Waqas Majeed1 and Arunark Kolipaka1
1Radiology, The Ohio State University Wexner Medical Center, Columbus, OH, United States
Synopsis
We demonstrate the feasibility of combining simultaneous multi-slice (SMS) excitation with in-plane acceleration to achieve highly accelerated MR elastography data acquisition. The proposed approach enables the acquisition of diagnostic liver MRE data in a single breath-hold, which was not possible using the previous approaches. Our results indicate excellent agreement between the data acquired with and without SMS.
Introduction
Magnetic Resonance Elastography (MRE) is
increasingly being utilized for the diagnosis and staging of liver fibrosis. Current clinical protocol for
liver MRE data acquisition requires four ~18 second breath-holds to acquire
4-slice diagnostic gradient recalled echo (GRE) MRE data 1. Chamarthi et al. proposed a method (referred to as rapid
MRE or MREr) to reduce the acquisition time by acquiring data with same motion
encoding gradient (MEG) polarities during opposing phases of mechanical
vibration 2. Although MREr results in shorter breath-holds, multiple breath-holds are
still needed for acquisition of diagnostic MREr data. We propose
incorporation of simultaneous multislice acquisition into MREr (SMS-MREr) to
further reduce the acquisition time, thereby enabling the acquisition of the
required images within one 16 second breath-hold. We demonstrate that
SMS-MREr can be used to obtain diagnostic quality liver elastograms and report
excellent agreement between the results obtained using SMS-MREr and MREr. The
use of SMS without in-plane under-sampling to accelerate MRE acquisition has been reported
previously 3. However, to our knowledge, this is the first
study that utilizes SMS along with in-plane acceleration for MRE.
Methods
23 healthy volunteers were imaged on a 1.5T MRI scanner (Magnetom Aera, Siemens Medical Solutions Inc., Erlangen, Germany). A commercial pneumatic driver system was used to
introduce 60 Hz mechanical vibrations in subjects’ liver (Resoundant, Mayo
Clinic Foundation, Rochester, MN).
We used MREr 2 and the developed SMS-MREr
sequences to acquire MRE data. Briefly, MREr reduces the scan time (relative to
the standard GRE based sequence) by playing MEGs with same polarity during
opposing phases of mechanical vibration cycle. SMS-MREr further reduces the scan time by
utilizing multiband excitation pulses. RF-based CAIPIRINHA was implemented to reduce g-factor penalty 4.
The following imaging parameters were identical for both the
MREr and SMS-MREr sequences: TE = 21.4 ms; TR =25 ms; FOV = 360 mm; Flip angle
20°; slice thickness = 5 mm; 4 slices; matrix = 128x64; vibration frequency =
60Hz; 4 MRE time offsets; MEG of 16.67 ms duration (60Hz) along the slice direction.
Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA)
acceleration factor of 2 with 16 reference lines was utilized in MREr to reduce
the breath-hold duration per slice to 8 seconds. For SMS-MREr, we used GRAPPA
acceleration factor of two with 8 reference lines which were not used in the reconstruction. This
resulted in 14 second acquisition for 4 slices. A two second long low
resolution GRE based reference scan was acquired immediately before SMS-MREr
acquisition to serve as a reference scan for split slice-GRAPPA reconstruction 5, resulting in a total
breath-hold duration of 16 seconds for the acquisition of 4 slices. Images were constructed using custom developed MATLAB scripts. Stiffness maps were generated using MRE Lab (Mayo Clinic, Rochester, MN). Identical ROIs were manually drawn on SMS-MREr and MREr derived stiffness maps to obtain average stiffness values.Results
Magnitude, wave snapshot and stiffness images from volunteers with elevated and normal liver stiffness are shown in Figures 1 and 2 respectively. Visual inspection suggests that the
simultaneously acquired slices were successfully separated using the split
slice-GRAPPA algorithm. Magnitude, wave and stiffness images acquired using
both methods show excellent visual agreement. Indeed, excellent linear relationship between the mean
stiffness values obtained using MREr and SMS-MREr is observed (Figure 3), albeit with a slight negative bias in SMS-MREr (p-value = 0.014). Additionally,
Bland-Altman analysis demonstrates excellent agreement between the stiffness
values obtained using MREr and SMS-MREr for all subjects except one. Maximum difference of 0.36kPa was observed between MREr and SMS-MREr derived stiffness values.Discussion and Conclusions
In
this study, we demonstrate, for the first time, that it is possible to combine
SMS and in-plane acceleration for diagnostic MR elastography. This, combined
with the rMRE approach, enables diagnostic MRE data acquisition with
unprecedented speed. The stiffness values obtained with MREr and SMS-MREr
exhibit agreement in general. The negative bias observed for SMS-rMRE may be
caused by the reduction in SNR due to g-factor penalty, since an inter-slice
gap of 0mm was used in this study.
The inherent variation between
breath-holds results in variation in the orientations of the slices acquired
during multiple breath-holds. Also, the inter-slice gap may not be uniform
across the slices. The proposed approach overcomes these limitations on the
anatomical precision by enabling the acquisition within a single breath-hold.
Future work
will focus on further improving the SMS-rMRE image quality by optimizing sequence
parameters and RF profile.
Acknowledgements
Authors would like to thank NIH-NHLBI for grant support (R01HL124096). References
- Dzyubak, B., et al. (2016). "Automated liver
elasticity calculation for MR elastography." Journal of Magnetic
Resonance Imaging 43(5):
1055-1063.
- Chamarthi, S. K., et al. (2014). "Rapid
acquisition technique for MR elastography of the liver." Magnetic
resonance imaging 32(6):
679-683.
- Guenthner, C., et al. (2017). "Simultaneous Multislice Acquisition for Magnetic Resonance Elastography." Proceedings of 25th ISMRM Annual Meeting: Abstract 1137
- Breuer, F. A., et al. (2005). "Controlled
aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for
multiāslice imaging." Magnetic resonance in medicine 53(3): 684-691.
- Cauley, S. F., et al. (2014). "Interslice
leakage artifact reduction technique for simultaneous multislice
acquisitions." Magnetic resonance in medicine 72(1): 93-102.