 relaxation times [7]. Each vial was segmented
using region growing, leading to 12 3D masks each containing approximately 2500
voxels. In vivo feasibility was demonstrated by imaging the brain of healthy
volunteers.Results and Discussion
Accurate T2 mapping is possible with T2-BURZTE, as demonstrated in
the reference relaxation phantom (Figs.2 and 3). The impact of different T1
times is relatively small, because of fitting the three echo times, hence
reducing the impact of initial magnetisation. The main impact of long T1 times on
the T2 estimation is a decrease in SNR, hence reduced accuracy.
Brain scans with two different resolutions (2mm)3 and
(1.7mm)3 are shown in Figs.4 and 5, respectively. A main limitation
of T2-BURZTE in vivo is SNR, hence a relatively large flip angle of 5° was
chosen with recovery times of 250ms and 200ms, respectively. The resulting
acquisition times increase rapidly with 3D isotropic encoding, because the
number of spokes is proportional to the squared matrix size and because of the
required waiting times for double spin-echo and recovery at the end.
The power deposition (SAR) is limited because of choosing an optimal
adiabatic pulse obeying both constant adiabaticity and offset independent
adiabaticity with a small sweep width (500Hz) and flip angle (integral of 400°
under magnitude), hence making this sequence a good candidate for T2 mapping also
at higher fields (7T).
Burst encoding never made it into the clinic [2], mainly because of
SNR limitations and calibration requirements. The SNR is improved considerably
by a full 3D-encoding, while the fully compensated gradient trajectory and the
double spin-echo removes unknown phases, hence obviating the need for any
calibration.
This novel T2 mapping method might enable more quantitative and
comparable imaging approaches in MRI, as compared to standard T2 weighted
imaging.
Acknowledgements
No acknowledgement found.References
-
Burst imaging. Hennig J, Hodapp M.
MAGMA. 1993; 1:39-48.
- Burst imaging—Can it ever be useful
in the clinic? Doran SJ, Bourgeois ME, Leach MO. Concepts Magn Reson Part A.
2005; 26A(1):11-34.
- Fast imaging in liquids and solids
with the back-projection low angle shot (BLAST) technique. Hafner S. Magn Reson
Imaging 1994;12:1047-1051.
- Ultra-fast imaging using low flip
angles and fids. Madio DP, Lowe IJ. Magn Reson Med 1995;34:525–529.
- Density of organic matrix of native
mineralized bone measured by water- and fat-suppressed proton projection MRI. Wu Y,
Ackerman JL, Chesler DA, Graham L, Wang Y, Glimcher MJ. Magn Reson Med.
2003;50:59-68.
- Design of a constant adiabaticity
pulse for selective population inversion. Mitschang L, Ittermann B,
Schubert F, Rinneberg H. J Magn Reson 2004;168:103-109.
- Eurospin Test Object
5. http://www.diagnosticsonar.com/mritext.html