Jing-Huei Lee1, Jullie W Pan2, and Hoby P. Hetherington2
1Biomedical Engineering, University of Cincinnati, Cincinnati, OH, United States, 2Radiology, University of Pittsburgh, Pittsburgh, PA, United States
Synopsis
A novel dynamic spatially selective dephasing (DSSD)
for outer volume suppression (OVS) is introduced. This technique uses high-order
shim pulses to accomplish OVS by creating a low and high field gradient within
and outside of region of interest. The effectiveness of DSSD is investigated with
phantom and human using an 8-channel parallel transmit 7 Tesla Siemens system. Similar
suppression performance was obtained in phantom and human brain using DSSD with
a circular and an elliptical shape, respectively. The results show that OVS can be achieved by
using a high degree shim insert to generate a steep B0 gradient at periphery of
target region.
Introduction
Outer volume suppression (OVS) is commonly used in MR
spectroscopic imaging (MRSI) to reduce contamination from strong lipid and
water signals of peripheral regions. Traditionally, these signals are suppressed
using slice-selective pulses that are placed around the brain to approximate
the elliptical shape of the skull (1). The transverse magnetization generated
is then dephased with large crusher gradients. Recently, an in-plane OVS was
introduced using RF shimming and an eight-element transceiver array which
excited a ring about the periphery of the head and left the central brain
regions largely unaffected (2). Another
approach for OVS introduced the so-called gradient rotating outer volume
excitation method which uses a single RF pulse together with a varying gradient
to produce a frequency-swept excitation of a desired region (3). In this work,
we introduce a novel OVS method, dynamic
spatially selective dephasing (DSSD), which uses high-order shim pulses to
create a low and high field gradient within and outside of region of interest
(ROI), respectively. Thus, OVS is accomplished by de-phasing the transverse
magnetization outside of ROI due to the high shim field gradient. Transverse magnetization within ROI is
maintained in phase since overall spatial derivative of B0
distribution is small.Methods
High-order shim gradient pulses required for OVS were
carried out using custom-written spherical harmonic gradient waveforms
optimization software in Matlab (The MathWorks, Inc., Natick, MA, USA). The
program calculated and stored multiple shim gradient strengths so that the
required shims could be effectively transferred to scanner and examined. The
OVS performance was initially tested and optimized using a 17cm water-filled spherical
MRI phantom. Healthy volunteers (n = 5) participated in the study after giving
informed consent according to procedures approved by the Institutional Review
Board of the University of Pittsburgh. All scans were conducted on
an 8-channel parallel transmit 7 Tesla (T) Siemens system (Erlangen, Germany). An
18-channel shim insert coil with 38-cm ID, 43cm OD, and 70-cm length, 18.5cm
from patient end to magnetic center, consisting of 3rd, 4th, and two 5th-degree
shims (ZC4 and ZS4) with 10A shim supplies (Resonance Research, Billerica, MA
USA) was used for high-degree shimming. The B0 field used for OVS optimization
was obtained using the B0 loop
encoded readout (Bolero) method (4). Shimming was performed by optimizing the
shim current values using a least-squares algorithm based on a calibrated
spherical harmonic model (4) to minimize the standard
deviation of the B0 field over the target ROI.
The pulse sequence used for acquiring 2D MRSI is shown in Figure 1. The B0 shims are
pulsed during a spin echo sequence either single waveform or paired waveform of
opposite sign or potentially distribution so as to enhance dephasing.Results
The results show that the DSSD pulse with optimized amplitudes can
selectively suppress the target outer volume . Figures 2A and 2D show the images
of a phantom from NAA without and with the DSSD pulse,
respectively. The corresponding spectra from the phantom were acquired with the
DSSD pulse off (Fig. 2B and 2C) and on (Fig 2E and 2F). Figures 2G and 2H
display the retention of NAA magnetization for the human
brain with red=1.0 for full signal retention and blue=0.0 for full suppression.Discussion
The results show that OVS can be achieved by using a high degree shim
insert to generate a steep B0 gradient at periphery of the target
region. The DSSD pulse is a transient change in B0 shim values, and
is added onto the optimal setting only during the shim pulse. Results also show
that DSSD requires less SAR and time than traditional OVS schemes.Conclusion
This report outlines the theory and experimental verification of the novel DSSD approach.
The efficacy of this new technique is demonstrated both in phantom and human
subjects.Acknowledgements
No acknowledgement found.References
1. Duyn
JH, Gillen J, Sobering G, et al. Radiology. 1993; 188:277.
2. Hetherington HP,
Avdievich NI, Kuznetsov AM, et al. Magn Reson Med. 2010; 63:9.
3. Powell NJ, Jang A, Park JY, et al. Magn Reson Med. 2015;73:139.
4. Hetherington HP, Chu WJ,
Gonen O, Pan JW. Magn Reson Med 2006;56:26–33.