Lauren Watkins1, Feliks Kogan2, Marianne Black3, Marc Levenston1,2,3, and Garry Gold1,2
1Bioengineering, Stanford University, Stanford, CA, United States, 2Radiology, Stanford University, Stanford, CA, United States, 3Mechanical Engineering, Stanford University, Stanford, CA, United States
Synopsis
GagCEST is a quantitative MR technique that
shows promise at 7T to specifically detect cartilage glycosaminoglycan content;
however, its potential at 3T is still uncertain. This study utilizes a new
optimized 3D GagCEST sequence to maximize SNR and GagCEST contrast at 3T.
Comparison of GagCEST asymmetry maps obtained at 3T and 7T suggest that GagCEST can be used to distinguish
zonal differences in cartilage composition at both 3T and 7T. This work
demonstrates potential for whole joint GagCEST knee imaging at 3T with improved
dynamic range.
Introduction
Osteoarthritis is a progressive and debilitating
disease whose early stages are marked by cartilage degradation1. At
this phase, enzymatic degradation of cartilage results in glycosaminoglycan
(GAG) depletion. While many quantitative MR imaging techniques have been
developed to detect GAG content in articular cartilage, their application has
been limited by long scan times, the need for specialized hardware, or lack of
specificity to GAG1. Chemical exchange saturation transfer of GAG (GagCEST)
is a recently-developed quantitative MR technique that uses chemical exchange
of specifically saturated exchangeable protons on the hydroxyl groups of GAG molecules
and bulk water protons to provide contrast and monitor cartilage GAG content in vivo2. GagCEST has been
validated against sodium MR at 7T2,3; however, its potential at 3T
is still uncertain. While previous studies have shown that the dynamic range of
GagCEST asymmetry in healthy volunteers is low at 3T4, studies that
assess its specificity to GAG distribution are limited. In this work, we
utilize a new optimized 3D GagCEST sequence to maximize SNR and GagCEST
contrast and assess the relationship between GagCEST effects at 3T and 7T using
immature bovine articular cartilage. Methods
Four intact, immature bovine stifles
(San Jose Valley Veal, Santa Clara, CA) were imaged at room temperature on 3T and
7T MR scanners (GE Healthcare) using a 16-channel and 32-channel coil,
respectively. GagCEST images were acquired using a magnetization prepared 3D spoiled
gradient-echo sequence (Figure 1) with imaging parameters shown in Table 1. CEST
saturation parameters were optimized for bovine cartilage at each respective
field strength. CEST analysis was performed using in-house MATLAB scripts to
correct for B0 and B1 field inhomogeneities as described previously5.
CEST asymmetry due to GAG was calculated using the normalized B0
corrected signal intensity at ±1.0 ppm, the chemical shift of GAG hydroxyl
protons, using the equation: $$ (S-1.0ppm - S+1.0ppm)/S0 $$. For each
bovine specimen, femoral cartilage was divided into two layers representing
either superficial or deep zones. These were then subdivided into 19-23 2x4x3mm
regions from equivalent regions on 3T and 7T GagCESTasym maps. An average GagCESTasym was
calculated for each masked region on a slice-by-slice basis. A Wilcoxon signed
rank test (α = 0.05) was used to compare differences in the average GagCESTasym in the
surface and deep zones at different field strengths. Further, the correlation
between GagCESTasym measurements at 3T and 7T were measured
using a Pearson’s correlation coefficient.
Results
Zonal differences in GagCESTasym were observed at
both 3T and 7T (Figure 2). There was a significant difference between the average
GagCESTasym between
the superficial and deep zones at both 3T and 7T (p < 0.01 for both). While
the average cartilage GagCESTasym was
greater for 7T than for 3T, the results were positively correlated (R = 0.431, p < 0.01). When
the zones were compared separately, there was a similar correlation between GagCESTasym at 3T and at 7T for
cartilage within the superficial zone (R = 0.409, p < 0.01) and
within the deep zone (R = 0.458, p < 0.01)
(Figure 3). Discussion
While GagCEST has shown to be promising for
direct detection of GAG at 7T, its utility at 3T remains in question. In
healthy bovine cartilage, there was a clear laminar distribution in the GagCESTasym maps at 7T. This
pattern was less distinct, though still significant, at 3T. These results are
consistent with the known GAG distribution in articular cartilage6. Despite
the lower dynamic range and reduced SNR in comparison to 7T, the 3T images
acquired here demonstrate improved dynamic range compared to previous literature4.
These results suggest that GagCEST can be used at either field strength to
distinguish zonal differences in cartilage composition. It is important to note
that we studied bovine samples at room temperature, which likely resulted in
different exchange conditions compared to in
vivo human knees. Additionally, the images were not automatically segmented
or registered, which could lead to some discrepancies in the correlations. Further
work will assess if this relationship is present within healthy human
cartilage, and if it can be used to observe early cartilage degeneration in
subjects with osteoarthritis. Conclusion
This work
demonstrates potential for whole-joint GagCEST knee imaging at 3T with improved
dynamic range. Results suggest that within juvenile femoral articular
cartilage, GagCEST can be used to distinguish zonal differences in cartilage
composition at both 3T and 7T.Acknowledgements
This work was funded by the William K. Bowes Jr.
Stanford Graduate Fellowship, GE Healthcare, and the National Institutes of
Health (NIH) grants K99EB022634,
RO1 EB002524, and K24
AR062068.References
1. Matzat S, van Tiel J, Gold G, Oei EHG.
Quantitative MRI Techniques of Cartilage Composition. Quantitative Imaging in
Medicine and Surgery, 2013; 3(3):162-174.
2. Ling W, Regatte RR, Navon G, Jerschow
A. Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer
(gagCEST). Proceedings of the National Academy of Science,
2008;105(7):2266-2270.
3. Schmitt B, Zbyn S, Stelzeneder D,
Jellus V, Paul D, Lauer L, Bachert P, Trattnig S. Cartilage quality assessment
by using glycosaminoglycan chemical exchange saturation transfer and (23)Na MR
imaging at 7T. Radiology, 2011; 260(1):257–264.
4. Singh A, Haris M, Cai K, Kassey VB,
Kogan F, Reddy D, Hariharan H, Reddy R. Chemical exchange saturation transfer
magnetic resonance imaging of human knee cartilage at 3 T and 7T. Magnetic
Resonance in Medicine 2012; 68(2):588–594.
5. Kogan F, Hargreaves B, Gold G.
Volumetric Multislice GagCEST Imaging of Articular Carilage: Optimization and
Comparison with T1rho. Magnetic Resonance in Medicine, 2017;77(3):1134-1141.
6. Maroudas A. Transport of solutes
through cartilage: permeability to large molecules. Journal of Anatomy, 1976;
122:335–347.
7. Krishnamoorthy,
G., Nanga, R. P. R., Bagga, P., Hariharan, H. and Reddy, R. High quality
three-dimensional gagCEST imaging of in vivo human knee cartilage at 7 Tesla. Magnetic
Resonance in Medicine, 2017;77: 1866–1873.