Hsuan Wen Yu1,2, Feng Mao Chiu3, Cheng Ping Chien2, and You Yin Chen1
1National Yang-Ming University, Taipei, Taiwan, 2Taipei Beitou Health Management Hospital, Taipei, Taiwan, 3Philips Healthcare, Taipei, Taiwan
Synopsis
Diffusion Tensor Imaging (DTI) is a reliable
tool for investigating renal microstructure and renal function, the imaging stability
remains challenging. Recently, the image-quality improvement by zoom DTI
technique (reduced Field-Of-View diffusion) is reported1. We scanned
10 healthy volunteers by this technique via the respiration-triggered
acquisition, and we assessed different ROIs within the medulla and the cortex
of the kidney. In this study, the reproducibility between different subjects in
zoom DTI was more promising when compared to full-FOV DTI. More DTI scalars were
compared between zoom and full-FOV DTI in cortex and medulla and these may be
potential parameters to detect pathological changes in kidney.
Introduction
Previous studies revealed that Diffusion
Tensor Imaging (DTI) is a reliable tool to provide the information about renal
microstructure and function1. Diffusion anisotropy in the medulla is
significantly higher than the cortex, and medullary Fractional Anisotropy (FA)
demonstrates a highly positive correlation with eGFR (estimated Glomerular
filtration rate)1. However, only the FA and Apparent Diffusion
Coefficient (ADC) values were discussed in related studies2,3, and
geometric distortion and imaging blurring of diffusion tensor Echo Planar
Imaging (EPI) remains challenging due to field inhomogeneity, long echo trains
and the respiratory motion. The zoom diffusion imaging employs non-coplanar
excitation to acquire within a small FOV, and the signal outside the FOV will
not be wrapped into the image. According to recent reports, the zoom diffusion
technique not only improves image spatial resolution, but also reduces the
distortion artifacts2,3. The aim of this study is to assess the
inter-subject reproducibility of DTI parameters, FA, Radial Diffusivity (RD),
Relative Anisotropy (RA), Axial Diffusivity (AD), and ADC in zoom diffusion and
full-FOV images with identical geometric parameters. Material and Methods
10 healthy volunteers (5 men and 5 women;
mean age, 27±3 years) without any renal pathology were examined on a 3T MR
scanner (Philips Healthcare, Eindhoven, The Netherlands), and both zoom DTI and
full-FOV DTI sequences were performed in the same position. The zoom DTI uses
non-coplanar excitation to acquire the signal only inside FOV, and the full-FOV
DTI uses the oversampling way with additional FOVs to prevent the aliasing
artifact. The following were parameters of zoom DTI and full-FOV methods
respectively: TR varies with the respiratory cycle of the subjects. (Range:1-3s),
TE=40/63ms, EPI factor=41/45, SENSE factor=3/1, slice thickness=6mm,
FOV=140x140mm, acquisition matrix=64x64, b-values=0 and 250 s/mm2, diffusion
directions=6, slice orientation=coronal, and phase encoding direction in RL. Each
subject was examined with the respiratory trigger. The image analysis was
performed on the workstation (IntelliSpace Portal, Eindhoven, The Netherlands).
We evaluate the tensor maps (ADC, AD, RD, FA and RA maps) by measuring regions
of interest (ROI) at medullary and cortical area based on b0 image. Two
radiology technologists contoured all ROIs respectively, and both technologists
are with 10 years of experience in radiology. The statistics analysis was
executed on SPSS version 20.0 (SPSS Inc., Chicago, III)
Results
Figure 1 shows a better boundary between cortex
and medulla in zoom DTI. For quantitative analysis, and the averaged DTI
parameters of cortex and medulla are shown in Figure 2. The averaged FA value
in medulla is higher than in cortex, and the mean ADC value in medulla is lower
than in cortex, and these results are similar to previous reports1,3.
In addition, we observed that the ADC and RD values are higher in cortex, and
the AD, FA and RA values are lower in medulla in both zoom and full-FOV DTI. It
shows that the coefficient of variation (CV) of zoom DTI is lower than full-FOV
method (Figure 3). The assessment of inter-observer variability were examined
with paired t-tests, and there is no significant difference (all p
value≥0.801). Discussion and Conclusion
We found that zoom diffusion imaging gives
sharper images, even the same geometrical setting in both two imaging methods,
and this might be caused by the shorter echo trains in zoom DTI. Our result
showed the lower CV in zoom DTI parameters, and it could be the improvement of
the image quality. It is important to get stable measurements for long-term
observation, and zoom diffusion imaging could help to improve the
reproducibility. This technique could be applied to early detection, treatment
response and long-term observation in the renal disease.Acknowledgements
No acknowledgement found.References
-
Lu, Lan, et al. "Use of
diffusion tensor MRI to identify early changes in diabetic nephropathy."
American journal of nephrology 34.5 (2011): 476-482.
- Notohamiprodjo, Mike, et al.
"Diffusion tensor imaging (DTI) of the kidney at 3 Tesla–feasibility,
protocol evaluation and comparison to 1.5 Tesla." Investigative radiology
45.5 (2010): 245-254.
- Palmucci, Stefano, et al.
"Diffusion weighted imaging and diffusion tensor imaging in the evaluation
of transplanted kidneys." European journal of radiology open 2 (2015):
71-80.