Jianxun Qu1, Tianye Lin2, Xiaocheng Wei1, Bing Wu1, and Feng Feng2
1GE Healthcare, Shanghai, China, 2Radiology, Peking Union Medical College Hospital, Beijing, China
Synopsis
Phase cycling was used to address the banding artefact in DANTE prepared black blood imaging. Simulation, phantom and in-vivo experiment were performed to illustrate and validate the effectiveness
Introduction
Delayed alternating with nutation for
tailored excitation (DANTE) preparation module has gained increasingly popularity
in black blood imaging. An essential steady state free precession (SSFP) sequence,
composed of non-selective radiofrequency pulse (RF) and flow dephasing gradient
of the same moment and polarity, is used to achieve flow suppression while
maintain steady state for static tissue. Banding artefact, as induced by
constant phase offset in present of dephasing gradient, persists in DANTE
prepared images. Increasing dephasing gradient moment could reduce the spatial
period of banding and thus make the artefact inconspicuous. However, banding
was merely hided but not truly eliminated and its effectiveness was usually compromised
at the cost of increased eddy current and prolonged unit duration. In cases
where dark band resides adjacent and parallel to small vessels, the lumen size
could potentially be exaggerated. Phase cycling is a conventional method adopted
in SSFP for banding suppression. In high resolution vessel wall imaging (VWI),
multiple acquisition is usually performed to guarantee sufficient signal to
noise ratio (SNR). This study aims to achieve banding free DANTE prepared VWI
incorporating multi-acquisition and phase cycling.
Method
The spatial period of banding artefact is
determined by the dephasing gradient moment in DANTE preparation. Such banding
pattern could be shifted via changing the phase increment of the series RF
pulses. In multiple NEX acquisition, banding patterns with different offset could
be averaged to reduce the spatial period and thus reduce the artefact. A phase
cycling scheme could be adopted for phase increment in multiple NEX acquisition.
Simulation, phantom, and in-vivo exam were performed to illustrate and validate
its improvement. First, banding artefact pattern with phase increment of 180
degree and 0 degree was simulated to visualize the pattern shift. The
simulation system parameters are: T1/T2 1000/100ms, dephasing gradient moment
0.714 pi/mm, flip angle 10 degree. The DANTE preparation unit was repeated for
100 times with a time step of 2us. Then, DANTE prepared 3D fast spin echo (FSE)
sequence (Figure.1A) with and
without phase cycling was implemented in a 3.0T whole body system (Discovery
750, GE) for phantom and in-vivo study. The DANTE preparation parameters for
phantom validation are: dephasing gradient moment 0.563 pi/mm in readout
direction, flip angle 10 degree, repeated number 100, unit time 2ms. The
banding artefact was visualized by 3D FSE sequence with in-plane resolution of
0.5*0.5mm. Later, a healthy volunteer was recruited with consent acquired prior
to examination. In-vivo experiment was performed with clinically adopted settings:
DANTE gradient moment of 1.126 pi/mm in slice selective, phase encoding, and readout
direction with flip angle, unit duration, and repetition number the same as
that of phantom study. A sagittal slab was prescribed for cranial black blood
imaging with TE/TR 13/700ms, FOV 200*200mm, Matrix 288*288, slice thickness
1.2mm, slice number 64, echo train length 32, parallel imaging acceleration factor
2, and average number 2. The banding artefact with and without phase cycling
was evaluated visually. Results
The simulation result with different phase
increment scheme was presented in Figure.1B.
The spatial period of banding artefact was 2.8mm, corresponding to (2pi) /(0.713pi/mm).
A half period shift was observed between #1 (phase increment 180) and #2 (phase
increment 0). In averaged pattern, the banding period was reduced by half and
the depth of dark band was also elevated, which was also validated in phantom
study. Figure.2A adopted single phase
increment of 180 degree, while Figure.2B
used phase-cycling scheme. In-vivo experiment result was shown in Figure.3. In
zoomed region of interest from different slices (Figure.3 A D), conspicuous banding artefact was observed in images
without phase-cycling (Figure.3 B E),
compared to those with (Figure.3 C F).Conclusion
In this work, multiple acquisition and phase
cycling are incorporated and implanted in DANTE prepared VWI. Simulation and
phantom experiment were used to demonstrate and validate its effectiveness. The
signal lost in the banding artefact was retrieved with multiple acquisition and
phase cycling. In-vivo exam was also performed. Banding artefact was greatly
reduced or eliminated with phase cycling modulation. Its effectiveness is to be
further evaluated, especially for small vessels. Acknowledgements
No acknowledgement found.References
[1] Linqing Li, and etc. Radiology, 2014
[2] Linqing Li, and etc. MRM, 2012