Shengzhen Tao1, Yunhong Shu1, Joshua Trzasko1, Myung-Ho In1, Erin Gray1, John Huston III1, and Matt Bernstein1
1Radiology, Mayo Clinic, Rochester, MN, United States
Synopsis
Recently, a low-cryogen, compact 3T MRI system optimized for brain, extremity and infant imaging was developed. The system is
equipped with a high slew rate gradient capable of 80 mT/m maximum gradient
amplitude and 700 T/m/s slew rate. Due to its reduced imaging volume (26-cm
diameter-spherical-volume), the high gradient amplitude and slew rate can be
achieved simultaneously on this system with substantially less peripheral nerve
stimulation. Here, we investigate the benefit of performing spiral imaging using
high gradient performance available on this system. We demonstrated that the
high slew-rate can significantly reduce spiral readout time and therefore
reduce off-resonance-induced blurring.
Purpose
The specialized MRI platforms optimized for
imaging particular organs (e.g., brain) have generated interest due to their
advantages compared to the conventional, whole-body MRI, including easier
siting and reduced risk of peripheral nerve stimulation (PNS)1-3. Recently,
a low-cryogen, compact 3T (C3T) MRI system capable of imaging brains,
extremities, and infants was developed1. It has a 26-cm diameter-spherical-volume,
and is equipped with a high-performance gradient (80mT/m maximum gradient
amplitude and 700T/m/sec slew-rate), in comparison to the 200T/m/s, 50mT/m conventional,
60-cm bore whole-body system like a GE MR750. On the whole-body systems, the
PNS concerns usually impose further, stringent limitations on gradient performance.
Therefore, the maximal 200T/m/sec slew-rate (SR) is not always available during
readout4. However, due to its reduced size, high gradient amplitude
and SR can be achieved simultaneously on the C3T with substantially less PNS1.
The high SR can particularly benefit acquisitions such as echo-planar-imaging (EPI)
by reducing readout time and consequently susceptibility artifacts5.
Similarly, spiral acquisitions can benefit from a high SR. Here, we evaluate
the readout time reduction for spiral acquisition based on the standard gradient
performance available on a GE750 whole-body system, and the high SR on the C3T.
Phantom experiments were performed to demonstrate the reduced susceptibility
effect using high SR. Methods
Two-dimensional fully-sampled Archimedean6
and variable-density spiral7 readouts were generated in simulation. On
the MR750 system, the SR is by default limited to 70T/m/sec due to PNS risk4.
We therefore chose 50mT/s gradient amplitude and 70T/m/sec SR to represent the
whole-body system. Simulations were also performed assuming the full gradient SR
of the whole-body gradient (200T/m/s) and the C3T (700T/m/s). Simulations were
performed assuming a brain scan FOV=24cm, combined with different resolutions (1~10mm),
number of interleaves (2~32), and various bandwidth (±250/±125/±62.5kHz). The variable-density
spiral was designed using the method described by Lee et al.7. The
trajectory is fully-sampled at the k-space center; as the trajectory departs
the center, the sampling density gradually decreases in a linear fashion. At
each sampling point specified by its k-space radial distance ($$$k_r$$$), the instant sampling rate $$$\Delta k(k_r)=1/FOV-k_r/k_m(1-\alpha)/FOV$$$, i.e., an undersampling factor α is
achieved at the end when the maximal radial sampling distance $$$k_m$$$ is reached ($$$\Delta k(k_m)=\alpha/FOV$$$). To demonstrate the advantage of higher SR,
the American College of Radiology (ACR) MRI phantom8 was scanned on
the C3T using a variable-density spin-echo spiral (FOV=24cm, acquisition resolution=3.5mm,
BW=±125kHz, readout=1.4ms, 8 interleaves, α=3, flip angle=90°, 3mm slice, TR/TE=100/7.1ms).
The low-resolution setup used here was relevant for applications such as spiral-based
arterial spin labeling (ASL)9. The slew rate was set to 200 T/m/s
for this initial test. The first/zeroth-order concomitant fields present on the
asymmetric gradients were compensated by gradient pre-emphasis10 and
frequency tracking11. Images were reconstructed onto a 256×256
matrix using iterative-SENSE solved with conjugate gradient12 without/with
off-resonance correction using time segmentation13,14. Results
Figures 1 shows the fully-sampled spiral
trajectory and readout gradient designed assuming (a) standard whole-body system
SR=70T/m/sec, (b) high SR=200T/m/s, and (c) full SR=700T/m/s on the C3T, all
with BW=±125kHz. Figure 2 shows the same plot for a variable-density spiral (α=3).
The readout time is reduced from 4.0ms (SR=70) to 2.4ms (SR=200) and 1.9ms (SR=700)
for fully-sampled spiral, and from 2.4ms (SR=70) to 1.4ms (SR=200) and 0.8ms (SR=700))
for variable-density spiral. The readout time ratio between the two high SR
setups and the standard SR setup was shown in Figure 3 as a function of resolution
and number of interleaves for fully-sampled and variable-density spiral (α=3). Figure
4 shows the readout time ratio as a function of undersampling factor α and
resolution. A lower number suggests greater readout time reduction. Figure 5
shows an ACR phantom image acquired using variable-density spiral before/after
off-resonance correction, as well as the off-resonance map. About 400 rad/s off-resonance
was observed at the periphery of the phantom. Due to the short readout time (1.4
ms), off-resonance blurring was much less apparent using a variable-density
spiral and high SR. Discussion
These results demonstrate that the high SR
available on the C3T system can reduce readout time for spiral acquisitions.
Higher SR can especially benefit acquisition typically executed with high BW
and moderate-to-low resolution, such as ASL9, fMRI15, and
MRE16. Greater readout time reduction is observed when a variable-density
spiral is used. The phantom results demonstrate that the high-slew rate can greatly
reduce the susceptibility induced blurring. Increasing SR to 700T/m/s can
further reduce readout time, but further attention for eddy-current effects may
be required17. Conclusion
We demonstrated that the high slew-rate available
on the C3T system can reduce spiral readout time and therefore reduce
off-resonance-induced blurring.Acknowledgements
This work was supported by NIH grants U01 EB024450-01 and R01 EB010065.References
1. Lee SK, Mathieu JB, Graziani D, et al.
Peripheral nerve stimulation characteristics of an asymmetric head-only
gradient coil compatible with a high-channel-count receiver array. Magn Reson
Med 2016;76:1939–1950.
2. Setsompop K, Kimmlingen R, Eberlein E, et
al. Pushing the limits of in vivo diffusion MRI for the Human Connectome
Project. NeuroImage 2013;80:220–233.
3. Weiger M,
Overweg J, Rösler MB, et al. A high-performance gradient insert for rapid and
short-T2 imaging at full duty cycle. Magn Reson Med 2017. DOI: 10.1002/mrm.26954
4. King KF, Schaefer
DJ. Spiral scan peripheral nerve stimulation. J Magn Reson Imaging
2010;12:164-170.
5. Tan EK, Lee SK,
Weavers PT, etc. High slew-rate head-only gradient for improving distortion in
echo planar imaging: Preliminary experience. J Magn Reson Imaging
2016;44:653-664.
6. King KF, Foo
TKF, Crawford CR. Optimized gradient waveforms for spiral scanning. Magn Reson
Med 1995;34:156-160.
7. Lee JH,
Hargreaves BA, Hu BS, Nishimura DG. Fast 3D imaging using variable-density
spiral trajectories with applications to limb perfusion. Magn Reson Med 2003;50:1276-85.
8. Phantom test
guidance for the ACR MRI accreditation program.Reston: The American College of
Radiology, Reston, Virginia, 2005.
9. Alsop DC, Detre
JA, Golay X, et al. Recommended implementation of arterial spin-labeled
perfusion MRI for clinical applications: A consensus of the ISMRM perfusion
study group and the European consortium for ASL in dementia. Magn Reson Med
2015;73:102-116.
10. Tao S, Weavers
PT, Trzasko JD, Shu Y, Huston J, Lee S, Frigo LM,Bernstein MA. Gradient
pre-emphasis to counteract first-order concomitant fields on asymmetric MRI
gradient systems. Magn ResonMed 2017;77:2250–2262.
11. Weavers PT, Tao
S, Trzasko JD, Frigo LM, Shu Y, Frick MA, Lee SK, Foo TKF, Bernstein MA. B0
concomitant field compensation for MRI systems employing asymmetric transverse
gradient coils. Magn Reson Med 2017. doi:10.1002/mrm.26790.
12. Pruessmann KP,
Weiger M, Bornert P, Boesiger P. Advances in sensi-tivity encoding with
arbitrary k-space trajectories. Magn Reson Med2001;46:638–651.
13. Noll DC, Meyer
CH, Pauly JM, Nishimura DG, Macovski A. A homogeneity correction method for
magnetic resonance imaging with time-varying gradients. IEEE Trans Med Imaging
1991;10:629-37.
14. Sutton BP; Noll DC;
Fessler JA. Fast, iterative image reconstruction for MRI in the presence
of field inhomogeneities. IEEE Trans Med Imaging 2003;22:178-188.
15. Glover GH.
Spiral imaging in fMRI. Neuroimage 2012;15:706-12.
16. Johnson CL, McGarry
MDJ, van Houten EEW, et al. Magnetic Resonance Elastography of the brain using
multi-shot spiral readouts with self-navigated motion correction. Magn Reson
Med 2013:70:404-412.
17. Addy NO, Wu HH,
Nishimura DG. A simple method for MR gradient system characterization and
k-space trajectory estimation. Magn Reson Med 2012;68:120-129.