Ed Boskamp1, Zhentian Xie2, Victor Taracila1, Amy Stephen2, Mike Edwards2, Tim Skloss2, Ralph Hurd3, Fraser Robb1, and Joe Murphy-Boesch4
1G. E. Healthcare Technologies, Aurora, OH, United States, 2G. E. Healthcare Technologies, Waukesha, WI, United States, 3Radiological sciences lab, Stanford University, Palo Alto, CA, United States, 4NINDS-NIH, Bethesda, MD, United States
Synopsis
Hyperpolarized 13C enhances
the SNR of signals from 13C metabolites. Separate transmit and receive coils are
inserted into the magnet bore to image 13C, limiting patient
space. Here, a dual tuned 13C
/1H body coil is developed that is capable of imaging both proton and
13C in one exam. The coil has the same 70 cm inner diameter as the standard
body coil and can be used stand-alone as the Tx/Rx coil, or as the transmit
coil for proton and 13C receive arrays. The efficiency for proton
excitation is comparable to that of the standard proton only body coil.
Introduction
Hyperpolarization of 13C
is known to enhance the SNR of the 13C resonance by orders of
magnitude1 and has been used to image prostate
cancer in humans at 3T.2 For prostate imaging, a 13C
transmit coil is positioned around the patient separate from the 13C
endorectal receive coil, limiting space for the patient.2 Here, a
dual tuned whole body coil based on the four-ring birdcage3 is described
that is capable of imaging both 1H and 13C
simultaneously. The coil has the same 70 cm diameter and proton field of view as
the standard body coil. It can be used in
Tx/Rx mode, or it can be de-tuned for use with receive-only arrays for either
nucleus. Methods
Using ½ inch
copper tape conductors, 8 inch diameter, 8¼ inch long scale models were built of (1) a
regular high-pass 16-rung proton birdcage, (2) overlapped 16 rung birdcages
(proton, HP and 13C, LP)4 with ½ inch gap between end-rings, and (3)
a dual tuned 4 ring birdcage type ( ½ inch between end-rings) first reported by
Murphy-Boesch et al.3 All coils were surrounded by a 9¼ inch
diameter RF shield. Based on the results of scale model testing, the four-ring
birdcage was chosen and developed as a full scale 70 cm whole body coil (Fig 1).
The low-pass 13C mode is developed by the inner 16-rungs birdcage; the
high-pass proton mode by the outer birdcages coupled through the inner birdcage.
The rung capacitance was 40 pF and the outer end-ring capacitance was 33 pF per
segment. 1 nF capacitance was inserted on the inner rings to prevent gradient
eddy currents. 127 MHz baluns were applied to the carbon-mode and proton-mode
drive cables to prevent E-field induced shield currents. Both modes were
matched to a 75kg patient and driven in quadrature. FEM analysis using HFSS
(Ansys,Inc) was used to calculate the fields and RF current density patterns.
The body coil was mounted in a standard 3T 70 cm MRI system with multi-nuclear
capability (Discovery MR750W, GE Healthcare, Waukesha, WI). The net forward
power to the coil was measured with directional couplers during a calibrated
pulse sequence with known B1+ magnitude derived from a 10 cm spherical Si oil
phantom for proton and an 18 cm spherical ethylene glycol phantom for 13C.
The field of view in the z-direction was measured using pick up loops. Lab
tests on volunteers using two non-resonant pick up loops showed the
relationship between patient size / shape and the Q of the coil in both modes.Results
At 127 MHz, scale
models showed that coil losses for the four-ring birdcage were 0.7 dB worse
than for the standard birdcage, but 0.9 dB better than the overlapped
birdcages. The carbon efficiency (32 MHz) in the four-ring birdcage was also
better than the overlapped birdcages by 4.5 dB. For the body coil, the low-pass modes of the
inner birdcage and the high-pass modes of the coupled outer birdcages were well
separated. FEM analysis (Figs 2 and 3)
of the body coil at 127.7 MHz indicated that 31% of the current in the outer
rings is returned by their neighboring inner rings, so ring currents, normalized
to the outer rings, distribute as 1.0/-0.31/0.31/-1.0. The -6dB Z-FOV measured in the system was 52
cm for proton and 37 cm for 13C. The unloaded B1+ efficiency
measured in the system was: 0.238 mT/sqrt(P) for 127 MHz and 0.271 mT/sqrt(P) for 32 MHz, where power P
is in Watts. The unloaded Q’s were 280 for the proton mode and 180 for the carbon
mode. The Q ratios measured in the lab using people of different weights were:
Qempty/Q75kg = 2.8 for 127 MHz and Qempty/Q75kg =1.1 for 32 MHz. Phantom images
showed both good uniformity and good SNR at 32 MHz (Fig. 4). Mode isolation was
<-32dB at both frequencies with bandstop filters.Discussion
A 70 cm dual
tuned birdcage with close coupled RF shield was developed. The gap between end-rings is narrower than in
previous work3 (16% versus 25% of coil length) providing greater FOV
for the carbon mode. The efficiency for
proton excitation is comparable to that of a regular proton body coil, and the
transmit efficiency for carbon is adequate to produce a 24 mT B1 field at the coil
center with 8 kW of RF power. Addition of bandstop filters will allow
simultaneous excitation at both frequencies.Conclusion
A dual tuned
body coil can replace custom-built transmit coils for imaging hyperpolarized 13C
without sacrificing performance of a standard body coil.Acknowledgements
No acknowledgement found.References
1. Ardenkjaer-Larsen, JH, Fridlund, B, Gram, A, et al. Increase in
signal-to-noise ration >10,000 times in liquid-state NMR. PNAS 2003;100:10158.
2. Nelson, SJ, Kurhanewicz, J, Vigneron, D, et al. Metabolic Imaging of
Patients with Prostate Cancer
Using Hyperpolarized [1-13C]Pyruvate. Sci Transl Med 2013;5(198):1.
3. Murphy-Boesch, J, Srinivasan, R. et al. Two Configurations of the
Four-Ring Birdcage Coil for 'H Imaging and 'H-decoupled 31P Spectroscopy of the Human Head. JMR
1994;B103:103.
4. Fitzsimmons, JR, Beck, BL, and Brooker, HR. Double Resonant
Quadrature Birdcage. Mag Reson Med 1993;30(1):107.