Modern magnetic resonance spectroscopic (MRS) pulse sequences frequently overlook the issue of unwanted coherence pathways. A novel and robust algorithm which only requires input of the desired coherence(s) was developed to optimally crush all unwanted coherence pathways for any MRS pulse sequence. Experiments were performed on the GE BRAINO phantom comparing crusher schemes obtained from the literature with those obtained from the developed optimization algorithm for sLASER and MEGA-sLASER. The results demonstrate that the effects of unwanted coherences can be drastically reduced through the implementation of an optimized crusher scheme, without the need for additional or stronger crushers.
In a magnetic resonance spectroscopy (MRS) pulse sequence a scheme of pulsed magnetic field gradient "crushers" are played out during times in-between radiofrequency (RF) pulses to dephase unwanted coherence pathways which are incidentally perturbed by the RF pulses. Figure 1 and Figure 2 are pulse sequence and coherence pathway diagrams employing 5 and 7 RF pulses, respectively. The number of measured coherence pathways with maximum coherence order of 1 can be expressed by $$$3^{n-1}$$$, where n is the number of RF pulses. The number of unwanted coherences is therefore largely increased for modern sequences such as LASER [1], sLASER [2] or MEGA-sLASER [3] over three-pulse experiments and the challenge of designing an appropriate crusher scheme becomes more difficult. A strategy for designing a crusher scheme is to adjust crusher areas until the experimentally measured spectrum appears free from the typical unwanted pathway characteristics, however unwanted pathways can be easily overlooked. As such there can be found examples in the literature of MRS sequences which do not crush all unwanted pathways, for example the crusher schemes used by Landheer [4], Scheenen [5], Prinsen [6], and Andreychenko [3] do not crush 1, 26, 8 and 8 of the 8, 242, 728 and 728 unwanted coherence pathways, respectively. Here we present a fully automated algorithm, which requires no experimental iteration, for the design of optimal crusher schemes for MRS sequences, as well as its application to modern MRS sequences (sLASER and MEGA-sLASER).
1. Garwood, M. & DelaBarre, L. The Return of the Frequency Sweep: Designing Adiabatic Pulses for Contemporary NMR. J. Magn. Reson. 153, 155–177 (2001).
2. Scheenen, T. W., Heerschap, A. & Klomp, D. W. Towards 1H-MRSI of the human brain at 7T with slice-selective adiabatic refocusing pulses. MAGMA 21, 95–101 (2008).
3. Andreychenko, A., Boer, V. O., Arteaga de Castro, C. S., Luijten, P. R. & Klomp, D. W. J. Efficient Spectral Editing at 7 T: GABA Detection with MEGA-sLASER. Magn Reson Med 68, 1018–1025 (2012).
4. Landheer, K., Sahgal, A., Das, S. & Graham, S. J. Constrained Source Space MR Spectroscopy: Multiple Voxels, No Gradient Readout. Am J Neuroradiol 37, 1–8 (2015).
5. Scheenen, T. W. J., Klomp, D. W. J., Wijnen, J. P. & Heerschap, A. Short echo time 1H-MRSI of the human brain at 3T with minimal chemical shift displacement errors using adiabatic refocusing pulses. Magn Reson Med 59, 1–6 (2008).
6. Prinsen, H., de Graaf, R. a, Mason, G. F., Pelletier, D. & Juchem, C. Reproducibility measurement of glutathione, GABA, and glutamate: Towards in vivo neurochemical profiling of multiple sclerosis with MR spectroscopy at 7T. J Magn Reson Imag 45, 187–198 (2017).
7. Mitschang, L., Ponstingl, H., Grindrod, D. & Oschkinat, H. Geometrical representation of coherence transfer selection by pulsed field gradients in high-resolution nuclear magnetic resonance. J. Chem. Phys. 102, 3089–3098 (1995).