Matthew M. Willmering1, Ryan K. Robison2, Hui Wang3, James G. Pipe4, and Jason C. Woods1,5
1Center for Pulmonary Imaging Research, Cincinnati Children’s Hospital Medical Center, Cincinnati, OH, United States, 2Phoenix Children’s Hospital, Phoenix, AZ, United States, 3Philips Healthcare, Gainesville, FL, United States, 4Barrow Neurological Institute, Phoenix, AZ, United States, 5Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, United States
Synopsis
MRI
of lungs is inherently challenging due to the short T2* and
intrinsic motion from the respiratory and cardiac cycles. Ultrashort echo-time (UTE)
sequences are often implemented for their shorter echo times and relative
insensitivity to motion. Spiral UTE sequences have been touted recently as
having greater k-space sampling efficiencies than radial UTE, but few are
designed well for the shorter T2* of lung. In this study, FLORET (Fermat looped,
orthogonally encoded trajectories), a recently-developed spiral 3D UTE
sequence, was implemented in human lungs for the first time and outperformed
traditional radial UTE for imaging of lung tissue.
Purpose:
Diagnostic-quality MRI of
the lungs is challenging due to low lung parenchymal proton densities, intrinsic
physiologic motion (respiratory/cardiac cycles), and inherently short T2*
of the lung parenchyma (approximately 0.75 msec at 3T1). Ultrashort echo-time (UTE) MRI
sequences overcome the loss of signal due to T2* by
shortening the echo time to less than 100 µs in many cases. However, most radial
UTE acquisitions require long scan times and oversample small k-values. Spiral
trajectories to acquire k-space have been shown to be more robust to motion
artifacts2 but typically require long T2*
components to image well. Recently, FLORET (Fermat looped, orthogonally encoded
trajectories), a spiral acquisition technique, was developed and exhibits high
signal-to-noise and sampling efficiencies compared standard 3D UTE sequences.3,4 We demonstrate the ability of the
FLORET sequence to obtain faster and higher-quality lung images in human
subjects compared to traditional “fast” UTE techniques (here, stack-of-stars (SOS)).Methods:
FLORET is based on the
Fermat spiral to avoid oversampling low k values, allowing for higher sampling
efficiencies4. Each spiral is projected onto a single
cone (between +45° and -45°). Two orthogonal sets of cones were acquired to fully
measure k-space. The initial angle of the spiral out of k=0 was rotated via the
golden angle. Additionally, rapid gradient spoiling (directly after the
readout), was implemented.
The FLORET sequence
calculates ideal spiral parameters to maximize sampling efficiency for a given
systems gradient amplitude and slew rate (31 mT/m and 74 mT/m/s used here,
respectively on Philips 3.0 T, R5.1.7). The chest of a healthy adult male
subject (age 33) were imaged using a Philips 32 channel anterior/posterior chest
coil during free-breathing. The scan parameters were: 0.093 ms echo time (TE),
5° flip angle, 3.92 ms repetition time (TR), 1.489 mm isotropic voxels, 268
isotropic matrix, 43,112 total acquisitions, and a 100% sampled k-space. The
full acquisition of FLORET took just under 3 minutes.
Radial stack-of-stars
was acquired on the same subjects with the same parameters, coil, and magnet, though
the minimum TE was very slightly greater, at 0.122 ms. SOS allows for one non-isotropic
dimension, allowing for fewer slices (160) in the foot to head dimension, thus
saving time. However, the full image required 85,760 FID acquisitions and 7
minutes for the full acquisition.
All
acquisitions were gridded and reconstructed similarly using Graphical
Programming Interface (GPI).5 The GPI reconstruction implemented a
sampling-density correction to grid the acquisitions into a Cartesian matrix.
Following 3D FFT, images were corrected for off-resonance effects and B1
inhomogeneity.
Results:
Both acquisition
techniques result in diagnostic-quality lung images, with parenchyma signal visible,
as seen in Figure 1, despite the 2.5x difference in acquisition length.
Comparing the images, FLORET outperforms SOS in some image quality metrics. The
signal to noise ratio (SNR) of the lung parenchyma was 4.3 for FLORET and 2.9
for SOS UTE, and the parenchyma/muscle ratios were 0.34 for FLORET and 0.42 for
SOS. The vasculature was
visualized with
greater contrast using FLORET, as illustrated in Figure 2. SOS UTE exhibited
stronger non-Cartesian artifacts compared to FLORET, as seen near the heart in
the axial views of Figures 1 and 2. However, the proprietary image
reconstruction removes the artifacts and improves SNR. Both methods appear to
perform similarly with respect to motion artifacts during these free breathing
images, though FLORET appears to have more noticeable off-resonance affects
(most notable near the spine in these images), which may be overcome by
advanced focusing techniques.
Importantly, the SNR and vascular contrast per rf excitation was higher in FLORET than in SOS, providing a time and/or
spatial resolution advantage.Discussion:
By using GPI-reconstruction,
direct comparisons of image quality for the sequences can be performed. FLORET
was near-equivalent to SOS UTE even with a much smaller number of k-space
trajectories. Non-Cartesian artifacts are much less noticeable with FLORET but
the longer FID acquisition time makes the off-resonance effects more
noticeable. Importantly, the ~2.5 times faster image acquisition (at identical
spatial resolutions) is ideal for less-compliant or pediatric patients.Conclusion:
The FLORET sequence allows for faster acquisition of high
diagnostic-quality lung images and its short T2* components
without sacrificing signal to noise or quantification of abnormalities. This
sequence has potential for clinical and research translation in pulmonary MRI, particularly
in pediatric or less-compliant adult patients with the shortened scan time. The
ability to quantify lung parenchyma accurately with a short scan time can
facilitate phenotyping in both emphysematous (e.g. COPD, BPD) and fibrotic
(e.g. IPF, NSIP) pulmonary conditions.Acknowledgements
The
authors thank the following sources for research funding and support: NIH R01 HL131012 and NIH R44 HL123299.
Additionally, the authors would like to thank Ashley G. Anderson III PhD
(Philips Healthcare) for the ability to implement their off resonance focusing
code.References
1. Roach DJ, Crémillieux Y,
Serai SD, et al. Morphological and quantitative evaluation of emphysema in
chronic obstructive pulmonary disease patients: A comparative study of MRI with
CT. J Magn Reson Imaging. 2016;44(6):1656-1663. doi:10.1002/jmri.25309.
2. Salerno M, Altes TA,
Brookeman JR, Lange EE De, Mugler III JP. Dynamic Spiral MRI of Pulmonary Gas
Flow Using Hyperpolarized 3He : Preliminary Studies in Healthy and Diseased
Lungs. Magn Reson Med. 2001;46:667-677.
3. Pipe JG, Zwart NR,
Aboussouan EA, Robison RK, Devaraj A, Johnson KO. A new design and rationale
for 3D orthogonally oversampled k-space trajectories. Magn Reson Med.
2011;66(5):1303-1311. doi:10.1002/mrm.22918.
4. Robison RK, Anderson AG,
Pipe JG. Three-dimensional ultrashort echo-time imaging using a FLORET
trajectory. Magn Reson Med. 2017;78(3):1038-1049. doi:10.1002/mrm.26500.
5. Zwart NR, Pipe JG.
Graphical programming interface: A development environment for MRI methods. Magn
Reson Med. 2015;74(5):1449-1460. doi:10.1002/mrm.25528.