Emilee S. Minalga1, Robb Merrill1, Dennis L Parker1, Allison H Payne1, and J. Rock Hadley1
1Utah Center for Advanced Imaging Research, University of Utah, Salt Lake City, UT, United States
Synopsis
Hardware requirements can be a roadblock to implementing
procedure-specific coils in magnetic resonance guided focused ultrasound. In order to more effectively implement coils in the system, the effects of the focused ultrasound transducer’s
position on SNR needs to be considered. This work characterizes the SNR
and noise correlation variability of the RF coils by evaluating the SNR
tradeoffs and noise correlation as a function of device orientation and
transducer position and report such variances. Understanding the SNR tradeoffs of system
placement during treatment can aid in increased SNR within the treatment volume
and can be a factor to consider in treatment planning.
Introduction
Due
to the hardware requirements of magnetic resonance guided focused ultrasound
(MRgFUS), procedure-specific coils are not often used. As RF coil integration
advances in this field, effects of the focused ultrasound transducer’s position
on SNR need to be considered. The Muse MR guided focused ultrasound (FUS)
treatment device and associated radio frequency (RF) coils (1-7) is a
breast-specific design that allows for rotation of the entire device and variable
transducer positioning to target tumors throughout the breast. The system
adjustability results in potential variations in SNR and noise correlation due
to reduced signal sensitivity and shielding of the RF coils by the solid
transducer ground plane (8).
This
work characterizes the SNR and noise correlation variability of the RF coils by
evaluating the SNR tradeoffs and noise correlation as a function of device
orientation and transducer position. Accurate characterization of
device-position dependent SNR variability could be potentially used during
patient treatment planning. Such information could be used to adjust pulse
sequence parameters to improve treatment plans and monitoring techniques to
compensate for any expected SNR reduction.
Methods
A
cross-sectional view of the Muse MRgFUS device is shown in Fig.1A. The transducer
is 256-element phased array with a 14.5x8 cm aperture size and conductive
ground plane (Imasonics, Inc., Bensacon, France). The 5 degree-of-freedom mechanical
steering and volumetric electrical steering results in 1001 cm3 a
treatment volume. The treatment cylinder
(14cm diameter, 11cm height) integrates an 8-channel RF imaging coil array. A
six channel ladder geometry (9,10) phased array (11) is wrapped around the cylinder
(elements 5cm wide x 9cm high), a 17 cm diameter chest loop coil encircles
the top of the cylinder, and a 10cm x 15cm loop surrounds the transducer opening.
To test for variations in RF coil performance with device
position and transducer orientation, coronal SNR scans (GRE: TR/TE 500/10ms, flip angle=90°,
FOV=256x256mm, Resolution=1x1x5mm, 5 Averages) were taken
of a homogeneous Cu2SO4 cylindrical phantom at 8
different transducer positions. The transducer rotation, slider position and tilt
values are summarized in Table 1. SNR and noise correlation were calculated for
each treatment configuration using methods published previously. SNR
measurements were obtained from seventeen ROIs distributed throughout the
phantom (see Fig.2 and Fig.3). One ROI was at the center of the treatment tank
and 2 sets of eight were placed radially at a distance of 40 and 55 mm from the
treatment cylinder center.Results
SNR images at the 8 orientations are shown in Fig.3. The largest variation in SNR occurs with rotation angle, with the largest variance occurs towards the edge of the cylinder. When the system was rotated the average percent change of all the ROIs was 53.5%(compare Fig.3A and Fig.3E). The SNR at the center of the phantom varies
from 372 (Fig.3E) to 692 (Fig.3G). The largest variation is seen at left edge
of the tank with a minimum of 330 (Fig.3C) to a maximum of 2148 (Fig.3E). The slider position results in a large change in SNR at the 90°
rotation. The smallest variation in SNR results from the transducer tilt; seen when comparing SNR maps Fig.3C to Fig.3D and Fig.3G to Fig.3H in Figure 3 where the average
percent change of all the ROIs was 5.2% (compare Fig.3G and Fig.3H). The correlation matrices obtained from each
treatment configuration are shown in Fig.4.
The highest value for each configuration is indicated in the figure with
the max value in any configuration varying from 0.31 to 0.46. Discussion
The Muse MRgFUS
system has been engineered to allow treatment from many transducer orientations
relative to the breast. Although the SNR variation with rotation is large at
the periphery, it is reduced in the center of the treatment cylinder. Therefore
SNR variability due to device positioning will likely not impact treatment of
centrally located tumors. Conversely, considerations of SNR variability will
likely need to be considered in treating tumors closer to the skin. The noise
correlation does demonstrate a small change mainly a function of the transducer
slide position, but the magnitude of this effect is small enough that it should
not affect the overall SNR. Also, as shown in a previous
study (8), segmenting the ground plane during the transducer design
process would further allow for more stable SNR images as a function of
transducer position with the tank.Conclusion
Variations in SNR can
impact all aspects of an MRgFUS treatment including planning, temperature
monitoring and assessment. Characterization of the SNR profiles in different
treatment configurations will allow the use of SNR information to be
incorporated into the patient treatment planning process informing treatment configurations
selection that provide the best SNR for a given treatment volume. Acknowledgements
Funding received from grant, NIH RO1 CA172787.References
1. Minalga E, Merrill R, Parker DL, de
Bever J, Hadley JR, Payne A. Evaluation of a MRgFUS breast system at 1.5 and 3
Tesla.; 2015 April 17, 2015; ISTU; Utrecht, The Netherlands. p Abstract:
2162666
2. Minalga
E, Merrill R, Parker DL, Payne A, Hadley JR. Comparison of improved breast
magnetic resonance guided focused ultrasound system with improved radio
frequency phased array coils. ISMRM. Toronto, Ontario, Canada
; 2015.
3. Minalga
E, Merrill R, Todd N, Parker DL, Hadley JR, Bangerter NK. A 10-channel RF coil
for use in magnetic resonance guided high intensity focused ultrasound of the
brain. 2013; ISMRM; Salt Lake City, Utah. p 0832.
4. Minalga
E, Merrill RP, Todd N, Payne A, Parker DL, Hadley JR. A 6-channel brain coil
for MR guided high intensity ultrasound. 2012; ISMRM; Melbourne, Australia.
5. Minalga
ES, Payne A, Merrill RP, Todd N, Vijayakumar S, Parker DL, Hadley JR. Design
and Evaluation of RF Coils for Magnetic Resonance Guided High Intensity Focused
Ultrasound. . 2011; ISMRM; Montreal, Canada. p 1726.
6. Payne
A, Merrill R, Minalga E, Vyas U, de Bever J, Todd N, Hadley R, Dumont E,
Neumayer L, Christensen D, Roemer R, Parker D. Design and characterization of a
laterally mounted phased-array transducer breast-specific MRgHIFU device with
integrated 11-channel receiver array. Medical physics 2012;39(3):1552-1560.
7. Payne
A, Minalga E, Todd N, de Bever J, Dumont E, Neumayer L, Christensen DA, Roemer
RB, Parker DL. Initial in vivo evaluation of a breast-specific MRgHIFU system,
. 2012; ISMRM; Melbourne, Australia.
8. Minalga
E, Payne A, Merrill R, Parker DL, Hadley JR. Effects of ultrasound transducer
ground plane configuration on SNR in MR guided focused ultrasound therapies
2017 April 22 - 27, 2017; ISMRM; Hololulu Hawaii. p #783.
9. Jevtic
J. Ladder networks for capacitive decoupling in phased-array coils. 2001 21-27
April; ISMRM-ESMRMB Joint Annual Meeting; Glasgow, Scotland, UK. p 17.
10. Wang J.
A novel method to reduce the signal coupling of surface coils for MRI. 1996 27
April - 3 May; ISMRM; New York, New York. p 1434.
11. Roemer
PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array.
Magnetic resonance in medicine : official journal of the Society of Magnetic
Resonance in Medicine / Society of Magnetic Resonance in Medicine
1990;16(2):192-225.
12. Kellman
P, McVeigh ER. Image reconstruction in SNR units: a general method for SNR
measurement. Magn Reson Med 2005;54(6):1439-1447.