4D-flow sequences on two 1.5T scanners from different vendors (Philips Achieva dStream and Siemens MAGNETOM Aera) were validated head-to-head. 4D-flow in a pulsatile flow phantom showed high accuracy and precision compared to reference laser particle image velocimetry for both scanners. 2D-flows in ascending and descending aorta and pulmonary trunk were compared to 57 4D-flow scans in 10 subjects. Lower bias for flow volumes was found on Aera (-4.7±12.5%) compared to Achieva (-18.0±20.1%). Kinetic energy showed lower bias for repeated examinations at the same scanner compared to different scanners the same day. 4D-flow without respiratory gating on Aera showed acceptable quality.
Experiments were performed on an Achieva dStream (Philips Healthcare, Best, The Netherlands) and a MAGNETOM Aera (Siemens Healthcare, Erlangen, Germany) 1.5T scanners. Phantom validation (Figure 1a) was performed using a setup (1) enabling measurement of a pulsatile and fully three-dimensional water flow with 4D-flow MRI and laser particle imaging velocimetry (PIV) as the reference standard. Three cardiac MRI examinations for each of the ten healthy subjects were performed: 2D-cine in long- and short-axis planes, 2D-flow of the aorta and pulmonary artery and 4D-flow using a product and a prototype sequence on Achieva and Aera, respectively. Examinations were performed consecutively on the same day on both scanners with one additional repeated scan within two weeks. Retrospective ECG-triggered 4D-flow with (Resp+) and without (Resp-) respiratory gating were acquired (Figure 2). Images were analysed using custom plug-ins based on Segment (http://segment.heiberg.se) (2).
Image quality in the 4D flow data was scored by two experienced observers, with a score of 0 being near-perfect quality and 3 the worst. Results are presented both for all acquired data and with suboptimal quality scans excluded. Second observer analysis was performed for 66 vessels from 22 4D-flow scans and the same vessels with 2D-flow.
In vivo flow was calculated from the ascending and descending aorta and pulmonary artery from 2D and 4D-flow acquisitions in identical planes. The contours from 2D-flow were transferred to 4D-flow images for stroke volume estimation. KE for all time phases and voxels in the ventricle (as delineated in the cine short-axis stack) were calculated as KE=½mv2, with voxel mass m and velocity v.
For the phantom validation, voxel-wise comparison of velocities between 4D-flow and laser PIV showed a strong correlation and low bias for both scanners (Figure 1 b,c,d). KE showed good agreement compared to laser PIV on Aera and a slight underestimation on Achieva.
In total 57 in vivo 4D-flow scans were acquired. The re-scans were performed 6±3 days after the first examination.
Average 4D-flow scan durations on Aera and Achieva were 9±3 (Resp+) / 6±2 (Resp-) minutes and 17±3 (Resp+) / 10±1 (Resp-) minutes, respectively. Image quality was better on Aera as on Achieva for both Resp+ (0.7±0.6 vs. 2.0±0.7) and Resp- (0.9±0.6 vs. 2.5±0.5) acquisitions. There was no intra-scanner difference in image quality between 4D-flow Resp+ and Resp- acquisitions. On Aera, no 4D-flow acquisitions had suboptimal image quality for analysis, but four Resp+ (31%) and eight Resp- (57%) acquisitions had suboptimal image quality on Achieva.
2D-flow before and after 4D-flow acquisitions showed low bias and variability (0.6±8.0 ml). Compared to 2D-flow, 4D-flow underestimated the flow on both scanners (Figures 3 and 4). 4D-flow Resp+ on Aera showed the lowest bias and higher precision, both compared to 4D-flow Resp- (p=0.005) on Aera and 4D-flow Resp+ on Achieva (p<0.001). Bias was not different between 4D-flow Resp+ and Resp- (p=0.87) on Achieva. Excluding acquisitions with suboptimal image quality on Achieva did not improve the bias (Figure 4).
Peak KE in systole and diastole (Figure 5) was similar on Aera (3.9±1.5 and 5.3±1.9 mJ) and Achieva (4.7±1.4 and 5.6±1.6 mJ). Mean KE from 4D-flow was lower on Aera (1.7±0.6 mJ) compared to Achieva (2.3±0.5, p<0.01).
Inter-observer variability for 4D-flow measurements (66 vessels) was 1.3±2.1 ml and for 2D-flow was 0.2±5.9 ml.
1. Töger J, Bidhult S, Revstedt J, Carlsson M, Arheden H, Heiberg E: Independent validation of four-dimensional flow MR velocities and vortex ring volume using particle imaging velocimetry and planar laser-Induced fluorescence. Magn Reson Med 2016; 75:1064–1075.
2. Heiberg E, Sjögren J, Ugander M, Carlsson M, Engblom H, Arheden H: Design and validation of Segment - freely available software for cardiovascular image analysis. BMC Med Imaging 2010; 10:1.