Dahan Kim1,2, Katrina Ruedinger3, David Rutkowski3, Alejandro Roldán-Alzate4,5, Patrick Turski2,5, and Kevin M Johnson2,5
1Department of Physics, University of Wisconsin - Madison, Madison, WI, United States, 2Department of Medical Physics, University of Wisconsin - Madison, Madison, WI, United States, 3Department of Biomedical Engineering, University of Wisconsin - Madison, Madison, WI, United States, 4Department of Mechanical Engineering, University of Wisconsin - Madison, Madison, WI, United States, 5Department of Radiology, University of Wisconsin - Madison, Madison, WI, United States
Synopsis
We examined velocity measurements of standard (STD-PC) and
ultra-short echo (UTE-PC) phase-contrast MRI in three different study cases, to
characterize the effect of shortening echo time on artifacts from flow and
metal. We found that UTE-PC measures higher velocity magnitudes not only in
disturbed flow but also in normal vasculature, that UTE-PC results in smaller
divergence of velocity field but no difference in erroneous flux through
arterial wall within metal aneurysm stent, and that UTE-PC higher velocity
magnitudes, less signal loss, and coherent flow directions in both untreated
and un-treated aneurysm phantom.
Introduction
Artifacts from flow and metal are often
sensitive to echo time and by shortening the echo time (TE) intra-voxel
dephasing based artifacts can be reduced. For 4D-flow imaging, the shortest possible
echo times can be achieved using center out 3D radial sampling1, so called ultra-short echo time phase contrast (UTE-PC).
Unfortunately, center-out radial also has lower sampling efficiency and higher
noise than other radial and Cartesian sampling strategies. Hence, UTE-PC may
result in degraded image quality in areas without disturbed flow, metal devices,
or disturbed flow. The general
consistency between STD-PC and UTE-PC has not been well characterized in
clinical intracranial scans or in the presence of metal. In this work, we
compared STD-PC and UTE-PC in three different study cases, and intended to
assess the accuracy of each technique with different data fidelity measures. Methods
Using a 3T scanner (MR750, GE Healthcare,
WI, USA) with a 32-channel head coil (Nova Medical, MA, USA), standard echo
(STD-PC) and ultra-short echo (UTE-PC) phase-contrast MRI were performed in (1)
seven human subjects with known neurovascular disease, (2) a patient treated
with a stent for intracranial aneurysm, and (3) two patient specific
intracranial aneurysm phantoms, one untreated and the other pseudo-treated with
endovascular coiling. Both scans were acquired with 3D radial sequences1,2 with 0.6mm isotropic spatial resolution, and were
registered (3D, rigid) to each other using their magnitude images with a mutual
information metric (ANTs)4. For (1), both
STD-PC and UTE-PC were reconstructed using standard (STD) and iterative
compressed sensing (CS) reconstructions. Flow vectors were compared voxel-wise between
STD-PC and UTE-PC for blood vessels over the whole brain, identified with a
manual threshold in the PC-derived angiogram (PC-MRA). For (2), STD-PC and
UTE-PC were assessed for self-consistency using the following two measures: (i)divergence
of velocity fields, normalized to flow magnitudes, and (ii)flux of blood flow
through the vessel wall. The through-wall blood flux was calculated by taking
the dot-product between interpolated velocity vectors at the vessel wall and
normal vectors at the vessel wall surface, evaluated over all surface area
elements in the surface mesh of the segmented vessel. For (3), the two phantoms
were connected in series and imaged side-by-side using a flow pump at steady
flow rate 0.5L/min.Results
Comparison between STD-PC and UTE-PC from
the 7 subjects shows that UTE-PC measures higher velocity than STD-PC on the
average (Fig.1a-b), when examined over the entire vasculature of the brain. Compressed
sensing (CS) did not change that trend or the magnitude of the disagreement
between the two techniques. When each of STD-PC and UTE-PC were compared with
itself between standard (STD) and CS reconstructions(Fig.2a-b), CS also did not
make any meaningful difference in the flow values of PC-MRI. STD-PC and UTE-PC
shows an excellent agreement in the flow directions, in both STD and CS
reconstructions.
Divergence of flow vectors, normalized
relative to the flow magnitude, was higher within the metal stent, which is the
area of signal loss, compared to the volume outside the stent(Fig.2). Within
the stent, the normalized divergence was lower in UTE-PC than in STD-PC,
indicating greater self-consistency of UTE-PC in its flow values. When the
blood flux was calculated through the stent wall, there was no meaningful
difference between STD-PC and UTE-PC, despite both having higher flux within
the stent than outside the stent(Fig.4) .
UTE-PC in the aneurysm phantoms are
characterized with more homogeneous signal, higher velocity magnitudes, and
directions in agreement with general flow. STD-PC shows areas of signal loss
near both the treated(Fig.5b) and untreated(Fig.5d) aneurysms.Discussion/Conclusion
Although shortening TE has been shown to
reduce flow artifacts in stenotic phantoms5,6,
we find it true even when the entire cerebral vasculature was examined, not
just the areas of disturbed flow.
Divergence must be zero for a completely
accurate measurement of velocity field of blood, and was used here as a metric
to assess the accuracy of velocity measurements, without reference to ground
truth. We found lower divergence in UTE-PC than STD-PC, and that UTE-PC yields
more self-consistent velocity fields.
Since macroscopic blood flow does not
occur through the arterial vessel wall, we computed the through-wall flux as a
measure of velocity field accuracy. However, we did not find any meaningful
difference with this metric between UTE-PC and STD-PC in the stent examined,
especially with the signal loss in the area.
Although UTE-PC showed higher velocity
magnitudes, we could not characterize the effect of metal due to presence of
disturbed blood flow, suggesting improved phantom design free of disturbed flow
for future studies.Acknowledgements
No acknowledgement found.References
1. Johnson KM. 5-Point, Ultra-Short TE, 3D
radial Phase Contrast: Improved characterization of Complex and Turbulent flow.
2010. Proc of 18th Meeting and Exhibition of the International Society for
Magnetic Resonance in Medicine. Stockholm, Sweden. Abstract #66
2. Johnson KM, Lum DP, Turski PA, Block WF,
Mistretta CA, Wieben O. Improved 3D phase contrast MRI with off-resonance
corrected dual echo VIPR. Magn Reson Med. 2008;60(6):1329-1336.
3. Johnson KM, Fain SB, Schiebler ML, Nagle S.
Optimized 3D Ultrashort Echo Time Pulmonary MRI. Magn Reson Med.
2013;70(5):1241–1250.
4. B.B. Avants, N.J. Tustison, G. Song, P.A.
Cook, A. Klein, J.C. Gee. A reproducible evaluation of ANTs similarity metric
performance in brain image registration. Neuroimage. 2011;54(3)2033-2044
5. O’Brien KR, Cowan BR, Jain M, Stewart
RA, Kerr AJ, Young AA. MRIphase contrast velocity and flow errors in turbulent
stenotic jets. JMagn Reson Imaging 2008;28:210–218.