Nicolas Boulant1, Vincent Gras1, and Pierre-François Van de Moortele2
1NeuroSpin, CEA, Saclay, France, 2CMRR, University of Minnesota, Minneapolis, MN, United States
Synopsis
Recent numerical studies have shown that temperature could be controlled
explicitly at the pulse design stage in parallel transmission, furthermore reporting that further scan
performance could be gained if the more relevant safety metric that is temperature, instead of the specific absorption rate, was controlled. Bioheat models yet are still undergoing experimental validations. So as a first step, this work reports an experimental demonstration of flip-angle homogenization with temperature control on a water phantom with parallel transmission at 7T.
Introduction
Although
temperature rise induced by RF waves is recognized as a more relevant safety
measure in MR exams than the Specific Absorption Rate (SAR), the latter has
been ubiquitously used as gold standard for simplicity. SAR on the other hand
is just one of the multi-factors impacting temperature and the lack of direct
relationship between the two metrics has worsened with the advent of parallel
transmission (pTx). Recently, it was shown that temperature could be controlled
explicitly at the pulse design stage with so-called temperature matrices and
resulting compressed Temperature Virtual Observation Points (TVOPs)1.
The study was numerical but nonetheless showed that further scan and safety
performance could be gained if temperature, instead of SAR, was controlled. As
a first step, hence we here present an experimental demonstration of flip-angle
(FA) homogenization with temperature control on a water phantom with pTx. Methods
Experiments
were performed on a 7T Magnetom scanner (Siemens Healthcare, Erlangen, Germany)
and with the 8Tx-8Rx RapidBiomed coil (Rapid Biomedical, Rimpar, Germany). The
sample was a 16 cm diameter agar-gel phantom with 4 g/l NaCl (s=0.78 S/m and er=72).
Electric and magnetic RF fields for each channel were returned by simulation
using the HFSS (Ansys, Canonsburg, PA, USA) software after tuning/matching the
array with cosimulation. B1 field map measurements were performed and compared
with the simulated maps. A calibration procedure based on this comparison was
then applied to correct for unaccounted losses, phase offsets and additional
coupling effects in the numerical domain2, which was propagated to
the simulated electric fields. SAR Q-matrices were then computed and fed to the
heat diffusion equation (30 min sequence) with Neumann’s boundary conditions
(h=10W/m2/K) to build temperature matrices, which were then
compressed into 493 TVOPs1. These matrices were directly used into a
non-selective pulse design1 consisting of 11 kT-points
targeting a 60° FA (duty-cycle of 3.75%, TR=80 ms) with explicit constraints
of: maximum temperature rise (0.4°C), maximum average power (1.9 W/channel) and
peak voltage (160 V/channel). The FA homogenization of the pulse was first
verified with the AFI sequence3. Second, MR thermometry4 was used to measure the temperature rise induced by the pulse sequence. For
that purpose, a 3D-GRE sequence was applied (TE=28ms) to return a reference
phase φ1 for
each voxel. The kT-points pulse was then played in a 30 min-long
gradient-free RF-heating sequence (TR = 80 ms), after which a second GRE
sequence was implemented to return a locally-shifted phase φ2.
Temperature rise was then computed using T(r)=(φ2 - φ1)/(αTE) with α being the electron-screening constant4.
Field stability prior to heating was verified by repeating the GRE sequence 10
times which showed over 10 minutes a fluctuation of ~0.02 rad.Results
RF
pulse design could be performed in about 10s and returned a pulse which
saturated all constraints. In agreement with the simulations, the FA measurements
provided in Fig.1 show good FA homogenization, although a slight offset most
likely due to a B1 map or pulse implementation uncertainty appears. Fig. 2
reports the simulated temperature rise induced by each of the 11 kT-point
sub-pulses separately (illustrating temporal averaging), the overall effect of
the pulse being the average of the temperature rise maps across all sub-pulses
(yellow-framed image). The latter is reproduced in Fig.3 over 3 orthogonal
slices (top row) and compared with temperature rise measurements (bottom row). For this sequence duration, by
simulation we found that heat diffusion had for effect to roughly halve the
maximum temperature rise. Discussion and conclusion
This
work reports the first in vitro demonstration of explicit temperature control in
pulse design. The maximum temperature rise in the design here was set to 0.4 °C
and could be verified experimentally (Fig. 3). A separate numerical
optimization revealed that the maximum temperature rise obtainable with the
same average power constraints was 0.97 °C but, interestingly, not more than
0.4 °C rise was necessary to achieve FA-homogenization optimality (perhaps due
to temporal averaging). In this case it appeared that the most limiting constraint
was the average power enforced for coil protection. Thus for this coil, given
the long sequence duration and the lack of efficient cooling mechanism
(perfusion) in the sample, it appears that significant tissue heating in normal
in vivo conditions is unlikely. To conclude, given the very good agreement
between temperature simulations and measurements, controlling temperature in
vivo through the use of TVOPs appears to be within reach, although significant
efforts in bioheat equation validations must be invested first.Acknowledgements
Acknowledgements: ERC EXPAT (grant number 309674).References
[1] N. Boulant, X. Wu, G. Adriany, S. Schmitter, K. Ugurbil, P-F. Van de Moortele. Direct control of the temperature rise in parallel transmission by means of temperature virtual observation points: Simulations at 10.5 tesla. Magn Reson in Med 75:249-256 (2016).
[2] A. Beqiri, J. W. Hand, J.V. Hajnal, S. Malik. Comparison between simulated decoupling regimes for specific absorption rate prediction in parallel transmit MRI. Magn Reson in Med 74:1423-1434 (2015).
[3] V. Yarnykh. Actual flip angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson in Med 57:192-200 (2007).
[4] V. Rieke and K.Butts-Pauly. MR thermometry. J of Magn Reson Imag 27:376-390 (2008).