Johannes Strasser1, Franz Fazekas1, and Stefan Ropele1
1Medical University of Graz, Graz, Austria
Synopsis
In
Magnetic Resonance Elastography (MRE), mechanical tissue parameters are
assessed by sampling shear wave propagation via a set of motion encoded phase
offset images. We here investigate a multiple phase offset image acquisition
strategy based on displacement encoding with stimulated echoes (DENSE) for multi-slice
human brain MRE together with a low frequency mechanical excitation and short
echo times. Clear wave images could be acquired using the proposed imaging
approach and estimates of the magnitudes of the complex shear modulus could be
calculated from the derived wave image data set.
Introduction
In magnetic
resonance elastography (MRE), mechanical tissue properties are assessed by
gathering information about mechanical wave propagation induced by externally
applied vibrations [1-5]. In conventional MRE, tissue displacement is assessed
by using bipolar motion encoding gradients at distinct offsets to the
mechanical vibration in order to get a time-resolved series of phase offset
images showing the propagation of the waves. Alternatively, tissue displacement
imaging can be performed via displacement encoding with stimulated echoes
(DENSE) [6] which has been proposed for human heart MRE [3]. We recently have presented
an accelerated DENSE-MRE approach with multi phase offset readout acquisition [4].
This approach is very beneficial for low frequency excitations which are
receiving increased interest in human brain MRE [1,5]. Such vibrations
penetrate deeper into the brain due to less damping and thereby regions of low
wave amplitude are minimized. Consequently they allow the evaluation of larger
cross-sectional areas in the calculated elastograms. While conventional MRE needs
long echo times (TEs) for the motion encoding of low frequency mechanical
excitations, the encoding gradient duration in DENSE-MRE is not coupled to the
wave period [3,4]. Additionally, using the multi phase offset readout strategy,
the complete time resolved wave image series can be acquired in a single
sequence run. Thus besides TE also the acquisition time (TA) and thus possible
unwanted patient movement related artifacts can be reduced. We here investigate
the application of the multi phase offset readout DENSE approach for multi-slice
MRE in the human brain.
Methods
MRE acquisitions and viscoelastic properties
were investigated in the brain of two healthy volunteers using the multi phase
offset readout DENSE approach. A 3T MRI scanner (MAGNETOM Prisma fit, Siemens
Healthcare, Erlangen, Germany) and a 20ch head coil were used for all
experiments. Sinusoidal mechanical vibrations at 20Hz were excited via a
piezoelectric actuator and released to a head cradle. The excitation was
triggered by the scanner each 50ms to achieve a continuous vibration over the
whole sequence run. All of the 4 time resolved phase offset images for one
encoding direction were acquired in a single run with proper mixing time (TM) settings
(TM=8.01ms+[0/12.5/25/37.5]ms). A variable flip angle scheme was used to take
an average T1 signal decay at the different TMs into account (α=30°/35°/45°/90°).
Further imaging parameters were: FOV=300x300mm2, matrix=128x128,
slice thickness=4mm, TE=7.71ms, TR=2500ms, 5 slices, GRAPPA parallel imaging
factor 2, Partial Fourier 6/8, TA=2:30min per encoding direction. The sequence
was repeated three times with altered displacement encoding axis x, y and z.
All raw phase images were spatially and temporally mean filtered to eliminate
background phase information and ensure clear visibility of the imaged waves.
First harmonics were extracted after temporal Fourier Transform of the time
resolved images for each encoding direction. Estimates of the complex shear
modulus were obtained using an MDEV
[5] reconstruction.Results
Low
frequency excited displacement encoded phase images could be acquired in all
encoding directions using the multi phase offset readout DENSE-MRE approach.
The phase images are superimposed with background phase information in the
first place as shown for the 4 time resolved phase offsets of one encoding
direction of a volunteer (Fig. 1a). Elimination of the background phase by
mean filtering revealed the wave propagation information (Fig. 1b). First
harmonic selection revealed the 20Hz-corresponding complex wave images (Fig. 2).
Estimates of the magnitude of the complex shear modulus were obtained for all 5
slices (Fig. 3). Brain tissue’s mean values of the magnitude of the complex
shear modulus over all slices were 0.37 kPa and 0.32 kPa for the two healthy
volunteers.Discussion and conclusion
Our
results show that brain MRE acquisitions at low frequencies can be reliably performed
using the proposed multi phase offset readout DENSE approach. The acquisition
scheme is able to capture low vibration wave fields in the brain in a fast way while
keeping TE short and with no susceptibility artifacts. Multi phase offset as
well as interleaved multi-slice acquisition is possible in a single sequence
run.Acknowledgements
References
[1] Dittmann F et
al., MRM 76:1116-1126 (2016)
[2] Muthupillai R et
al., Science 269:1854-1857 (1995)
[3] Robert B et al., MRM
62:1155-1163 (2009)
[4] Strasser J et
al., Proc ISMRM 25:1370 (2017)
[5] Streitberger KJ et al.,
PLoS ONE9(10):e110588 (2014)
[6] Aletras AH et al., J Magn Reson 137:247-252 (1999)