Luigia D'Errico1, Jens Wetzl2, Michaela Schmidt2, Aurelien F. Stalder2, Christoph Forman2, and Bernd J. Wintersperger1
1Department of Medical Imaging, University of Toronto, Toronto, ON, Canada, 2Siemens Healthcare, Erlangen, Germany
Synopsis
Iterative reconstruction
methods can improve vessel depiction in thoracic contrast-enhanced MR
angiography, particularly in small vasculature. This improvement can be
ascribed to the reduced temporal footprint for iterative reconstruction
compared to view sharing for standard reconstruction. An investigation of the
effects of this new reconstruction on quantitative vessel characteristics found
no bias in vessel diameter measurements compared to the reference.
Introduction:
Dynamic
contrast-enhanced MR angiography (CE-MRA) allows for assessment of vascular
dynamics and tissue perfusion. However, common techniques for dynamic CE-MRA
such as TWIST1 or TRICKS2 employ view sharing to increase
temporal resolution, but with a temporal footprint that is considerably longer
than the temporal resolution. This may introduce temporal blurring,
particularly in small vasculature. Recent iterative reconstruction techniques3
have enabled improved temporal footprints in dynamic MRA, demonstrating
improved signal characteristics and visual image quality. However, changes in
signal characteristics may affect quantitative parameters such as vessel
diameter measurements. This study aims to assess the impact of iterative
reconstruction in dynamic CE-MRA of the thorax with respect to signal and
quantitative vessel characteristics.Methods:
34 patients referred for cardiovascular MR with
known/suspicion of thoracic aortic abnormalities were prospectively enrolled.
Examinations were performed at 3T (MAGNETOM Skyrafit, Siemens
Healthcare, Erlangen, Germany) employing a standard parasagittal dynamic MRA
(TWIST; R=4x2 GRAPPA) with injection
of 2mL Gadobutrol diluted with 6mL NaCl (3mL/s). The 3D slab covered 105.6mm
with a 1.2x1.0x1.2mm3 voxel size. TWIST sampling included a central k-space region (A=15%) and differently
sampled peripheral k-space regions
(B=20%, B1-B5) with an acquisition time of 1.2s per individual k-space part (A/B1-B5). In addition to
standard product TWIST reconstruction, a prototype iterative reconstruction
(IT-TWIST) was employed on identical k-space
raw data sets for reconstructions on the scanner. Detailed reconstruction
parameters including the regularization were based on previous empirical
testing3. Signal intensity (SI) analysis in matching ROIs of both
reconstructions at various aortic levels (ascending, isthmus, hiatus) and
within the pulmonary artery (PA) tree (large/mid/small vessels) was performed
using ImageJ4. Automated centerline segmentation of the proximal
descending aorta with identical centerlines for TWIST and IT-TWIST data was
performed using a dedicated tool5 (Figure 1). Subsequently,
quantitative assessment of the aortic diameter was performed on time frames
with maximal signal intensity at 3 predefined levels with 1cm intervals
starting distally to the last aortic arch branch. Wall sharpness was defined as
the inverse of the distance of the 20% to 80% points between minimum background
SI and maximum vessel SI averaged over 10 profile lines perpendicular to the
centerline segmentation6 and the vessel diameter was found by
averaging the full width at half maximum (FWHM) values of these profile lines
(Figure 1). Statistical comparisons were performed using a Wilcoxon rank-sum
test and ANOVA.Results:
IT-TWIST data sets consistently
demonstrated significantly higher SI compared to TWIST at all evaluated
vascular levels and territories (Figure 2, 3). Differences were most pronounced
in mid- to small-sized pulmonary vessels. Along the pulmonary artery tree, the
signal ratio from large to small branches increased with IT-TWIST from 0.24
[0.17, 0.39] to 0.67 [0.57, 0.77] (P<0.0001), while the ratio of large
pulmonary artery signal to ascending aortic signal did not change (1.00 [0.90,
1.20] vs. 0.96 [0.90, 1.08]; P=ns). The respective average vessel wall
sharpness demonstrated only minimal differences with 0.126mm-1 for
TWIST and 0.119mm-1 for IT-TWIST reconstructions (Figure 4). The
vessel diameters for TWIST and IT-TWIST data (1cm: 22.35 [20.28, 26.41]mm vs.
22.30 [20.41, 26.61]mm; 2cm: 23.52 [21.92, 27.18]mm vs. 22.76 [22.13, 27.54]mm;
3cm: 23.77 [21.07, 26.71]mm vs. 23.84 [22.17, 27.10]mm) only showed differences
in the range of 0.24-0.43mm, well below the image resolution, and can thus be
considered equivalent.Discussion:
The results of this study demonstrate that the application
of iterative reconstruction in dynamic CE-MRA benefits respective signal
parameters. Signal characteristic improvements are most prominent in mid- to
small vessels (Figure 5). Respiratory motion and fast contrast agent transit
times are two possible explanations why blurring can occur for longer temporal
footprints, and so explain the improved depiction when the temporal footprint
is shortened. Results for vessel sharpness and vessel diameter measurements for
both reconstruction techniques were on par with each other. While this
evaluation could only be reliably performed in larger vessels due to the poor
depiction of smaller vessels in the standard reconstruction, we expect that
iterative reconstruction does not induce bias on vessel diameter measurements
of smaller vessels either.Conclusion:
The use of iterative
reconstruction in dynamic CE-MRA of the thoracic aorta improves the overall
image quality in terms of improved signal characteristics in large- and
small-sized vessels. It specifically improves the signal characteristics in
peripheral small pulmonary vessels making them better visualized and assessable
to the eye. Furthermore, quantitative vessel characteristics can be considered
identical between both types of reconstruction for clinical aspects.Acknowledgements
JW, MS, AFS and CF are employees of Siemens HealthcareReferences
- Lim RP, Shapiro M, Wang EY et al. 3D
time-resolved MR angiography (MRA) of the carotid arteries with time-resolved
imaging with stochastic trajectories: comparison with 3D contrast-enhanced
Bolus-Chase MRA and 3D time-of-flight MRA. Am J Neuroradiol. 2008;29:1847–54.
- Korosec FR, Frayne R, Grist TM et al. Time-resolved
contrast-enhanced 3D MR angiography. Magn Reson Med. 1996;36:345–51.
- Wetzl J, Forman C, Wintersperger BJ et
al. High-resolution dynamic CE-MRA of the
thorax enabled by iterative TWIST reconstruction. Magn Reson Med. 2017;77(2):833–840.
- Schindelin J,
Rueden CT, Hiner MC et al. The ImageJ ecosystem: An open platform for biomedical image
analysis. Mol Reprod Dev. 2015;82(7-8):518–529.
- Schwemmer C, Forman C, Wetzl J et al. CoroEval: a multi-platform, multi-modality
tool for the evaluation of 3D coronary vessel reconstructions. Phys Med Biol.
2014;59(17):5163-74.
- Li D, Carr J, Shea S
et al. Coronary arteries: magnetization-prepared contrast-enhanced 3D
volume-targeted breath-hold MR angiography. Radiology 2001;219:270–7.