In this work, we quantitatively studied the impact of the local shim coil to RF coil in a combined B0 shim coil and RF coil system in terms of SNR, transmit B1+ and receive B1. By using the results as a design guideline, a 5-channel shim coil was constructed, of which interferences on RF coils were minimized with the appropriate shim coil diameters, number of turns and distances between the shim coil and the sample.
Introduction
In MRI, shim coils can compensate for B0 variations to avoid geometric distortion and signal dropout. This is particularly important in echo-plane imaging (EPI), susceptibility weighted imaging (SWI) and T2* weighted imaging1. However, in an integrated B0/Rx coil system2, losses in SNR and reductions in transmit coil efficiency were observed due to the addition of B0-shim coils1. In this work, the perturbations of the transmit field (B1+ field) and receive sensitivity were firstly investigated quantitatively by using a loop with different diameters, turns and distances from the sample. Based on the study, a 5-channel shim coil was constructed for small animal imaging at 3T and then the transmit efficiency of RF coil and SNR losses were evaluated through imaging experiments on a 3T MR scanner.Methods
As a simplified model, a circular copper loop that connected to the shim power controller was placed on the surface of a 4 cm diameter phantom, as shown in Figure 1 a. The impacts of circle loops with different diameters (D=3 cm, 4 cm, 5 cm, 6 cm), turns (N=1, 2, 4, 8) and distances from the phantom (d=1 cm, 2 cm, 3 cm, 4 cm) on SNR and B1+-field were evaluated at a 3 T Siemens MRI system. A 2D GRE sequence (TR/TE=500 ms/10 ms, flip angle=60o, FOV=100 mm, matrix size =64x64, slice thickness=5 mm, slices=1, bandwidth=390 Hz/pixel) was applied to acquired images for SNR calculation with equation: sqrt(2)·sgnal/σ (σ is the standard deviation of noise image). To assess the transmit field, the double-angles method3 was applied, of which images were acquired with a 2D GRE sequence (TR/TE=3000 ms/10 ms, flip angle=45o/90o). To diminish the impact of transmit field magnitude on SNR comparison, the normalized SNR4 was calculated to evaluate the receive sensitivity. However, multi-coil shimming is more efficient in practical applications than a single loop shimming. Hence, based on the above investigations, the interferences of a 5-channel shim coil that designed for small animal temperature mapping at 3T with RF coils were also studies. The coil elements with a 4 cm diameter were placed on a polycarbonate material former, which had 2 cm distances away from the phantom, as shown in Figure 1 b. Notably, the 5-channel shim coil was set with the optimized currents to shim B0-field corresponding with the actual application. Additionally, the maps of the shim coil with the transmit voltage set as the voltage loading only phantom was acquired.Discussion
The degradation of SNR, transmit efficiency and the receive sensitivity with more loop turns may be caused by the close distance from the sample, which can be reduced by increasing the distance between coil and phantom or decreasing the diameter of coil element. Most importantly, the B0-shim performance of the multi-coil should be focused since the interferences with RF coils can be neglected in an appropriate design.Conclusion
Interferences of local B0-shim coils with RF coils were investigated experimentally, finding that a local shim coil with less number of turns, smaller coil loop sizes and farther distances to imaging samples would generate less interferences to RF coils. By using this conclusion as a design guideline, a 5-channel shim coil was designed and constructed. The impact of the shim coil on the RF coils for this specific coil system in the aspects of SNR, B1+ field and receive sensitivity was minimal and could be omitted in practice.1. J.P. Stockmann, L.L. Wald. In Vivo B0 Field Shimming Methods for MRI at 7T [J]. NeuroImage, 2017. http://dx.doi.org/10.1016/j.neuroimage.2017.06.13.
2. J.P. Stockmann, T. Witzel, J. Blau, J.R. Polimeni, W. Zhao, B. Keil, L.L. Wald, Combined shim-RF array for highly efficient shimming of the brain at 7 Tesla, ISMRM, 2013, 21:665.
3. C.H. Cunningham, et al. Saturated double-angle method for rapid B1+ mapping [J]. Magnetic Resonance in Medicine, 2006, 55(6): 1326-1333.
4. X. Yang, et al. A Scalable Constellation Coil Design for 3T Body Imaging. ISMRM, 2013, 4388.