Sila Kurugol1, Onur Afacan1, Catherine Seager1, Richard S Lee1, Jeanne S Chow1, and Simon K. Warfield1
1Boston Children's Hospital and Harvard Medical School, Boston, MA, United States
Synopsis
Dynamic Radial VIBE DCE-MRI enables
motion-robust imaging with high spatiotemporal resolution for accurate estimation
of kidney function. However, in
feed and wrap DCE-MRI, bulk motion during infant’s sleep reduces the quality of
images affected by motion and limits clinical utility of this method for
imaging without sedation. This work evaluated the ability of detecting bulk
motion using the center-of-k-space line, removing corrupted volumes, and compensating
for motion using non-rigid registration for improved parameter estimation
accuracy. Results showed that volumes affected by motion were successfully detected
and removed in all patients, and the goodness-of-fit to the tracer kinetic model was improved.
Purpose
To evaluate if the effect of bulk
motion in feed and wrap (FW) Dynamic
Radial VIBE (DRV) DCE-MRI can be
compensated for by using bulk motion removal and non-rigid registration, and if
this
technique improves the glomerular filtration rate (GFR) parameter
estimation.Introduction
Dynamic contrast-enhanced
(DCE) MRI can be used to measure single kidney glomerular filtration rate (GFR)
and differential renal function (DRF) in babies with antenatally detected
hydronephrosis.
DRF and GFR are used by clinicians
to decide who is likely to benefit from surgery. Sedation is required to eliminate
infant motion in the scanner. The risk associated with sedation is a limiting
factor. Therefore we aim to perform FW
DCE-MRI without sedation by compensating for subject motion. The effect of
respiratory motion can be minimized using a Dynamic Radial VIBE (DRV) sequence1. However, bulk motion of infant subjects during sleep prevents accurate GFR
estimation. In this work we propose to use the center of k-space line to detect
and remove DRV volumes corrupted by bulk motion, compensate for the effect of
motion using non-rigid registration, and compare our motion-corrected filtration
rate parameter estimation to standard DRV.Methods
FW DCE-MRIs of 4 infants (1.9-4.5months) and non-FW
imaging of 2 children were performed at 3T using Gadavist. Infants were fed, swaddled and rocked to sleep. Images were
then acquired for six minutes using a radial “stack-of-stars” 3D FLASH sequence
(TR/TE/FA 3.56/1.39ms/12o, 32 coronal slices, voxel
size=1.25x1.25x3mm). The mean temporal resolution was 3.3 sec for the arterial
phase (2 minutes) and 13 sec for the remaining phases (4 minutes). Bulk motion
events were present in the image series. We first registered each volume in the
dynamic image series to a reference volume in image space2. The center of
k-space point of each radial line was then used to detect motion3,4. The center of k-space points in 3D corresponds
to the center of a k-space line. We
first applied principal component (PC) analysis to the center of k-space line and
kept 2 PCs from each channel. We then constructed the measurement vector (m(t))
of size 2xnumber of channels at each time point t. We computed the correlation
coefficient (CC(t)) between m(t) and m(t-1). 1-CC(t) was used to detect motion5. 1-CC(t) larger than a selected threshold indicated a motion event. Volumes
corrupted by bulk motion were removed before tracer kinetic (TC) model fitting6 and filtration rate parameter estimation. Results
We compared
the model fitting performance of the original signal without motion compensation
(MC) with signal after motion compensation with non-rigid registration and
after additional compensation with removal of bulk motion in Figure 1. The
model-fitting root mean square (RMS) errors were reduced from 0.21 and 0.16 to 0.12 and 0.12 for
right and left kidneys respectively after non-rigid registration. The volumes
corrupted by bulk motion were removed (Figure 2). The errors were further
reduced to 0.09 and 0.09 after removal of bulk motion. In all 6 cases, motion-corrupted volumes were successfully
detected and removed resulting in improved model fitting and parameter
estimation.Conclusions
We demonstrated
that the bulk motion events during feed and wrap DCE-MRI reduce the accuracy of
TC model estimation and the reliability of computation of important clinical
markers of kidney function (GFR and DRF). The model-fitting error was reduced
after removal of volumes affected by bulk motion and registration of volumes to
a common reference image. The proposed
DRV bulk motion detection and removal method is a potentially useful method for
non-sedated imaging of infants with hydronephrosis and potentially impaired
renal function without exposing patient to radiation, as opposed to clinical
nuclear scintigraphy.Acknowledgements
This
work is supported by Crohn’s and Colitis Foundation of America’s Career
Development Award, the National Institute of Diabetes and Digestive and Kidney
Diseases (NIDDK) of the NIH under award R01DK100404 and by the National Institute of Biomedical Imaging and Bioengineering (NIBIB) of
the NIH under award R01EB019483.References
1. Feng L, Grimm R, Tobias Block K, et al.
Golden-angle radial sparse parallel MRI: Combination of compressed sensing,
parallel imaging, and golden-angle radial sampling for fast and flexible
dynamic volumetric MRI. Magn Reson Med. 2014;72: 707–717.
2.
Commowick,
O., Wiest-Daessl´e, N., Prima, S., 2012. Automated di↵eomor- 529 phic registration of
anatomical structures with rigid parts: Application to 530 dynamic cervical
MRI, in: MICCAI. Springer, pp. 163–170.
3.
Cruz,
G., Atkinson, D., Buerger, C., Schaeffter, T. and Prieto, C., 2016. Accelerated
motion corrected three‐dimensional
abdominal MRI using total variation regularized SENSE reconstruction. Magnetic
resonance in medicine, 75(4), pp.1484-1498.Vancouver
4.
Feng,
L., Axel, L., Chandarana, H., Block, K.T., Sodickson, D.K. and Otazo, R., 2016.
XD‐GRASP:
Golden‐angle
radial MRI with reconstruction of extra motion‐state dimensions using compressed
sensing. Magnetic resonance in medicine, 75(2), pp.775-788.
5.
B. Stemkens et al., “Adaptive bulk motion exclusion
for improved robustness of abdominal magnetic resonance imaging,” NMR Biomed., vol. 30, no. August, p.
e3830, 2017.
6.
Sourbron,
Steven P., et al. "MRI-measurement of perfusion and glomerular filtration
in the human kidney with a separable compartment model." Investigative
radiology 43.1 (2008): 40-48.