Chan Hong Moon1, Michael Schwerter2, Jullie W. Pan1, N Jon Shah2, and Hoby Hetherington1
1MRRC, Dept. of Radiology, University of Pittsburgh, Pittsburgh, PA, United States, 2Medical Imaging Physics (INM-4), Institute of Neuroscience and Medicine, Jülich, Germany
Synopsis
Dynamic shimming compared to global shimming improves
B0 homogeneity over whole brain. However, typical shim optimization
could result in high currents, particularly for small volume, e.g. small slab or slice. To
address this we investigated the extent to which large reductions in aggregate
shim current could be achieved through constrained optimization at 7T. We also
investigated the extent to which alternate imaging orientation could improve
homogeneity. The results show constrained optimization can provide robust shimming
at substantially reduced aggregate current values, and angulated imaging improved
shimming condition. The proposed methods enhanced multi-slice imaging by
reducing both susceptibility-induced signal losses and distortions.
Introduction
Typically optimization of B0 homogeneity
within a 3D volume is achieved through least square minimization of the
residual B0 field constrained only by the maximum output of the shim
current supplies. Unfortunately, with higher order shim systems (e.g., 1st-4th)
this can result in high currents for volumes of limited size, e.g., small slab
or slice. The higher currents used place greater demands on the accuracy of the
calibration of the shims, including any imperfections, thereby requiring longer
iterative approaches. Further, eddy current effects are increased when the high
valued shims are switched dynamically. To overcome this effect we investigated
the extent to which large reductions (>60%) in aggregate shim current could
be achieved through constrained optimization without significant loss in B0
homogeneity. We also investigated the extent to which alternate imaging plane
orientations could improve homogeneity.Methods
All data were acquired on a 7T Siemens whole body
system (with 1st - 2nd shims embedded) using a high
degree head shim insert coil (RRI, Resonance Research Inc., MA) providing 3rd
and 4th degree shims1. Dynamic updating of 2nd-4th
degree shims was achieved using an RRI power supply. Offsets for 1st
degree shims (gradient offset) were controlled via imaging sequence in MR console.
The shim unit and scanner were synchronized through trigger signal which
generated in the scanner sequence. Data processing and shim controls were done
by a PC with home-made MATLAB program (MathWorks, MA). The B0 field
was mapped using BOLERO with incremental B0 encoding intervals (TE =
1,2,4 and 8 ms)2. Data was acquired using an 8×2 transceiver array in pTx mode. B0 maps,
33-41 contiguous slices, 3mm thickness, matrix size 64×64 over a FOV of 192×192mm2 were acquired by using multi-echo
GRE sequence. The shim optimization was applied for 3D volume in static shimming and 2D multi-slices
in dynamic shimming. In multi-slice 2D imaging, a constrained optimization (15A
total aggregate current) was used to reduce the optimal shim currents. For
dynamic shim update, eddy currents due to dynamic updating of the shims was
reduced further by using a 2ms ramp to switch the shims. For the feasibility
test to real MR application, the developed shim methods were applied to
multi-slice 2D GE EPI with identical spatial parameters as B0
mapping sequence.Results
Fig. 1 shows the dependence of the achieved B0 inhomogeneity
(SD) to the aggregate current applied for all 2nd, 3rd
and 4th degree shims using static shimming; the unconstrained
optimizations required 42.9, 46.5, 24.1 and 8.4A of aggregate current for TL,
SCN, CBL, SMA, respectively. However, constraint of the aggregate current to
15A, results in increases of 0% - 8% for the four loci. Thus a >60%
reduction in aggregate current can be achieved with only minimal loss in B0
homogeneity. For multi-slice studies the B0 homogeneity can be optimized
by updating the values of the shims for each slab (3 adjacent 3mm slices) per slice,
allowing each slice(s) to have different values. For axial slices, B0
shimming over the temporal and inferior frontal lobe is often complicated due
to the need to simultaneously correct for different susceptibility effects
arising from both the frontal sinuses and lateral ear canals. These effects can
be decoupled from each other by angulating the slice orientation along the
temporal pole. In addition to providing an optimal view of the hippocampus
imaging, this approach places the frontal susceptibility artifact in more
superior slices. Fig. 2 shows the B0
map acquired for 36 3mm slices acquired parallel to the planum temporale. The
overall inhomogeneity across all brain pixels was measured as 11.7Hz with
dynamic shim, while it was18.9Hz under static shimming condition. Fig. 3 shows resting
state EPI data acquired by using dynamically updated and static 1st-4th
degree shims. The improvement in B0 homogeneity from the anterior
region above the sinuses reduces dropout, distortion and thus increases SNR and
temporal stability.Conclusions
Constrained optimization of high order B0
shimming can provide robust shim optimization at substantially reduced
aggregate current values (up to 70%) with minimal loss (<10%) in B0
homogeneity. Angulation of the imaging plane along the planum temporale
spatially segregates the frontal and temporal susceptibility effects enabling
improved shimming condition for the temporal lobe and inferior frontal regions
when dynamically updating the shims. Combined these methods enhance the quality
of multi-slice imaging by reducing both susceptibility induced signal losses
and distortions in difficult brain regions.Acknowledgements
The study is supported by NIH EB011639, EB009871, NS090417,
NS081772.References
1.
Pan JW, Lo KM, Hetherington HP. Magnetic resonance in medicine.
2012;68:1007-1017.
2. Hetherington H, Chu WJ, Gonen O, Pan JW, Magnetic resonance in
medicine, 2006;56:26-33.