In this work we demonstrate the use of two dedicated anisotropic diffusion fibre phantoms for the study of free-water elimination DTI. In particular, we make use of the recently proposed approach in which an extra dimension to the diffusion weighting, namely transverse relaxation weighting, is added to the model.
Model. The model assumes that the diffusion MRI signal originates from two compartments with different transverse relaxation times and diffusivities1,3,5 $$S\left( T_\mathrm{E},b ,\mathbf{g}\right)=S_0\left[ f_\mathrm{w}e^{-\frac{T_\mathrm{E}}{T_{2,0}}}e^{-bD_0}+\left(1-f_\mathrm{w}\right)e^{-\frac{T_\mathrm{E}}{T_{2,\mathrm{t}}}}e^{-b\mathbf{g}^\mathrm{T}\mathbf{D}\mathbf{g}}\right]\mathrm{, (1)}$$ where S0 is the proton density, fw, T2,0 and D0 are the fraction, transverse relaxation time and diffusion coefficient of the free-water compartment, respectively. T2,t and D denote the transverse relaxation time and diffusion tensor for the tissue compartment. The experimental parameters are the strength and direction of the diffusion weighting gradient, b and g, and the echo-time, TE.
Phantoms. Both phantoms were constructed using polyethylene fibres (~8 μm in radius) wound on Perspex platforms, and submerged in a container with distilled water (Fig. 1).6,7
Experiments and data analysis. Measurements were performed on a 3T Siemens Prisma scanner (Siemens, Germany). A diffusion-weighted double-echo sequence with bipolar pulsed-field gradients was used. Experimental parameters included (Fig. 2): for the T2-dimension, 10 measurements with TE in the range TE $$$\in$$$ [70,400] ms, b-value = 0 s/mm2, 5 repetitions per TE. For the diffusion-dimension, b-values = 0 (8 repetitions), 400 (21 directions) and 1000 (35 directions) s/mm2; 5 repetitions. Other parameters were voxel-size = 23 mm3; matrix-size = 96×96×50; GRAPPA acceleration factor = 2.
FWET2 parameter estimation was performed in three steps with in-house Matlab scripts:
Fig. 2 shows the signal attenuation in the fibre area for 70 ms ≤ TE ≤ 400 ms and b = 0 s/mm2 (volumes 0-50) diffusion-attenuated signal (volumes 51-370) for b = 0, 400, 1000 s/mm2 (blue line). The red line depicts the fits of Eq. (1) to the experimental data.
Fig. 3 summarises the results for Phantom 1. Mean diffusivity (MD) and fractional anisotropy (FA) from conventional DTI were evaluated for comparison (Fig. 3a-b). MD, FA and fw from FWET2 are shown in Fig. 3c-e. It was found that fw decreases from left to right, according to the increase in FD which is observed in the profiles shown in Figs. 3f-g.
Fig. 4 summarises the results for Phantom 2. MD and FA from conventional DTI are shown in Fig. 3a-b for comparison. MD, FA and fw from FWET2 are shown in Fig. 3c-e. The thinnest fibre bundle (leftmost) appears blurred in DTI metrics, whereas it is fully recovered in FWET2 maps. This is shown in the profiles in Figs. 3f-h, taken along the ROI depicted by the red line in Fig. 4e. Grey zones represent voxels affected by PVE. It is clearly shown that FWET2 eliminates the bias observed in conventional DTI metrics.
[1] Pasternak O, et al. Free Water Elimination and Mapping from Diffusion MRI. Magn Reson Med 2009;62:717-730.
[2] Pasternak O, et al. Estimation of Extracellular Volume from Regularized Multi-Shell Diffusion MRI. Med Image Comput Comput Assist Interv 2012;15:305-312.
[3] Hoy AR, et al. Optimization of a free water elimination two-compartment model for diffusion tensor imaging. Neuroimage 2014;103:323-333.
[4] Neto Henriques R, et al. Exploring the potentials and limitations of improved free-water elimination DTI techniques. Proc Intl Soc Magn Reson Med 2017;1787.
[5] Collier Q, et al. Solving the free water elimination estimation problem by incorporating T2 relaxation properties. Proc Intl Soc Magn Reson Med 2017;1783.
[6] Farrher E, et al. Novel multisection design of anisotropic diffusion phantoms. Magn Reson Imaging 2012;30:518-526.
[7] Grinberg F, et al. Complex patterns of non-Gaussian diffusion in artificial anisotropic tissue models. Microporous Mesoporous Mater 2013;178:44-47.