Valentina Taviani1, Ann Shimakawa1, Lloyd Estkowski1, Arnaud Guidon2, Ersin Bayram3, and Robert Peters4
1Global MR Applications and Workflow, GE Healthcare, Menlo Park, CA, United States, 2Global MR Applications and Workflow, GE Healthcare, Boston, MA, United States, 3Global MR Applications and Workflow, GE Healthcare, Houston, TX, United States, 4Global MR Applications and Workflow, GE Healthcare, Waukesha, WI, United States
Synopsis
MUltiplexed Sensitivity Encoding (MUSE) has been
successfully used to correct for motion-induced phase errors in multi-shot
diffusion-weighted imaging. However, this technique relies heavily on parallel
imaging (PI) and can result in residual aliasing and excessive noise amplification when
the number of shots is similar to the number of receiver coil elements. We
propose a navigated multi-shot approach with multiplexed sensitivity encoding to
handle cases where the coil geometry would otherwise limit the maximum number
of interleaves. We show that both PI and 2D-selective excitation pulses can be
used to reduce the scan duration, while maintaining similar levels of
distortion.
Introduction
MUltiplexed
Sensitivity Encoding1 (MUSE) has been successfully used to correct
for motion-induced inter-shot phase errors in multi-shot diffusion-weighted
imaging (DWI). Despite the better conditioning of the MUSE formulation over
conventional parallel imaging (PI), this technique relies heavily on coil
sensitivity variations to correct for motion-induced phase errors, making its
performance dependent on the specific receiver coil used for signal reception.
In addition, especially for body applications, while high-density coils are
often used, it is not uncommon to have only two receiver elements in the phase
encode direction, which also limits the maximum number of interleaves that can
be reconstructed with MUSE without incurring in unwanted residual aliasing or
excessive noise amplification. We propose a navigated multi-shot Echo-Planar
Imaging (EPI) acquisition with multiplexed sensitivity encoding2 to extend
the MUSE approach to cases where the coil geometry or the specific prescription
would otherwise limit the maximum number of interleaves. All the advantages of
the non-navigated approach in terms of SNR performance and improved matrix
conditioning are maintained. In addition, depending on the application, a
combination of PI and 2D-selective excitation pulses3 are shown to
reduce the overall scan duration, while maintaining similar levels of
distortion. Methods
We implemented a twice-refocused, navigated, multi-shot,
diffusion-weighted EPI pulse sequence, with a low-resolution, fully sampled 2D navigator
acquired after each imaging readout (Figure 1a). Phase correction and ramp
sampling correction were performed separately for imaging and navigator data. Complex
coil combination of each navigator echo was performed using coil sensitivities
estimated from a separate conventional 3D gradient echo scan4. Local
unwrapping and smoothing of the navigator phase using a 17x17 boxcar filter was
performed to reduce phase inconsistencies in low-SNR regions. For each
interleave, the corresponding motion-induced navigator phase was demodulated
from the coil sensitivities, yielding Nshots x Ncoils pseudosensitivities. A
generalized SENSE reconstruction1 was performed to recover the final
phase-corrected, eventually unaliased, imaging volume (Figure 1b). For
prostate, a 2D excitation pulse was used to limit the phase encoded FOV to the
anatomy of interest. Imaging was performed both at 1.5T (brain - MR450w, GE,
Waukesha, WI) and 3T (prostate – MR750, GE, Waukesha, WI), using an 8-channel brain
coil and a 32-channel cardiac array for brain and prostate imaging,
respectively. Results
Figure
2a shows a comparison between conventional and navigated MUSE in a healthy
volunteer. The non-navigated approach results in unacceptable noise amplification
and residual aliasing when the number of interleaves is the same as the number
of available receiver coils, while both of these issues are completely resolved
when the motion-induced phase estimates are obtained from the navigator data. Scan
time scales linearly with the number of interleaves, however, distortion is
proportional to the effective phase-encoded FOV, so that the same or even lower
levels of distortion can be achieved in significantly less time using PI to
synthetize missing samples. As an example, increasing the number of interleaves
from 2 to 4 reduces distortion by a factor of 2, while doubling the scan time. Further
undersampling a 4-shot acquisition by a factor of 2 results in distortion
levels comparable to an 8-shot acquisition, with no additional scan time
penalty (Figure 2b). Figure 3 shows two representative slices of a high-resolution
brain dataset. In this case, leveraging PI in combination with multi-shot to
speed up the acquisition while reducing distortion comes at the cost of visibly
reduced SNR. Depending on the specific application, reduced FOV (r-FOV) imaging
together with multi-shot can be used instead of PI to effectively reduce
distortion without incurring in prohibitively long scan times. Figure 4 shows significant
distortion reduction in the prostate of a healthy volunteer using r-FOV 2-shot
navigated MUSE compared to conventional r-FOV single-shot. Note that, in this
case, because of the axial prescription, there are only two receiver coil elements
in the phase encoding direction (anterior/posterior), however, no residual
aliasing and adequate SNR were obtained with the proposed navigated approach. Discussion
High
resolution DWI suffers from severe distortion and blurring due to the
underlying EPI readout. The use of 2D navigators with MUSE effectively
decouples the distortion-correction capabilities of this technique from the PI
performance of the specific coil used, thus extending its applicability to
cases with limited coil sensitivity variations in the phase encoding direction.
Nonetheless, PI and, in some cases, r-FOV imaging can be used together with
navigated MUSE to reduce the overall scan duration, while maintaining similar
levels of distortion. Conclusion
We
showed that 2D navigators, PI and r-FOV imaging used in conjunction with MUSE
allow high resolution imaging with minimal distortion and clinically acceptable
scan times. Acknowledgements
No acknowledgement found.References
1. Chen
N, Guidon A, Chang HC, et al. A robust multi-shot scan strategy for
high-resolution diffusion weighted MRI enabled by multiplexed
sensitivity-encoding (MUSE). Neuroimage 2013; 72:41-47.
2. Sundman MH, Chang HC, Xu D, et al. Enhancing
diffusion weighted image (DWI) quality with Navigator-MUSE. In Proceedings of
the 23rd Meeting of the ISMRM. Toronto, Canada. 2015, p.2802.
3. Saritas EU, Cunningham CH, Lee JH, et al. DWI of the spinal cord with reduced FOV single-shot EPI. Magn Reson Med.
2008; 60(2):468-473.
4. Roemer PB, Edelstein WA, Hayes CE, et al. The NMR phased array. Mag Res
Med. 1990; 16(2):192-225.