Je-Seok Ham1, Sang-In Bae1, Won-Joon Do1, Ki-Hun Jeong1, and Sung-Hong Park1
1Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, Daejeon, Republic of Korea
Synopsis
During brain surgery, location
of lesions can change in real-time due to leakage of cerebrospinal fluid.
Therefore, navigating an MR-Endoscope probe with real-time intraoperative MRI is
important in clinical application. However, conventional tracking system
attached to the endoscope probe induces severe artifacts and is expensive and
bulky. In this study, we propose a technique for navigating the endoscope probe
without additional tracking system through segmentation of signals from tunable
lens filled with gadolinium contrast agents. We also demonstrated tunable
liquid-filled lens endoscope for intraoperative MRI. The proposed system/approach
would be a good alternative as a tracking system for intraoperative MRI.
Introduction
In order to verify the
position and orientation of MR-Endoscope in human brain during intra-operative MRI,
it is crucial to navigate the MR-Endoscope. However, the conventional tracking
system during operation is required to additional tracking device, which is
expensive bulky, and when the detectable sensor of the device is scanned in
MRI, it could induce image artifact. Also, tunable focus lens endoscope can
vary the field-of-view (FOV) by changing the focal length and minimizing
mechanical movement. Tunable focus lens endoscopes are classified into
liquid-crystal (LC) type, electrowetting on dielectric (EWOD) type, and
liquid-filled type. Liquid-filled type is the most suitable type for
intraoperative MRI because it can be miniaturized by applying pneumatic
pressure and does not induce metal artifact with no electronic device. In this
work, we propose (i) MR-Endoscope tunable liquid-filled lens, which could be
controlled by changing the pressure as injected gadolinium (Fig. 1). Also, we
propose (ii) The technique navigating the MR-Endoscope lens, which is filled
with gadolinium contrast agents using MRI segmentation by thresholds and no
additional tracking device is needed (Fig. 4). In particular, for ultrafast MRI
imaging, balanced steady-state free precession (bSSFP) sequence is used since
it enables to have higher signal-to-noise ratio (SNR) per unit time, and reduce
the data acquisition time effectively, and get better T2/T1 contrast.Methods
All experiments were
performed on a 3 Tesla animal scanner (MR solutions, Guildford, UK). The 3D
bSSFP sequence was used with scan parameters of FOV= 60mm, matrix= 128x128x32,
flip angle=30, TR/TE= 2.6/5.4ms. To optimize the concentration of gadolinium
(Meglumine Gadoterate, Dotarem), phantoms with various concentrations of 1%,
2%, 3%, 4%, and 8% were generated and tested for MR imaging. To compare
differences in signal intensity, we set a region of interesting (ROI) and
calculated the signal-to-noise ratio (SNR). The gadolinium solution of the optimal
concentration was injected into lens designed to be tuned by adjusting the
pressure of the gadolinium solution. A homemade liquid-filled based lens was
attached inside a tube, fixed in a water phantom. Lens consisted of 4
liquid-filled based channels with diameter of 4.0mm. As the pressure of the injected
gadolinum solution increased (0, 3, 6.89, 13.79, 20.68, 27.58, 34.47, 41.37
kPa), it was observed that curvature of the lens relative to the focal point varied
during MRI scan. To reduce banding artifact, we acquired bSSFP images at multiple
phase cycling angles of 0°, 90°, 180°, and 270°. Navigation for MR-Endoscope were performed by
segmentation using Matlab (The MathWorks, USA). MRI segmentations were conducted
by intensity-based thresholding that demonstrated the tube which was filled
with the gadolinium solution. Images were acquired along the axial, sagittal,
and coronal directions and the axial plane with the highest number of pixels
above the threshold level for analysis. Also, we found the center of the gadolinium-filled
tube at the axial plane was also the middle of the coronal and sagittal planes.
Since the gadolinium tube was inserted into water phantom straightly, the
orientation was calculated using position vector.Results
SNR of the segmented
lens was maximized at 4% concentration, therefore, 4% gadolinium solution was used
to fill the tunable lens (Fig. 2). The curvature of the tunable lens was
logarithmically decreased with the pressure (Fig. 3-(d)). The information of
position and orientation of the lens could be displayed simultaneously after segmentation
and calculation with Matlab (Fig. 4). As shown in Fig. 4, the
orientation of the tube could be calculated as a unit vector in a similar way
to the conventional tracking devices.Discussion
The focal length of the tunable liquid-filled lens, which was filled
with the optimized concentration of gadolinium (4% in this study), was adjustable
by changing the pressure of the solution using a pneumatic pressure device. It
can enable us to view the wide tuning range and did not need a separate mechanical
adjustment through our homemade tunable liquid-filled lens. Upon connection of
the MR-Endoscope fiber with CCD camera and PC, it would be possible to
visualize the lesion in human brain during surgery. Since navigating the
tunable lens filled with MR contrast agent is also possible, it would be also possible
to navigate the MR-Endoscope during brain surgery in real-time MR imaging.Conclusion
The described technique
was designed to navigate the tunable liquid-filled lens filled with gadolinium for
MR-Endoscopy during human brain surgery using bSSFP. It does not require any
other expensive bulky tracking device. Also, our navigating system using
gadolinium segmentation would be suitable for clinical intraoperative MRI for
minimally excised surgery.Acknowledgements
No acknowledgement found.References
1.
D Beyersdorff, et al. MR imaging-guided prostate
biopsy with a closed MR unit at 1.5T: initial results. Radiology, 2005;
234:576-581
2.
Jeong, K-H, et al. Tunable microdoublet lens
array. Optics Express, 2004; 2494-2500
3. A de Oliveira, et al. Automatic passive tracking
of an endorectal prostate biopsy device using phase-only cross-correlation.
MRM, 2008;59:1043-1050
4.
GS Fischer, et al. Pneumatically operated
MRI-compatible needle placement robot for prostate interventions. IEEE, 2008;
doi: 10.1109/ROBOT.2008.4543587.