Rita Schmidt1,2, Amir Seginer3, and Assaf Tal3
1Radiology, Leiden University Medical Center (LUMC), Leiden, Netherlands, 2Neurobiology, Weizmann Institute of Science, Rehovot, Israel, 3Chemical physics, Weizmann Institute of Science, Rehovot, Israel
Synopsis
One of the powerful techniques for
faster spectroscopic imaging acquisition is so called proton echo planar
spectroscopic imaging (EPSI). An alternative method, which we here term COKE
(COherent K-t space EPSI), relies
on interleaving “blipped” PE gradients in between readouts to produce coherent
phase between the k-t space lines. COKE enables to double the SW in comparison
to EPSI or to halve the acquisition duration. We designed a sequence that
integrates a multi-band CAIPIRINHA with COKE acquisition for faster spectroscopic
imaging. This was demonstrated in both phantoms and in-vivo.
Introduction
Magnetic resonance spectroscopic imaging (MRSI) is a value added
modality, augmenting structural scans with metabolic spatial information.
However, it usually requires long scan durations. One of the powerful techniques for faster
acquisition is so called proton echo planar spectroscopic imaging (EPSI), which
delivers spectroscopic and 1D spatial information in a single excitation (Fig.
1a), with additional spatial dimensions undergoing phase encoding1-3.
EPSI trajectories suffer from undesired inconsistencies between readout lines
acquired with alternating positive and negative gradients4. To
overcome this, the positive and negative readouts are conventionally processed
individually5 and then combined, at the cost of halving the spectral
width (SW); or a flyback readout is used6, foregoing the negative readouts.
An alternative method7, which we here term COKE (COherent K-t space
EPSI), relies on interleaving “blipped”
PE gradients in between readout lines in order to produce coherent phase
between the k-t space readout lines (for a given PE). COKE introduces phase inconsistencies
along the PE dimension, but these are straightforward to correct as shown in
many EPI reconstruction schemes8. The full COKE readout (Fig. 1b)
doubles SW in comparison to EPSI. In an alternative realization, denoted here
as “diet” COKE (Fig. 1c), the SW is halved, yielding the equivalent EPSI
dataset at half the acquisition time. In this work, we integrated multi-band
CAIPIRINHA slice selection9 with the COKE interleaved PE gradients. By
introducing dual-band CAIPI COKE (Fig. 1d) and dual-band CAIPI “diet” COKE
(Fig. 1e) sequences, one can further accelerate the measurement two-fold. Methods
We
implemented and compared EPSI, COKE, diet COKE, and dual-band CAIPI variants
(Figs. 1a-e) in phantoms and in human volunteers at 3T (Tim Trio, Siemens,
Erlangen). Note that, in order to integrate the multi-band CAIPIRINHA into COKE,
one must switch the slice-selection gradient’s polarity between the even and
odd PE lines, respectively, since COKE interleaves the PE lines during each
excitation. The reconstruction of the data includes reordering and phase
corrections due to PE and CAIPI phase shifts. We compared EPSI, diet-COKE and
dual-band diet-COKE in phantoms mimicking brain metabolites and in a fat/water
phantom (based on white cheese and oil).
The single-band excitation pulse used was an SLR-90°
with bandwidth-time product of 18 and duration of 3.8 ms, whereas the dual-band
excitation pulse in Fig. 1d,e was based on SLR-90° with a bandwidth-time
product of 12 and duration of 5 ms. The
metabolite phantom included 10mM L-Glutamic acid,
10mM Creatine 8mM myo-Inositol, 2mM GABA, 2mM Choline chloride, 5mM
Choline chloride, 5mM Sodium lactate 12.5mM NAA and 0.5mL Gd for short T1
(T1 of water was 625 ms). The relevant scan parameters are
summarized in the relevant figure captions.Results
Fig. 2 shows phantom results comparing “diet”
COKE and EPSI. EPSI data was reconstructed using odd readouts only for spectrum
quality comparison of COKE reconstruction and equivalent SNR evaluation (having
the same as “diet” COKE SNR∕√2). The measured SNR ratio was 1.06, which is
close to the expected factor of unity. Fig. 3 demonstrates dual-band “diet”
COKE phantom results for spectroscopic imaging, with the brain mimicking
phantom and the fat/water phantom. A 5 cm gap between the slices was introduced
to ensure their contents differ sufficiently and for clarity of presentation;
however, the sequence works also well for adjacent slices (not shown), as was
shown before with CAIPIRINHA implementation8. Fig. 4 shows in-vivo brain MRSI acquisition using dual-band
COKE implementation, showing it is capable of producing multi-slice
spectroscopic images (with an acceleration factor of two due to dual-band and
using the double SW versus EPSI).Conclusions
In this study, we designed
a sequence that integrates a multi-band CAIPIRINHA with COKE acquisition for
faster MRSI. This was demonstrated in both phantoms and in-vivo. Alternatively, dual-band COKE can be used to
double the SW of EPSI, which is a limiting factor in-vivo, particularly at
ultrahigh (≥7T) fields where SW
requirements can become prohibitive. Acknowledgements
We are
grateful to Osnat Volovyk for preparation of the phantom mimicking brain
metabolites. We acknowledge the Monroy-Marks Career Development Fund and the historic generosity of the Harold Perlman Family.References
[1] Mansfield, P.
(1984) Magn. Reson. Med., 1(3), 370-386, [2] Posse, S., et.al
(1997) Magn. Reson. Med., 37(6), 858-865, [3] Lin F. H., et.al.
(2007) Magn. Reson. Med., 57(2), 249-257, [4] Posse, et.al. (2013) J. Magn. Reson. Imag., 37(6), 1301-1325, [5] Posse S., et al. (1994) Radiology, 192, 733-738, [6] Cunningham C.H. et. al. (2005) Magn. Reson. Med.,54,1286–1289, [7] Webb P., et.al. (1989) Magn. Reson. Med., 12(3), 306-315,[8] Chen N., et. al. (2004) Magn. Reson. Med., 51,1247–1253, [9] Setsompop K., et.al. (2012) Magn. Reson. Med., 67(5), 1210-1224.