Placental dysfunction is widely accepted as a major cause of common adverse pregnancy outcomes. However, current clinically available tools are limited in their ability to assess placental function. Here, we have incorporated quantitative MRI methods to estimate placental metabolic rate of O2 (PMRO2), which may contribute to the understanding of placental dysfunction. Presented results demonstrate the feasibility of estimating PMRO2 in vivo. Further, these data suggest, in a limited number of participants, that PRMO2 may in fact contribute to the understanding of pregnancy complications.
Five pregnant women (gestational age, GA=33±4wks) were scanned in a 1.5T (Siemens Avanto, Erlangen, Germany) scanner with two flexible body coils. MRO2 can be estimated from Fick’s principle:
MRO2=Ca·BFR·(SaO2-SvO2) $$$\space\space\space\space\space\space\space\space\space\space$$$ [1]
Here, Ca is the blood O2 carrying capacity in units of µmol O2/dL blood (Ca = 59.8 µmol O2/g Hb·[Hb], where [Hb] is the blood hemoglobin concentration in g/dL blood), BFR in mL min-1/kg of fetal mass (massf) and SaO2-SvO2 is the arterio-venous difference in oxygen saturation. PMRO2 can then be calculated by subtracting the contribution of fMRO2 to uMRO2:
PMRO2=uMRO2 - fMRO2
$$$\space\space\space\space\space\space\space\space\space\space\space$$$ =[Ca,m·BFRm·(SaO2,m-SvO2,m)]-[Ca,f·BFRf·(SaO2,m-SvO2,m)] $$$\space\space\space\space\space\space\space\space\space\space$$$ [2]
where the subscripts m and f indicate parameters measured in the maternal and fetal circulations, respectively. Location of measurements and techniques used are summarized in Figure 1B. Maternal hemoglobin (Hb) and hematocrit (Hct) values were obtained from the medical records closest to the MRI scan date. Fetal Hb values used were population averages.1 Flow rate of the uterine arteries was estimated with 4D-flow MRI. Imaging parameters: TR/TE=5.5/2.9ms, FA=8°, FOV=320×320×30mm3, voxel size=1.25×1.25×1.25mm3, segments=2, temporal resolution=44ms, acceleration factor=4.7, VENC=120cm/s. Blood flow velocity at the umbilical vein (UV) was measured using phase-contrast (PC) MRI during maternal breath-hold. Imaging parameters: velocity encoding (VENC)=20cm/s, TR/TE=10.6/6.5ms, FOV=200×200mm2, voxel size=0.625×0.625×5mm3, FA=20°, and scan duration of 10s. Oxygen saturation values at the ovarian vein (SvO2,m) and the umbilical vein (SaO2,f) were extracted using a background-suppressed (BS) T2-prepared balanced steady-state free precession (T2-bSSFP).2,3 Blood T2 values were converted to oxygen saturation using previously established calibration curves for adult and fetal blood, respectively.3,4 Fetal venous oxygen saturation, SvO2,f, was measured using susceptometry-based oximetry (SBO) in the fetal abdominal aorta (AAo) as it feeds the umbilical arteries carrying deoxygenated blood to the placenta after fetal oxygen extraction. SBO was measured with an axial single-slice three-echo RF-spoiled GRE acquisition used to estimate the difference of inter-echo phase accrual between blood in the abdominal aorta and surrounding tissue, Δφ. This sequence was interleaved with a dual-slice acquisition using simultaneous echo refocusing (SER), 10mm superiorly and inferiorly to the SBO imaging slice to estimate θ, the vessel tilt with respect to B0. Imaging parameters: TR=45ms, TEGRE=10ms, TESER,1=8.8ms, TESER,2=3.5ms, FOV=307×307mm2, voxel size=1×1×5mm3, FA=15°, and scan duration of 20sec. All regions of interest (ROIs) were manually traced on the magnitude image using ImageJ and applied to a bias field corrected phase map.5 Fetal mass was calculated from fetal volume.6
1. Jopling J, Henry E, Wiedmeier SE, Christensen RD. Reference ranges for hematocrit and blood hemoglobin concentration during the neonatal period: data from a multihospital health care system. Pediatrics 2009;123(2):e333-7.
2. Maleki N, Dai W, Alsop DC. Optimization of background suppression for arterial spin labeling perfusion imaging. MAGMA 2012;25(2):127-33.
3. Rodríguez-Soto AE, Abdulmalik O, Langham MC, Schwartz N, Lee H, Wehrli FW. T2 -prepared balanced steady-state free precession (bSSFP) for quantifying whole-blood oxygen saturation at 1.5T. Magn Reson Med 2017.
4. Rodríguez-Soto AE, Langham MC, Abdulmalik O, Englund E, Schwartz N, Wehrli FW. MRI Quantification of Human Fetal O2 Delivery Rate in the Second and Third Trimesters of Pregnancy. Magn Reson Med 2017. In Press.
5. Langham MC, Magland JF, Floyd TF, Wehrli FW. Retrospective correction for induced magnetic field inhomogeneity in measurements of large-vessel hemoglobin oxygen saturation by MR susceptometry. Magn Reson Med 2009;61(3):626-33.
6. Baker PN, Johnson IR, Gowland PA, et al. Fetal weight estimation by echo-planar magnetic resonance imaging. Lancet. 1994;343(8898):644-645.
7. Saini BS, Zhu M, Portnoy S, Porayette P, Lim J, Duan A, Sled JG, Wald R, Windrim R, Macgowan C, Kingdom JC, See M. OP29.07: Non-invasive in utero measurements of placental oxygen transport using MRI. Ultrasound Obstet Gynecol. 2016;48(S1):148.
8. Richardson B, Nodwell A, Webster K, Alshimmiri M, Gagnon R, Natale R. Fetal oxygen saturation and fractional extraction at birth and the relationship to measures of acidosis. Am J Obstet Gynecol 1998;178(3):572-9.