Kamlesh Pawar1,2, Zhaolin Chen1,3, Jingxin Zhang3,4, N Jon Shah1,5, and Gary F Egan1,2
1Monash Biomedical Imaging, Monash University, Melbourne, Australia, 2School of Psychological Sciences, Monash University, Melbourne, Australia, 3Department of Electrical and Computer System Engineering, Monash University, Melbourne, Australia, 4School of Software and Electrical Engineering, Swinburne University of Technology, Melbourne, Australia, 5Institute of Medicine, Research Centre Juelich, Juelich, Germany
Synopsis
Non-Fourier encoding such as random, noiselets
and chirp encoding have demonstrated image quality improvement in accelerated
compressive sensing (CS) MRI applications. However, implementation of the
non-Fourier encoding schemes in 2D spin echo sequence limits its use in
practice, due to the fact that spin echo is inherently a slow sequence. In this
work, we present a novel implementation of chirp encoding in fast 3D gradient
echo (GRE) and MPRAGE sequences. The chirp encoding scheme is compared with
conventional Fourier encoding for compressive sensing and susceptibility
weighted imaging applications. The evaluation demonstrates that chirp encoding
is able to preserve spatial resolution better than the Fourier encoding.
Introduction
Non-Fourier
encoding schemes such as random,1 chirp 2, toeplitz 3 and noiselets 4 have
shown performance improvement in compressive sensing and parallel imaging. Despite
the advantages provided by non-Fourier encoding, there are certain practical
limitations with the non-Fourier encoding. One of the limitations is the implementation
in an inherently slow spin echo sequence. Further, non-Fourier encoding
schemes are not compatible with the turbo spin echo sequence that limits its
use for practical purposes, as the turbo spin echo sequence is normally used in
the clinical setting. An alternative is to implement non-Fourier encoding in GRE
sequences, although slice selection is one of the problems with the 2D GRE
sequences. Therefore, in this work we present an implementation of chirp
encoding for the fast 3D GRE sequence where slice selection can be
achieved by oversampling in readout direction. A 3D chirp-encoded pulse
sequence is designed using real time spatially selective RF pulses, and the
application of chirp encoding in compressive sensing (CS) and susceptibility weighted
imaging (SWI) is evaluated.
Methods
Pulse
sequence design
A
3D GRE pulse sequence is designed as shown in Fig.1, that encodes the primary phase encoding (PE) direction with the chirp and other two directions with Fourier. For N PE, a total of N different spatially selective RF excitation pulses is
required to acquire the chirp-encoded k-space data. Due to the hardware constraint
on the scanner, which limits the number of different RF pulses to 128, the RF
pulses are prepared in real time for each PE. The gradient required during the
excitation RF pulse is given by:
$$G_y= \frac{1}{\gamma \hspace{3pt} FOV_{pe} \hspace{3pt} \Delta t_{p} }$$
where γ is gyromagnetic ratio, FOVpe is field of view in PE direction, Δtp is
dwell time of RF pulse.
CS Simulation
Simulations were performed on a phantom image to determine the effect of chirp encoding in CS reconstruction. The k-space of a phantom image was under-sampled by a factor of 4 and 6, and the reconstructed images using multichannel compressive sensing (MCS)4 was evaluated. To simulate chirp-encoded data, the PE direction was encoded with chirp basis and readout was encoded with Fourier basis. The chirp basis is constructed by modulating each row of the Fourier basis with a modulation function defined as:
$$C(n) = e^{j (\Delta c \hspace{1 pt} n^2 + \Delta c)}$$
where n is phase encode number and Δc is chirp encoding factor.
Multichannel compressive sensing reconstruction
The acquired data were under-sampled using complete random under-sampling for chirp and variable density under-sampling for Fourier. The images were reconstructed with MCS4 reconstruction by solving the following optimization:
$$Min_{\widehat{x}} \hspace{3mm} \lambda_1\|\Psi \widehat{x}\|_{l_1} + \lambda_2\| TV \widehat{x}\|_{l_1} + \|Y-E \widehat{x}\|_{l_2}^2$$
where, ψ is the wavelet transform operator, TV is the total variation operator and λ1, λ2 are the regularization parameters.
Imaging Parameters
Two sequences, 3D GRE and MPRAGE were evaluated for chirp and Fourier encoding. The imaging parameters for flow compensated GRE were: TE/TR = 20/30 ms, flip angle = 100, FOV = 256x256x192 mm3 and resolution 1 mm isotropic. For the MPRAGE sequence the TE/TR = 5/2000 ms, and the other parameters were same as GRE. The data were acquired in a 3T MAGNETOM Skyra, Siemens Healthineers, Erlangen, Germany.
Results
Fig.2
shows the simulation results for the acceleration factors of 4 and 6. It is evident
form the enlarged images and line plot that chirp encoding is able to preserve
the spatial resolution better than the Fourier encoding. Fig.3 shows the result
of MCS reconstruction from the acquired in-vivo
data that is consistent with the simulation results and shows improved
resolution in the reconstructed images. Minimum intensity projection (mIP)
across 20 axial slices from fully sampled GRE dataset is shown in Fig.4,
highlighting that the veins are more prominent and sharper in the chirp-encoded
mIP image.Discussion
The
improved resolution in the chirp-encoded images can be attributed to the fact
that chirp encoding spreads the energy in the measurement domain. A variable
density under-sampling performed in the Fourier-encoded compressive sensing
acquisition results in insufficient capture of high frequency components and hence
a loss in resolution. However, for chirp encoding the energy is completely
spread out in the measurement domain. A complete random under-sampling of
chirp-encoded data results in the preservation of sufficient high frequency
components and hence provides better resolution in the reconstructed
images. Conclusion
An
implementation of the chirp-encoded MRI is presented for the 3D GRE and MPRAGE
sequences. The application of chirp encoding to CS and SWI demonstrates that
chirp encoding outperforms Fourier encoding in preserving spatial resolution
for anatomical scans and is an excellent candidate for accelerated imaging.Acknowledgements
No acknowledgement found.References
- Justin P Haldar,
Diego Hernando, and Zhi-Pei Liang. Compressed-sensing MRI with random encoding.
IEEE transactions on Medical Imaging, 30(4):893–903, 2011.
- Gilles Puy, Jose P Marques, Rolf
Gruetter, Jean-Philippe Thiran, Dimitri Van De Ville, Pierre Vandergheynst,
and Yves Wiaux. Spread spectrum magnetic resonance imaging. IEEE transactions
on medical imaging, 1(3):586–598, 2012.
- Dong Liang, Guangwu Xu, Haifeng
Wang, Kevin F King, Dan Xu, and Leslie Ying. Toeplitz random
encoding MR imaging using compressed sensing. In Biomedical Imaging: From Nano to Macro, 2009. ISBI’09. IEEE International
Symposium on, pages 270–273. IEEE, 2009.
- Kamlesh Pawar, Gary Egan, and
Jingxin Zhang. Multichannel compressive sensing MRI using noiselet
encoding. PloS one, 10(5):e0126386, 2015.