0911

Accelerating Low-Rank Tensor Model Based Dynamic 31P-MRSI of Ischemia-Reperfusion in Rat at 9.4T
Bryan Clifford1,2, Yuning Gu3,4, Yudu Li1,2, Yuchi Liu3,4, Fan Lam2, Zhi-Pei Liang1,2, and Xin Yu3,4,5,6

1Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, IL, United States, 2Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, IL, United States, 3Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States, 4Case Center for Imaging Research, Case Western Reserve University, Cleveland, OH, United States, 5Department of Radiology, Case Western Reserve University, Cleveland, OH, United States, 6Department of Physiology and Biophysics, Case Western Reserve University, Cleveland, OH, United States

Synopsis

Dynamic 31P-MRS/MRSI is often used to assess mitochondrial oxidative capacity in skeletal muscle by monitoring the depletion and recovery of the phosphocreatine concentration during ischemia-reperfusion experiments. In animal models, standard methods are unable to provide the spatiotemporal resolution needed to discern spatial heterogeneity of the recovery process ($$$<$$$10 s/frame, $$$\approx$$$1 mm3 per voxel). To address this problem, we have improved a recently proposed low-rank tensor based method for accelerated high-resolution dynamic 31P-MRSI to provide in vivo results with 1.5x1.5x2 mm3 nominal spatial resolution, 36 ppm spectral bandwidth, 0.14 ppm spectral resolution, and 5.1 s temporal resolution.

Introduction

Mitochondrial oxidative capacity in skeletal muscle can be assessed by monitoring the depletion and recovery of the phosphocreatine (PCr) concentration during ischemia-reperfusion experiments. Dynamic 31P-MRS/MRSI is often the tool of choice for these experiments1-3 because it can provide non-invasive, in-vivo measurements of the concentrations of PCr as well as ATP and inorganic phosphate. In small animal models such as rats, the time constant of PCr recovery is approximately 60 s. Hence, a temporal resolution of less than 60 s is necessary to capture the kinetics of PCr recovery. Furthermore, the spatial resolution should be on the order of 1 mm3 if spatial heterogeneity is to be observed. Unfortunately, due to the inherently low SNR of the 31P-NMR signal, dynamic 31P-MRSI methods with adequate temporal resolution are typically limited to very low spatial resolution (on the order of 1 cm3).

Much work has been done to provide better combinations of spatial and temporal resolution4-8. Recently, a tensor-based method was introduced for providing dynamic 31P-MRSI reconstructions, achieving a resolution on the order of 1 mm3 and approximately 30 s temporal resolution9. However, for robust estimation PCr recovery rates, it is desirable to have temporal resolutions of less than 10 s. Therefore, in this work we present results from a modified acquisition and reconstruction framework which provides a factor of 6x acceleration compared to the previous work. Specifically we present results from 3 in vivo ischemia-reperfusion experiments on rat hindlimbs with 1.5x1.5x2 mm3 nominal spatial resolution, 36 ppm spectral bandwidth, 0.14 ppm spectral resolution, and 5.1 s temporal resolution.

Methods

In vivo data from three Sprague-Dawley rats were acquired on a 9.4T Bruker Biopic horizontal scanner. Each rat was anesthetized and had an inflatable cuff placed around its thigh. A Bruker 1H volume coil was used for shimming and the acquisition of the proton reference scans. A custom 31P saddle coil was placed around the animal's calf muscles to acquire MRSI data. The pressure cuff was inflated and deflated according to Fig. 1.

For each animal, we acquired two spectral subspace navigators, $$$\mathcal{D_{1,b}}$$$, $$$\mathcal{D_{1,r}}$$$, and two dynamic imaging acquisitions $$$\mathcal{D_{2,i}}$$$, $$$\mathcal{D_{2,r}}$$$ as shown in Fig. 1. $$$\mathcal{D_{1,b}}$$$ and $$$\mathcal{D_{1,r}}$$$ were acquired using a CSI-FID sequence with a 8x8x6 matrix size, 6 kHz readout bandwidth, 256 samples per readout and 8 averages. $$$\mathcal{D_{2,i}}$$$ was acquired with a 16x16x8 matrix size (stack of uniform density spirals), 90 kHz readout bandwidth, 4.66 ms echo-spacing, 17 echoes, and 4 averages, which corresponded to a nominal resolution of 1.5x1.5x2 mm3 with a FOV of 24x24x16 mm3 and a 5.1 s frame rate with a TR of 160 ms. $$$\mathcal{D_{2,r}}$$$ was similar to $$$\mathcal{D_{2,i}}$$$ except that it had 111 kHz readout bandwidth, 3.78 ms echo-spacing, and 20 echoes ($$$\mathcal{D_{2,i}}$$$ was designed with weaker gradients to reduce gradient heating during the long ischemia period). For all acquisitions the FOV, TR/TE, and flip angle were the same (TE = 0.69 ms, flip angle = 17 degrees). The total duration of each experiment was approximately 49.1 minutes.

The dynamic MRSI image was reconstructed by solving the following optimization problem:$$\begin{matrix}\{\hat{\upsilon}_m(\mathbf{x},T)\}_{m=1}^M &=&\ \arg\min\limits_{\{\upsilon_m\}_{m=1}^M}\left\Vert \mathbf{d}-\mathcal{F}\left\{r(\mathbf{x})\sum_{m=1}^M\upsilon_{m}\left(\mathbf{x},T\right) \hat{\phi}_m\left(f\right)\right\}\right\Vert_2^2+\mathrm{R}\left(\{\hat{\upsilon}_m(\mathbf{x},T)\}_{m=1}^M\right)\\&s.t.&\upsilon_m\left(\mathbf{x},T\right)=\sum_{p=1}^{P_m}\mu_{p}^{(m)}(\mathbf{x})\psi_p^{(m)}(T)\\&&\left\{\psi_p^{(m)}(T)\right\}_{m,p=1}^{M,P_m}\subset \mathrm{span}\left\{ e^{-(t-t_0)/T}u(t-t_0):t_0,T\in\mathbb{R}^+\right\}\end{matrix}$$where $$$\mathbf{x}$$$, $$$T$$$, $$$f$$$, are the spatial, temporal, and spectral coordinates, $$$\mathbf{d}$$$ is the $$$\mathcal{D}_{1,i/r}$$$ data, $$$\{\hat{\phi}_m\left(f\right)\}_{m=1}^{M}$$$ is the estimate of spectral subspace obtained from $$$\mathcal{D}_{1,b/r}$$$, $$$\mathcal{F}$$$ is the encoding and $$$B_0$$$ field inhomogeneity operator, $$$r(\mathbf{x})$$$ is a reference/support function determined from proton reference scans, $$$\left\{\psi_p^{(m)}(T)\right\}_{m,p=1}^{M,P_m}$$$ are the temporal basis functions, $$$\left\{\mu_{p}^{(m)}(\mathbf{x})\right\}_{m,p=1}^{M,P_m}$$$ are the spatial coefficients, $$$u(t)$$$ is the unit step function, and $$$\mathrm{R}\left(\cdot\right)$$$ is a regularization functional imposing a spatial smoothness prior. The model orders $$$M$$$ and $$$\{P_m\}_{m=1}^M$$$ were determined from an SVD-based analysis of the data, and given $$$\{\hat{\upsilon}_m(\mathbf{x},T)\}_{m=1}^M$$$, the image was computed as $$$\rho(\mathbf{x},T,f)=r(\mathbf{x})\sum_{m=1}^M \hat{\upsilon}_{m}\left(\mathbf{x},T\right) \hat{\phi}_m\left(f\right)$$$. The constraint on $$$\left\{\psi_p^{(m)}(T)\right\}_{m,p=1}^{M,P_m}$$$ (based on the biology of ischemia-reperfusion1-3) replaces the temporal smoothness regularization in our previous work, and combined with $$$r(\mathbf{x})$$$, enables us to reconstruct high SNR spatial coefficients without over smoothing.

Results

Figure 2 shows the proton reference images and some representative spectra from the reconstructions. Movies of the PCr peak integrals from the three reconstructions are shown in Fig. 3. Each reconstruction shows a gradual decrease in the peak integrals during ischemia, followed by rapid recovery during reperfusion, in good agreement with literature1-3.

Conclusions

We have improved a recently proposed low-rank tensor based method for accelerated high-resolution dynamic 31P-MRSI with application to the observation of the 31P metabolite dynamics during ischemia-reperfusion in rat hindlimbs. In vivo results obtained using a 9.4T animal scanner demonstrate a factor of 6x acceleration over the previous work, yielding an exciting combination of spatial and temporal resolution which provides the opportunity for assessing the heterogeneity of mitochondrial oxidative capacity in rodent models.

Acknowledgements

This work was supported in part by the National Institutes of Health (NIH-1R01- EB013695, NIH-R01-EB023704, NIH-R21EB021013-01, and NIH-R21HL126215).

References

1. S. Morikawa, C. Kido, and T. Inubushi, Observation of rat hind limb skeletal muscle during arterial occlusion and reperfusion by 31P MRS and 1H MRI. Magn. Reson. Imag. 1991; 9:269-274.

2. S. Morikawa, T. Inubushi, and K. Kito, Heterogeneous metabolic changes in the calf muscle of the rat during ischemia-reperfusion: in vivo analysis by 31P nuclear magnetic resonance chemical shift imaging and 1H magnetic resonance imaging. Cardiovasc. Surg. 1993; 1:337-342.

3. Y. Liu, X. Mei, J. Li, N. Lai, and X. Yu, Mitochondrial function assessed by 31P MRS and BOLD MRI in non-obese type 2 diabetic rats. Physiol. Rep. 2016; 4(15): 12890.

4. S. C. Forbes, J. M. Slade, R. M. Francis, and R. A. Meyer, Comparison of oxidative capacity among leg muscles in humans using gated 31P 2-D chemical shift imaging. NMR Biomed. 2009; 22:1063-1071.

5. R. L. Greenman, X. Wang, and H. A. Smithline, Simultaneous acquisition of phosphocreatine and inorganic phosphate images for Pi:PCr ratio mapping using a RARE sequence with chemically selective interleaving. Magn. Reson. Imag. 2011; 29:1138-1144.

6. R. Brown, O. Khegai, and P. Parasoglou, Magnetic resonance imaging of phosphocreatine and determination of BOLD kinetics in lower extremity muscles using a dual-frequency coil array. Sci. Rep. 2016; 6:30568.

7. A. I. Schmid, M. Meyerspeer, S. D. Robinson, et al., Dynamic PCr and pH imaging of human calf muscles during exercise and recovery using 31P gradient-echo MRI at 7 Tesla. Magn. Reson. Med. 2016; 75:2324-2331.

8. L. Valkovič, M. Chmelík, M. Meyerspeer, et al., Dynamic 31P–MRSI using spiral spectroscopic imaging can map mitochondrial capacity in muscles of the human calf during plantar flexion exercise at 7 T. NMR Biomed. 2016; 29:1825-1834.

9. C. Ma, B. Clifford, Y. Liu, et al., High-resolution dynamic 31P-MRSI using a low-rank tensor model. Magn. Reson. Med. 2017; 78:419-428.

10. J. He, Q. Liu, A. G. Christodoulou, et al., Accelerated high-dimensional MR imaging with sparse sampling using low-rank tensors. Magn. Reson. Med. 2016; 35(9):2119-2129.

11. F. Lam, C. Ma, B. Clifford, C. L. Johnson, and Z.-P. Liang, High-resolution 1H-MRSI of the brain using SPICE: data acquisition and image reconstruction. Magn. Reson. Med. 2016; 76(4):1059 — 1070.

Figures

Figure 1: Data acquisition began with low-resolution data, $$$\mathcal{D}_{1,b}$$$, acquired along with two proton reference scans (TR/TE1/TE2 = 500/4.1/4.6 ms). The blood-pressure cuff was then inflated, followed again by a set of reference scans. The 23 minute ischemia period was covered by a high-resolution spiral scan, $$$\mathcal{D}_{2,i}$$$. Then starting 1 minute before the cuff was deflated, another high-resolution scan, $$$\mathcal{D}_{2,r}$$$, was acquired, followed by a low-resolution scan, $$$\mathcal{D}_{1,r}$$$, and additional reference scans.

Figure 2: Center five slices of the mean anatomical reference image (a) and B0 inhomogeneity map (b) from the first rat. (c — e) show spectra over time from the locations in the reconstruction indicated by the red circle, blue square, and green triangle in the reference image respectively.

Figure 3: Movies generated from the PCr peak integrals of the center five slices of the reconstructions from the three rat data sets.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
0911