Thomas Christen1, Jia Guo1, Wendi W. Ni1, Audrey P. Fan1, Michael M. Moseley1, and Greg Zaharchuk1
1Radiology, Stanford University, Stanford, CA, United States
Synopsis
A new MR approach has been proposed to obtain simultaneous
measurements of blood oxygen saturation (SO2) and hematocrit (Hct)
by measuring and combining blood MR relaxation times. Although the first
results were encouraging, the method has not been properly validated. In this study,
we tested this approach in 10 volunteers subjected to gas challenges with the
intent to modify SO2 while keeping Hct constant. The method was also
tested in 10 Moyamoya patients and compared to photometric analysis. Results
suggest that reliable MR estimates of both SO2 and Hct can be
obtained in vivo.
Introduction:
An MR approach has been recently proposed to obtain
simultaneous measurements of blood oxygen saturation (SO2) and hematocrit
(Hct) by combining two or more in vivo
estimates of blood MR relaxation times1 (Fig.1a). Although the first
results obtained in 5 healthy volunteers were encouraging (with average
estimates in agreement with literature values), the method has not been
properly validated. In this study, we tested the approach in 10 volunteers
subjected to normoxic, hyperoxic and hypoxic gas challenges with the intent to
modify SO2 while keeping Hct constant. The method was also tested in
10 Moyamoya patients and compared to photometric analysis. Materials and Methods:
The local IRB committee
approved all studies. Experiments were performed at 3T (GE Healthcare Systems,
Waukesha, WI) with an 8-channel GE receive-only head coil. The MR acquisition
protocol included: (1) a blood T2 mapping sequence based on the TRUST approach2
(acq matrix=64x64, FOV=22x22cm2, NA=4, TR=10s, TEs=20-40-60-100ms,
Acq time=264s). (2) a blood T1 mapping sequence proposed by Valera et.al3-4
using an adiabatic nonselective 180° global inversion pulse, followed by a thin
single-slice selective Look-Locker echo-planar imaging (acq matrix=96x96, FOV=22x22cm2,
NA=6, TR=10s, first inversion time TI=50ms, ∆TI=150ms, 60 TI values, Acq
time=70s). Calibration curves1 were used to convert (T1, T2)
measurements into (SO2, Hct) estimates. Two types of experiments were
performed:
(1) Changes during Gas challenges: 10 volunteers were scanned
while breathing different gas mixtures (Hyperoxia (100%O2), Normoxia
(21%O2), hypoxia (14%O2)). Each epoch lasted 6 minutes
and acquisitions started after 1 min to ensure stable gas environment and
subject physiology. Hct
and SO2 were measured during each epoch for all volunteers (see
Fig.1b).
(2) Correlation with blood
gas analysis: 10 Moyamoya patients were scanned while breathing normoxic
gas. Before and after the MR acquisitions, small blood samples (<0.5mL)
taken from the patient’s IV were analyzed with a hemoglobin Meter (HemoPoint®,
Stanbio).
Results:
Averaged values, standard deviations as well as
individual estimates are shown in Figure 2 for all gas challenges. In normoxic
conditions, SO2 values measured in the superior sagittal sinus (SSS)
(55±13%) are in line with previous reports5. Hct results (45±9%) are
also in agreement with standard laboratories reports6. One can
notice (Fig.2b-2d) that higher SO2 and Hct values were obtained in
male volunteers6. Hyperoxic challenges resulted in a significant
increase of SO2 values (63±12%, p<0.01) while hypoxic challenges
resulted in a significant decrease (48±11%, p<0.01). On the contrary, Hct
estimates stayed stable during the challenges. A good correlation (high
correlation coefficient and correlation line close to unity) was found in
Moyamoya patients between MR and photometric measurements of Hct (Fig.3).Conclusion:
This study suggests that reliable MR estimates of
both SO2 and Hct can be obtained in
vivo. Several technical improvements can be foreseen (fast whole brain acquisition,
custom calibration models, combination of 3 or more parameter estimates, etc.)
and further validation steps are still required (analysis of sensitivity and
repeatability, effect of partial volume, etc.). Nonetheless, this approach
seems promising for non-invasive detection of hypoxic tissues in several
pathologies and could be used to calibrate other MR techniques such as Arterial
Spin Labeling.Acknowledgements
Supported in part by (NIH 5R01NS066506, NIH
2RO1NS047607, NCRR 5P41RR09784, R21-NS087491).References
[1] Christen et.al, Proc.
ISMRM, 2017 [2] Lu and Ge, MRM, 2008. [3] Varela et.al, NMR Biomed 2011 [4]
Zhang et.al, MRM 2013 [5] Jain et.al, JCBFM 2010 [6] University of Iowa
Diagnostic Laboratories (UIDL).