Quincy van Houtum1, Dimitri Welting1, Mark Gosselink1, Christopher Rodgers2, Wybe van der Kemp1, and Dennis W.J. Klomp1
1University Medical Center Utrecht, Utrecht, Netherlands, 2University of Oxford Centre for Clinical Magnetic Resonance Research (OCMR), Oxford, United Kingdom
Synopsis
We combined 1H MRI and 31P MRSI in the liver by
integrating a 31P RF body coil in a 7T MR system with a 16 channel 31P receive
array merged with 1H fractionated dipole antennas The setup facilitates uniform
31P excitation by the body coil with high sensitivity from the receiver array,
while providing B1 shimmed proton imaging with a dipole array.
Introduction
Conventional treatment of liver metastasis often
results in palliative care as the effectiveness of the therapy can only be
monitored after several months. Metabolite
imaging with 31P MRSI is therefore an upcoming field in MRI as it can provide
insight in tumor stage and therapy effectiveness, allowing for improved
personalized treatment. However, current applications in 31P MRSI are limited
by the use of surface transceivers consequently causing small field of views.
In addition, the very high RF field inhomogeneity results in an uneven flip
angle distribution. Adiabatic pulses to improve FA distribution are SAR demanding subsequently lowering the SNR. To
overcome these limitations, body coils for uniform 31P transmit at 7T have been
proposed1,2. In this work, we demonstrate the use of the 31P body coil
with a 16 channel receiver array merged to eight fractional dipole antennas to
combine metabolic imaging with conventional MRI.
Methods
Phosphorus MRSI was performed using an in-house
designed detunable birdcage body coil integrated into
a 7T MRI system (Philips Healthcare, Best, Netherlands). A 16-channel body
array receiver tuned to phosphor at 7T was used for acquisition. To allow for simultaneous
proton imaging, eight fractionated dipole antennas were merged with the
phosphor receive array and driven by the multi-transmit system (Philips
Healthcare, Best, Netherlands)3. A male volunteer was positioned in the MR bore
with the receive array wrapped around his torso and slightly skewed to the
right-side of the patient to properly cover the liver (Figure 1). Image based
B0 and B1 shimming was performed prior to calibration of the phosphor B1+
field, which was done by a flip angle sweep with scaling factors ranging from
250 to 750. Phosphor spectra were acquired using a 3D CSI sequence with an
isotropic resolution 24mm, a matrix size of 8x8x6 and 512 samples. Other
imaging parameters were BW, 8000Hz; TR, 2000ms; TE, 0.5409; NSA, 20; scan time,
20 minutes. Proton images were acquired using a multi-slice FFE with a total duration
of 23 seconds. Other imaging parameters were TR, 10ms; TE, 4.93ms; resolution,
0.78 x 0.78 x 10mm3; field-of-view, 500 x 500 x 30mm3 and
a flip angle of 15 degree. All data was processed in Matlab 2016b and data of
the receivers was combined using WSVD4.Results
The flip angle series enables calibration of
the RF power using 17KW as peak power (Figure 2). Coil combination was possible
with the WSVD method. The integrated body coil in combination with the receive
array for 31P MRSI showed high SNR (Figure 3) throughout the body. B0 shimming
and partial volume effects were suboptimal over such large FOV as shown by the
broad lines in the spectra and leakage of phosphocreatine (PCr) through all
voxels (Figure 4). As the antennas are inherently decoupled from the 31P coils,
good 1H MRI could be obtained (Figure 3, center and background).Discussion
Patient comfort during scanning is maintained
by merging the 31P receiver array with the 1H transceiver array. Moreover,
optimal 1H and 31P MRI can be obtained in one scan session without replacing RF
coils. While we have shown a uniform transmit, high sensitive receiving merged
with a 1H imaging setup, care must be taken in optimizing scan protocols for
motion corrections and synchronized B0 shimming.Conclusion
Localized
phosphorus MRSI by simultaneous proton imaging was successfully performed
within a large field of view at 7 Tesla. Future clinical studies will benefit
from the additional metabolite information gathered during imaging.Acknowledgements
No acknowledgement found.References
1. W. J. M. van der Kemp, V. O. Boer, P. R. Luijten, B. L. Stehouwer, W. B.
Veldhuis, and D. W. J. Klomp, ‘Adiabatic multi-echo 31P spectroscopic imaging
(AMESING) at 7 T for the measurement of transverse relaxation times and
regaining of sensitivity in tissues with short T2* values’, NMR Biomed.,
vol. 26, no. 10, pp. 1299–1307, Oct. 2013.
2. L. Valkovič et al., ‘Using a whole-body 31P birdcage transmit
coil and 16-element receive array for human cardiac metabolic imaging at 7T’, PloS
One, vol. 12, no. 10, p. e0187153, 2017.
3. A. J. E. Raaijmakers et al., ‘The fractionated dipole antenna: A
new antenna for body imaging at 7 Tesla’, Magn. Reson. Med., vol. 75,
no. 3, pp. 1366–1374, Mar. 2016.
4. C. T. Rodgers and M. D. Robson, ‘Receive array magnetic resonance
spectroscopy: Whitened singular value decomposition (WSVD) gives optimal
Bayesian solution’, Magn. Reson. Med., vol. 63, no. 4, pp. 881–891, Apr.
2010.