Aidin Ali Haghnejad1, Mark Gosselink1, Ingmar Voogt1, Dennis Klomp1, Peter Luijten1, and Alexander Raaijmakers1,2
1Radiology, UMC Utrecht, Netherlands, Utrecht, Netherlands, 2Eindhoven University of Technology, Biomedical Image Analysis, Eindhoven, Netherlands
Synopsis
Local
multi-transmit arrays at 3T provide reduced power requirements and reduced
local SAR. However, it requires 3T scanners with multi-transmit functionality
which are rare. This work presents add-on hardware that enables the use of
local transmit/receive arrays. An exploration on prostate imaging with fixed
phase settings using a 8-channel dipole array has been performed on four
subjects. B1+ levels range from 5 to 8.5 uT for 8 x 215-300 W input power. T2w
images have been acquired successfully for each subject. The modest
inter-subject variation in B1+ demonstrates the feasibility of this approach.
Purpose
The
birdcage body coil provides superior B1+ homogeneity for MR imaging at both 1.5
and 3T. However, the use of local multi-transmit arrays at 3T has demonstrated
considerable advantages in terms of reduced power requirements and reduced
local SAR1. This approach requires a so-called multi-transmit system where
the phases of all channels can be independently controlled. Currently only very
few 3T MRI systems have this functionality. In this work we present add-on
hardware to perform multi-transmit MR imaging at a conventional 3T system using
a local transmit/receive array. With this setup, prostate imaging performance
is explored for a 8-channel meandering dipole array using fixed phase settings.Methods
Two 4-channel power dividers have been
built using LC circuits (figure 1E). The resulting eight channels are connected
to eight custom built transmit/receive switches (figure 1C). A 8-channel
transmit/receive array1 was used consisting of dipole antennas of 30 cm
length where half of each leg of the dipole antenna was meandering with 30 cm width
and 2 meanders per cm (figure 1D). A
lattice balun ensured matching and differential drive. Local SAR distribution
was determined using numerical simulations (Sim4Life, Zurich Medtech). Maximum 10g-averaged
local SAR for 8x1 W input power was determined at 2.2 W/kg (figure 2). This
resulted in a maximum average input power of 9.1 W per channel. The two input
channels of the power dividers were connected to the two disconnected cables to
the body coil of a 3T Ingenia system (Philips Healthcare, Best, The
Netherlands). The allowed peak power arriving at the power divider for this
study was 1200/1400 W for channel 1 and 2 respectively. This resulted in 215 to
300 W power arriving at the TR-switch coil connector. The receive lines of the
TR-switches were connected to a D-stream box (MR Coils B.V., Zaltbommel, The
Netherlands) that connects to the receive interface of the scanner.
Optimal phases were obtained for one subject through addition of appropriate
cable lengths, determined by single-channel phase images. To investigate the
feasibility of one fixed cable length setup for multiple subjects, three other
subjects were scanned with the same cable lengths and B1+ levels in the prostate
were recorded. In addition, T1w and T2w images were acquired for each subject.Results and discussion
The B1+ levels acquired in the prostate
with the dipole array are indicated in table 1. On average, 6.6 uT is obtained. Figure 3 shows T1w images for all subjects.
The B1+ levels in the prostate vary and are generally low but no signal dropout
is ever observed inside the prostate for any subject. This combined with the relatively low input
powers enables the compensation of lower B1 levels by higher input power
similar to the way that input power to the body coil is optimized before every
scan. Figure 4 shows T2w images for each subject showing that good quality
images can already be obtained with this setup. However, in absence of a
dedicated receive array, the image quality should not be compared to
state-of-the-art image quality for prostate imaging at 3T.Conclusion
Dedicated add-on hardware was developed to
allow the use of local multi-transmit coil arrays at a conventional 2-channel
3T system. Using one optimized phase setting (acquired through additional cable
lengths) B1+ values in the prostate ranged from 5 to 8.5 uT for four different
subjects using 8 x 215-300 W input power. Future work will focus on exploring
2-channel RF shimming for B1+ improvement and adding in a comparison to a
transmit/receive loop coil array.Acknowledgements
This project is funded by STW project nr: 13783
References
1. A. Haghnejad, in Proceedings of the ISMRM 24th Annual Meeting 2016, #3174