Absolute quantification is a challenge with many paths to reach the final goal of quantifying metabolites in absolute units (e.g. Molarity and molality). Utilizing an external reference standard (ERF) is an attractive method for quantifying in vivo metabolites due to the ability for direct comparison between a known concentration of a metabolite and the in vivo data. A major concern in utilization of an ERF is the differences in coil loading between in vivo and in vitro measurements. To that end, this work describes a method to calibrate and adjust the transmitter voltage in order to maximize signal detection independently of coil load.
B1-mapping was utilized to observe the tip angles achieved in a localized 2x2x2cm3 voxel in-vitro at both 3T and 9.4T. A transmitter voltage correction was applied based on the tip angle achieved such that:
Vcorrected = (θtarget/θmeasured) x Vnominal.
Where Vnominal was selected as 250V at 9.4T and 220V at 3T for all scans and was selected as 90° for 9.4T and 8° at 3T. A STEAM sequence was utilized at 3T (TE/TM/TR = 20/10/4000 ms) and 9.4T (TE/TM/TR = 11/50/5000 ms) without and with water suppression for analysis of the water and NAA signals respectively. A comparison of B1-mapping sequences (TurboFLASH and AFI6) was performed to compare the accuracy of the measured flip angle for each B1 mapping sequence as input to calculate Vcorrected.
The designed coil loading scheme (Fig. 1) for testing the reliability of the method involved using two phantoms in similar manners at both field strengths. A spectroscopy phantom was placed securely in the head coil and was used to measure in the isocenter of the phantom while a large tube phantom (Fig. 1) was used to load the coil incrementally. The tube phantom was placed immediately at the foot of the spectroscopy phantom and incrementally adjusted away from the spectroscopy phantom, in the foot-direction, by 1cm. At 3T a 64-channel receive-only birdcage head coil was used for signal detection and the body coil was used as the transmitter. At 9.4T a homebuilt T/R phased-array head coil was used as transmitter and receiver7. At each loading position Siemens 2nd-order shimming was applied for both field strengths. MATLAB was used for reconstruction of data and statistical tests were performed using R (3.4.2).
Based on the water signal detected following the B1 mapping voltage adjustment, TurboFLASH yielded the more accurate flip angle optimization in comparison to AFI-mapping. This effect is seen in Fig. 2 where the water signal increased for TurboFLASH when comparing Vnominal and Vadjusted; whereas, the signal decreased when using AFI for calculation of Vadjusted. Adjustment of the transmitter voltage based on TurboFlash B1+ mapping allowed for increased signal detection at both 3T and 9.4T across a variety of coil loads. The voltage correction method applied with TurboFLASH increased the water signal by an average percent change of 17.5%±4.8% at 9.4T (Fig. 3) and 9.9%±0.1% at 3T (Fig. 4).
The range of the applied correction factor is relatively small; with the applied voltages ranging from 197.7-199.2V at 3T and 185.4-205.6V at 9.4T when loaded with both the spectroscopy phantom and tube phantom. NAA/water is more stable when TurboFLASH voltage correction is applied (Fig. 5). NAA/water signal is 9.9% increased when a localized transmitter voltage optimization is applied compared to when the nominal voltage is used for acquisition.
Top row: Flip angle distribution maps acquired at 9.4T once with AFI and once with the Turbo-FLASH (TFL) sequence at the nominal voltage (V=250V). The flip angle maps were used for local power optimization.
Bottom row: Unsuppressed water peaks acquired with the nominal voltage (left), with the adjusted voltage as calculated from the AFI B1+ map (middle), and with the adjusted voltage as calculated from the TFL B1+ map (middle)
Fig. 5A: NAA/water with TurboFLASH adjusted transmitter voltage
Fig. 5B: NAA/water with the nominal transmitter voltage
Fig. 5C: Variations of NAA/water for both nominal and adjusted transmitter voltages. The smaller variation of signals acquired with Vadj is suggestive of more stable signal acquisition throughout coil loading positions.