Michael Hock1, Maxim Terekhov1, David Lohr1, Maria Roxana Stefanescu1, Anja Schröder2, Heike Walles2, Christoph Juchem3, and Laura Maria Schreiber1
1Chair of Cellular and Molecular Imaging, Comprehensive Heart Failure Center (CHFC), University Hospital, Wuerzburg, Germany, 2Translational Center Regenerative Therapies (TLC-RT), Fraunhofer Institute for Silicate Research (ISC), Wuerzburg, Germany, 3Departments of Biomedical Engineering and Radiology, Columbia University, New York, NY, United States
Synopsis
Susceptibility-induced field inhomogeneities in both
space and time make B0-shimming a prerequisite for cardiac MRI at
ultra-high field. All individual terms of the static 3rd-order
spherical harmonics shim system were calibrated. Field mapping and calculation
of shim currents are performed in customized B0DETOX software. Analysis of B0-inhomogeneities
is later tested both in measurement of an ex-vivo pig heart and in-vivo in
humans. The adjustment of the shim volume to the three measured slices in a
healthy volunteer reduced the standard deviation of the field map by 4%, 19%
and 18% compared to shimming of the global heart.
Introduction
Magnetic resonance imaging of the heart at ultra-high
field (UHF) is challenging because of a variety of factors such as strong
susceptibility-induced B0-field variations occur because of the
complex structure of both the heart and surrounding tissue1,2. These
variations result in a significant signal loss and in image distortions. Thus,
their correction both spatially and temporally is an absolute prerequisite3.
For the assessment of the spatial distribution of B0 in the volume
of interest, the phase method is often used. Here, sufficient SNR at the
longest TE is important for efficient phase-based B0-mapping. To analyze
the magnetic field generated by spherical harmonics (SH) shim coils, the
Siemens MAGNETOMTM Terra 7T 3rd-order SH shim
system was first calibrated using phase-mapping pulse sequence for cardiac application.
Customized B0DETOX4 software based on the calibration matrix later
was used for field analysis in ex-vivo pig hearts as well as in several healthy
volunteers.Methods
For 3rd-order shim calibration phase maps are
acquired using a multi-slice gradient-echo pulse sequence (FOV 186×220×173 mm3,
matrix size 108×128×64, TR=30 ms, FA=25°) measuring four single-TE scans (TE1/TE2/TE3/TE4
= 1.0/1.3/2.0/4.0 ms) for each shim setting with a 2-channel Tune-Up coil. B0-maps
are calculated on pixel basis from the phase evolution over TE using linear
regression. Spatial phase-unwrapping was performed using FSL5
algorithm. The magnetic field selected by region-of-interest (ROI) (Fig. 1) is
decomposed by customized B0DETOX software into SH components taking into
account that three 3rd-order terms (X3, XYZ and Y3) are not
implemented in the 3rd-order shim electronics of the Terra system. To
calibrate an individual shim term, its value was changed in seven steps relative
to a reference B0-map using approximately 10 % of the dynamic range
for the linear gradients, and 30 % for the higher orders. The calibration
matrix A connects the shim settings s with all SH components f before shimming6, $$$s = -f·A^{-1}$$$
,
and also yields information about the cross terms. B0DETOX-based
field mapping is finally used to analyze localized scanner-side “Brain” shimming protocol
which offers the shortest TE1 and ΔTE (TE1/TE2 = 1.02/3.06 ms)7. For
both ex-vivo measurements of excised pig hearts and in-vivo measurements in
healthy volunteers, a multi-slice gradient-echo pulse sequence is used (FOV 300×300×30
mm3, matrix 128×128×3, TR=100.0ms, TE1/TE2/TE3 = 1.70/2.83/3.96 ms).
Ex-vivo measurements are performed with excised pig hearts preserved in 0.9 %
NaCl solution. After granting permission by local ethics committee and
obtaining informed consent from the volunteers, in-vivo measurements are
performed under breath-hold using ECG-triggering. The improvement by a certain
shim can e.g. be described by the quality factor QS = σR/σT
which describes the decrease of the standard
deviation σR of the resulting field compared to the initial field
map σT.Results and Discussion
The SH shim system was successfully calibrated. The
total dynamic ranges for each term are X = -2124 Hz/cm, Y = 2120 Hz/cm, Z =
2129 Hz/cm, X2-Y2 = 6.1 Hz/cm2, ZX = 12.6 Hz/cm2, Z2 =
40.2 Hz /cm2, ZY = -12.3 Hz/cm2, XY = -6.2 Hz/cm2,
Z(X2-Y2) = -0.10 Hz/cm3, Z2X = -0.35 Hz/cm3, Z3 = -0.68
Hz/cm3, Z2Y = -0.34 Hz/cm3. The predicted residual field
of the spherical phantom used for calibration shows the lacking 3rd-order
in axial slice (Fig. 2). The quality of the prediction largely matches the
experimental outcome. Analysis of short- and long-axis view images and field
maps of the heart for a human participant are shown on Figures 3 and 4. Using 3rd-order
shim calculation by B0DETOX in the heart, one has to take oblique slice
orientation and resulting rotated coordinate system relative to the scanner’s
gradient axes into account when acquiring short-axis or four-chamber view field
maps. Scanner-side “Brain” shimming was both adjusted global on the heart and
specifically on measured slices (Fig. 5). Quality factors QS for the
three measured slices were QS,1=0.96, QS,2=0.81 and QS,3=0.82,
which means an improvement in standard deviation of 4%, 19% and 18%.Conclusion
Calibration-based B0DETOX field analysis shows the
potential for static 3rd-order SH B0-shimming
optimization in both ex vivo and human measurements. However, further steps
have to be done to perform proper in vivo shimming of the heart. These steps
include dynamic shimming adjusted both on cardiac phases and slices as well as
static and dynamic multi-coil shimming. In contrast to scanner-side shimming –
which would require a very long acquisition time - B0DETOX enables calculation
of field maps measured using ECG-triggering. Calculation can be performed for
specific heart phases, which will further improve shim quality.Acknowledgements
Financial support was obtained from the German Ministry
of Education and Research (BMBF) under grant #01E1O1504.References
[1] W. Mattar, C. Juchem, M. Terekhov, L.M.
Schreiber; Multi-coil B0 shimming of the human heart: A theoretical
assessment; ISMRM Annual Meeting, Singapore (2016)
[2] M.K. Atalay, B.P. Poncelet, H.L. Kantor,
T.J. Brady, R.M. Weisskopf; Cardiac susceptibility artifacts arising from the
heart lung interface; Magnetic Resonance in Medicine 45, 341-5 (2001)
[3] P.F. Ferreira, P.D. Gatehouse, R.H.
Mohiaddin, D.N. Firmin; Cardiovascular magnetic resonance artefacts; Journal of
Cardiovascular Magnetic Resonance, 15:41(2013)
[4] C. Juchem, S.U. Rudrapatna, T.W. Nixon, R.A.
de Graaf; Dynamic multi-coil technique (DYNAMITE) shimming for echo-planar
imaging of the human brain at 7 Tesla, NeuroImage 105 (2015), 462 - 472
[5] S.M. Smith, M. Jenkinson, M.W. Woolrich,
C.F. Beckman, T.E.J. Behrens, H. Johansen-Berg, P.R. Bannister, M. De Luca, I.
Drobnjak, D.E. Flitney, R. Niazy, J. Saunders, J. Vickers, Y. Zhang, N. De
Stefano J.M. Brady, P.M. Matthews; Advances in functional and structural MR
image analysis and implementation as FSL; NeuroImage, 23(S1): 208-19 (2004)
[6] C. Juchem, T.W. Nixon, P. Diduch, D.L.
Rothman, P. Starewicz, R.A. de Graaf; Dynamic Shimming of the Human Brain at 7
T, Concepts in Magnetic Resonance Part B (Magnetic Resonance Engineering), Vol.
37B (3) 116 – 128 (2010)
[7] M. Hock, M. Terekhov, D. Lohr, A. Schröder,
H. Walles, L.M. Schreiber; B0-Mapping and Shimming Efficiency for ex
Vivo MR Imaging of the Heart at Ultra-High Field – Validation of Standard
Shimming Protocols of MAGNETOMTM Terra 7 T Scanner; DGMP and DGBMT
Annual Meeting, Dresden (2017)