Pallab K Bhattacharyya1,2, Tanvir Baig3, Bhumi Bhusal3, Mark J Lowe1, Michael Martens3, and Stephen Jones1
1Imaging Institute, Cleveland Clnic, Cleveland, OH, United States, 2Radiology, Cleveland Clinic Lerner College of Medicine, Cleveland, OH, United States, 3Physics, Case Western Reserve University, Cleveland, OH, United States
Synopsis
RF-induced heating of stereo encephalography (SEEG)
electrodes during MRI scans could be of concern. The change in heating pattern
of an electrode in the presence of another electrode was investigated by
measuring the heating at the tip of a conducting and insulated (bare at tip) wire
parallel to B0-field and entering a poly-acrylic gel phantom along
left-right axis in the presence of another wire. While the resonance length for
maximum heating of the wires did not depend on the number of wires, the
temperature rise at the wire tips depended on the relative lengths (resonance /
anti-resonance) of the wires.
Introduction
Simultaneous functional MRI (fMRI) and stereo
electroencephalography (SEEG) improve identification of epileptogenic zone (EZ),
for subsequent resection as a cure for medically refractory focal epilepsy.1 RF induced heating
of SEEG electrodes during MRI scans in dielectric media similar to cerebral
tissues have been investigated.2,3 Certain
resonating lengths can cause excessive heating at the tip of a guidewire.4,5 In order to study heating during MRI with
multielectrode SEEG, it is pertinent to understand the effect of coupling of
more than 1 electrode. In this study we explored the heating at the tip of a
conducting and insulated wire (bare at tip) parallel to B0 field and
entering a poly-acrylic gel phantom along left-right (X) axis in the presence
of another wire. Methods
Experiments were performed in a
whole body 3T Prisma scanner (Siemens Erlangen, Germany) with a transmit-receive
head coil. Heating of single as well as 2 copper wires (each 0.7 mm diameter)
during a turbo spin echo (TSE) scan (TR: 6470 ms; TE: 71 ms; Flip angle: 1800;
Turbo factor: 15; Echo trains per slice: 18) were studied. The specific
absorption rate of the sequence generating 8.3 W of time averaged RF power were
whole body: 0.2 W/kg, exposed body: 2.8 W/kg, head: 2.8 W.kg. The wires had
polyvinyl chloride insulation throughout the length, except for 2 mm at the tip
(as in the tip of a real electrode). An ASTM torso phantom was filled up with
polyacryclic gel6 having similar
conductivity and dielectric constant as human tissue. Experiments were carried
out with (i) a single wire entering the phantom along the X-axis with the wire
bent and entering from the side of the phantom with 9 cm of its length in the
phantom, and (ii) a 2nd wire entering the phantom similarly but at 1 cm above
the 1st wire (Fig. 1). The lengths of the wires were changed by
cutting the portion of the outside end while keeping the inside portion lengths
fixed at 9 cm (similar to where midline would be in in vivo scan). Temperature
changes were measured using fluoroptic temperature sensor (model m3300, Luxtron
(Lumasense Technologies), Santa Clara, CA, USA). The system was cooled down to
baseline temperature by allowing sufficient time between subsequent measurements
For each configuration, the induced RF fields were calculated using finite-difference
time-domain (FDTD) method7 based software
(XFdtd 7.4; Remcom Inc., State College, PA, USA), followed by thermal
simulations to calculate the temperature rise using the same software. The RF
transmit coil was modeled as 16-rung high-pass quadrature birdcage coil with 17.5
cm radius, 33 cm length and tuned to 127.6 MHz.Results and Discussion
The temperature rise (from
experiment and simulation) with a single wire of different lengths is shown in Fig.
2. The 1st two resonances in both cases, were at total lengths of 70
cm and 185 cm (separated by half the wavelength of electromagnetic wave in air
at 3T). The experimental and simulation-generated temperature rise of the wire
tips during the TSE scan as a function of lengths for 2 wires entering the
phantom laterally are shown in Fig. 3 and Fig. 4 respectively. The TSE
sequence, not used in vivo, is used
here to generate high power (~4-5 times higher than echoplanar imaging for fMRI).
The results show that presence of the 2nd wire causes less
temperature rise than that with a single wire. While the 1st
resonance length of the 1st wire remained the same (~70 cm), that of
the 2nd (upper) wire was ~10 cm higher. However, the temperature
rise in the lower wire is lower in the experiment, while it is higher is
simulation. Also, with 70 cm (resonance length) lower wire and 120 cm (close to
anti-resonance length) upper wire combination, the temperature rise of the 2
wires were 11.2oC and 5.8oC respectively. While the
simulated and experimental resonance lengths agree with each other, temperature
rises differ between experiment and simulation, which may be because of some differences
in the properties of the coils (exact geometry etc.) and the gel (dielectric property etc.) used in
the 2 cases. This study suggests that coupling of E-fields close to each
electrode effectively results in less temperature rise when the lengths of both
wires are same. However, combination of resonance and anti-resonance lengths
can lead to higher temperature rise in the wire at resonance length.Conclusion
RF-induced heating of 2 wires parallel to
B0 and entering dielectric media along X-axis, depend on the
relative lengths (resonance / anti-resonance) of the 2 wires.Acknowledgements
Cleveland Clinic Epilepsy Center.References
1. Gonzalez-Martinez J,
Bulacio J, Alexopoulos A, Jehi L, Bingaman W, Najm I.
Stereoelectroencephalography in the "difficult to localize"
refractory focal epilepsy: Early experience from a North American Epilepsy
Center. Epilepsia.
2. Bhattacharyya PK, Mullin J, Lee BS, Gonzalez-Martinez JA,
Jones SE. Safety of externally stimulated intracranial electrodes during
functional MRI at 1.5T. Magn Reson
Imaging. 2017;38:182-188.
3. Carmichael DW, Thornton JS, Rodionov R, et al. Feasibility
of simultaneous intracranial EEG-fMRI in humans: a safety study. Neuroimage. 2010;49(1):379-390.
4. Armenean C, Perrin E, Armenean M, Beuf O, Pilleul F,
Saint-Jalmes H. RF-induced temperature elevation along metallic wires in
clinical magnetic resonance imaging: influence of diameter and length. Magn Reson Med. 2004;52(5):1200-1206.
5. Yeung CJ, Karmarkar P, McVeigh ER. Minimizing RF heating of
conducting wires in MRI. Magn Reson Med. 2007;58(5):1028-1034.
6. Baker KB, Tkach JA, Phillips MD, Rezai AR. Variability in
RF-induced heating of a deep brain stimulation implant across MR systems. J Magn Reson Imaging. 2006;24(6):1236-1242.
7. Taflove A, Hagness SC. Computational
electrodynamics: the finite-difference time-domain method. Norwood, MA:
Artech Hhouse; 2005.