2495

R1rho dispersion in human kidney
Ping Wang1 and John C. Gore1

1Radiology and Radiological Sciences, Vanderbilt University Institute of Imaging Science, Nashville, TN, United States

Synopsis

R (=1/T) imaging has been applied in many human organs to characterize tissue biochemical changes. However, R imaging in human kidney has been rarely reported partly due to the challenges associated with field inhomogeneities and respiratory motion. We developed an R imaging protocol for human kidney which used adiabatic half passage pulse and volume shimming to overcome field inhomogeneities. In addition, R dispersion was evaluated via a simple method with a fixed locking time but different locking frequencies. The volunteer scans exhibited characterized R maps in kidney, also there was greater R dispersion between locking frequencies of 100Hz and 300Hz.

Introduction

R (=1/T) is sensitive to slow macromolecular interactions and has been used to evaluate biochemical changes in many biological tissues. R variations with spin-locking fields, termed as R dispersion, provides more complete information of the system chemical and diffusive exchange. 1-3 Recent study showed that R dispersion (represented by the R difference between low and high locking fields) may be relevant to tissue fibrosis in human skeletal muscles associated with aging. 4 R imaging in human kidney is very rare, partially because of the field inhomogeneities and respiratory motion that often lead to significant artifacts. The purpose of this study was to develop quantitative R (and R dispersion) imaging for human kidney aiming to characterize biochemical changes (for instance fibrosis) in kidney diseases.

Methods

The study was conducted on a Philips 3T Achieva scanner (Philips Healthcare, Best, the Netherlands) and two volunteers were tested by a 16-channel torso coil (Philips Healthcare) with breathhold (duration: 21s) to control respiration motion. R imaging was performed with an adiabatic half passage (AHP) based on Hyperbolic Secant functions (Figure 1) combined with volume shimming to overcome field inhomogeneities. Other parameters: FOV = 322x378mm2, pixel size = 2x2mm2, single slice was acquired at a thickness of 4mm, spin-locking time (TSL) = [1, 21, 41, 61]ms, spin-locking frequency (FSL) = 300Hz. TSE (turbo spin echo) sequence was used for data acquisition, with TR/TE = 3000ms/10ms, TSE factor = 15 (low-high profile), fat suppression (SPAIR) was applied, and SENSE factor = 2, NSA = 1, resulting in a scan time of 01m 24s. R data was calculated by fitting the signal to a mono-exponential model S = A·exp(-TSL·R), where S is the acquired MR signal, and A the signal intensity without locking pulse.

To investigate R dispersion in kidney cortex, a simple scheme based on one single TSL = 40ms and different FSLs (=100Hz, 300Hz, 500Hz) was utilized to estimate ΔR as described below:

S = A·exp(-TSL·R), [1]

The division of the signal intensities between the low and high FSLs is

Slow/Shigh = exp[-TSL·(Rlow - Rhigh)] = exp(-TSL·ΔR), [2]

then the dispersion ΔR = -1/TSL · ln(Slow/Shigh). [3]

Results

Figure 2 shows the representative T weighted images for kidneys, the corresponding T and R maps are displayed in Figure 3. The measured values (Mean ± SD) in kidney cortex are T = 110.4 ± 6.0ms, and R = 9.1 ± 0.5Hz. Figure 4 displays the T weighted images at one fixed TSL = 40ms and three different FSLs (= 100Hz, 300Hz, 500Hz), the R dispersion (ΔR) calculated from Eq. [3] is also shown. Finally, the signal intensity change over kidney cortex at different FSLs is plotted in Figure 5. It is seen that signal intensities increased about 12.2% from FSL = 100Hz to 300Hz, resulting in a ΔR ~ 2.3Hz based on Eq. [3], however, there was nearly no signal change above FSL = 300Hz (only 1.2% increase from FSL = 300Hz to 500Hz).

Discussion

We developed an R imaging protocol that utilized AHP pulse combining with volume shimming to overcome field inhomogeneities. This imaging scheme was shown to be able to characterize kidney R value and R dispersion in a very short time, which potentially enables its applications in clinical kidney diseases. The greater R dispersion at lower FSL may reflect there may be more exchangeable protons associated with amides and/or amines than with hydroxyls in kidney cortex.

Acknowledgements

No acknowledgement found.

References

1. Cobb J, Xie J, Gore J. Contributions of chemical and diffusive exchange to T1ρ dispersion. Magn Reson Med. 2013; 69(5): 1357-1366.

2. Spear J, Gore J. New insights into rotating frame relaxation at high field. NMR Biomed. 2016; 29(9): 1258-1273.

3. Wang P, Block J, Gore J. Chemical Exchange in Knee Cartilage Assessed by R1ρ (1/T1ρ) Dispersion at 3T. MRI. 2015; 33(1): 38-42.

4. Wang p, Zhu H, Kang H, Gore JC. R1ρ dispersion and sodium imaging in human calf muscle. MRI. 2017; 42: 139-143.

Figures

Figure 1. Adiabatic half passage (AHP) T-prepulse, the AHP pulse is both amplitude modulated (AM) and frequency modulated (FM). When adiabatic condition is satisfied, the AHP pulse is able to tip the magnetization down by an exact 90° even in the presence of B1 inhomogeneity. A constant amplitude spin-locking block pulse is then applied (with spin-locking time TSL and spin-locking frequency FSL) to generate T contrast. Finally, the magnetization is tipped up by the reverse AHP pulse.

Figure 2. T weighted images of a 52 yo healthy male subject. Images from left to right corresponds to TSL = [1, 21, 41, 61]ms, FSL = 300Hz.

Figure 3. T map (left) and R map (right) of the same subject.

Figure 4. (A) T weighted images at a fixed TSL = 40ms and different FSLs (from left to right: 100Hz, 300Hz, and 500Hz). (B) ΔR calculated using Eq. [3], with FSL = 100Hz and 300Hz.

Figure 5. Signal change over the kidney cortex region at different FLSs (100Hz to 500Hz).

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
2495