Berk Silemek1, Volkan Açıkel2, Uğur Yılmaz1, and Ergin Atalar1,3
1National Magnetic Resonance Research Center (UMRAM), Bilkent University, Ankara, Turkey, 2REHIS Power Amplifier Technologies, Aselsan, Ankara, Turkey, 3Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey
Synopsis
In this study, we
proposed a prototype implant that can be controlled wirelessly to change the
connection impedance between the lead and the case. To
show the effect of the connection impedance between the lead and the case, the
tip temperature increase was calculated for 3
different lead lengths by placing a capacitance between the lead and the case. Two
different case to lead connection impedance values were switched in this work
as proof of a concept and tip temperature heating was changed during the scanning. It is demonstrated that by changing the
impedance, the lead tip heating properties can be changed allowing safe
scanning. it can
be useful for avoiding excessive heating if resonance condition occurs due to
environmental factors inside human body and position inside the scanner.
Introduction
MRI scanning of
patients with active implantable medical devices (AIMDs) is not allowed because
conductive metallic wires (i.e.implant lead) inside the scanner system couples
with the high power radiofrequency (RF) field then causes excessive heating.
Acikel et al. reported a modified transmission line method (MoTLiM) to analyze
RF coupling of AIMDs with scanner1. The exact heating of an AIMD during
an MRI scan of a patient, however, cannot be calculated to this date because heating
is a strong function of the implant position in the body and the position is
not precisely controlled during implantation procedure and it is unknown during
the scanning. It is feared that with a bad lead position during MR scanning, excessive
lead tip heating may occur during scanning. Acikel et al. showed that the lead
tip heating also depends on the connection to the implantable pulse generator
(IPG) case2. The IPG case model in MoTLiM indicates that by adjusting
the impedance between the case and the lead, the resonance length of the lead
can be changed. In this study, we proposed a prototype implant that can be
controlled wirelessly to change the connection impedance between the lead and
the case. It is demonstrated that by
changing the impedance, the lead tip heating properties can be changed allowing
safe scanning. Methods
To show the
effect of the connection impedance between the lead and the case, the tip
temperature increase when there is no perfusion, was calculated for 3 different
lead lengths by placing a capacitance between the lead and the case. The value of the capacitance was swept from
1pF to 500pF and the lead tip temperature was calculated inside the medium with
relative permittivity of 65, conductivity of 0.2S/m at 63.795MHz. The wavenumber of the lead was 6.5-j1.62. The case circuit parameters from MoTLiM
model, voltage source and an impedance, was Vc=5.9exp(j151)mV, Zc=0.9580-1.2879jΩ.
To demonstrate
proposed system, we modified the previously presented Temperature Sensor
Implant (TSI) system3 to switch between capacitor values for changing
effective electrical length of the lead. The modified circuit block
diagram is shown in Fig-1. The photograph of the circuit and the case is shown in Fig-2.
To change the capacitance values, three 1-4 ultra-low power (110µA) RF switches (PE42442-Z, Peregrine
Semiconductor, California, US) were placed. In total, 10-RF switchable
capacitance nodes were designed and connected to input/output port of the
Bluetooth chip (BLE112, Bluegiga Technologies, Espoo, Finland). These switch
controls were wirelessly adjustable. Temperature measurement module, dimensions
of the implant and connectivity were designed according to previous design3.
The lead length was 45cm.
MRI experiment was performed on 1.5T scanner (Scimedix Inc., Incheon, South Korea). System’s body coil was
used. A rectangular phantom (65x40cm) filled with
hydroxyethylcellulose solution (14g/L with 1.5g/L NaCl, Total:29L) was used
with RF only high SAR sequence (with scanner reported 4W/kg SAR, total time:
15-min). TSI was placed as shown in Fig-3, 4.5cm distant from the phantom edge.
Before RF was applied to the phantom, impedance between the case and lead was switched
to the open. Then, the sequence was started and the tip temperature was
recorded for 325s. While the sequence was running the lead and the case
connection was wirelessly changed to the short circuit and the tip temperature was
recorded for another 265s. Then, the lead and the case connection was switched back to open circuit. The tip temperature
transferred to the external computer wirelessly during the scanning.
Results
The
calculations from MoTLiM model is shown in Fig-4 for lead lengths 24cm, 35cm
and 45cm. Depending on the physical length of the lead increasing the
capacitance value might either increase or decrease the temperature increase at
the tip. In Fig-5, the tip temperature is
shown during the scanning. The time points when the
connection impedance was changed are marked with arrows. Tip temperature value decreased 5°C when
the connection impedance switched from open to short circuit. It was also increased back when the
connection impedance was switched back to open circuit.Discussion and Conclusion
We
demonstrated a prototype system for wireless heating control of an AIMD during
MRI. The system changes the connection impedance between the lead and the case.
Two different case to lead connection impedance values were switched in this
work as proof of a concept and tip temperature heating was changed during the
scanning. Different conditions are investigating to understand practical
challenges. Furthermore, the simulations indicate that temperature increase may
not be decreased to zero heating, however; significant heating reduction can be
achieved using this method. As a result, it can be useful for avoiding
excessive heating if resonance condition occurs due to environmental factors
inside human body and position inside the scanner.Acknowledgements
No acknowledgement found.References
- Acikel V, Atalar E. Modeling of radio-frequency
induced currents on lead wires during MR imaging using a modified transmission
line method. Med Phys 2011;38:6623–6632.
- Acikel V, Uslubas A, Atalar E. Modeling of
electrodes and implantable pulse generator cases for the analysis of implant
tip heating under MR imaging. Med Phys 2015;42:3922–3931.
- Silemek, B., Acikel, V., Oto, C., Alipour, A.,
Aykut, Z. G., Algin, O. and Atalar, E. (2017), A temperature sensor implant for
active implantable medical devices for in vivo subacute heating tests under
MRI. Magn. Reson. Med. doi:10.1002/mrm.26914
.