Feliks Kogan1, Evan Gregory Levine1, Akshay Chaudhari1, Uchechukwuka D. Monu1, Kevin Epperson1, Edwin Oei2, Garry Gold1, and Brian Hargreaves1
1Radiology, Stanford University, Stanford, CA, United States, 2Radiology, Erasmus MC, Rotterdam, Netherlands
Synopsis
Osteoarthritis (OA) is
commonly a bilateral disease. While long scan time and costs have precluded
separate scanning of both knees in clinical MRI, there is evidence that
bilateral examinations are beneficial for evaluation of OA changes, especially
for longitudinal studies. In this study, we demonstrate that a bilateral
coil-array setup can image both knees simultaneously in similar scan times as
conventional unilateral knee scans with comparable image quality and
quantitative accuracy. This has the potential to improve the value of MRI knee
evaluations.
Introduction
Osteoarthritis(OA) remains
a tremendous burden to society, affecting the majority of the population by age
65[1]. OA is commonly a bilateral disease[2]. While long scan time
and costs have precluded separate scanning of both knees in clinical MRI, there
is evidence that bilateral examinations are beneficial for evaluation of OA
changes, especially for longitudinal studies[3].
Simultaneous imaging of both knees without added scan time can drastically
reduce scan costs, improve patient comfort and retention, and eliminate
potential parameter/sequence differences between scans of each knee. In this work, we
describe a bilateral coil setup using two 16-channel flexible coil-arrays to
scan both knees simultaneously with similar scan time, image quality, and
quantitative accuracy compared to single knee acquisitions. Methods
All imaging experiments
were performed on a 3.0T scanner (GE Healthcare) with university IRB approval.
Hardware modifications and safety testing were performed to enable MR imaging
with two 16-channel flexible phased-array, receive only, medium sized extremity
coils (NeoCoil, Pewaukee,WI) (Fig. 1). One volunteer was
imaged to determine the noise covariance and parallel imaging geometry factors
(g-factors) of the dual-coil-array
configuration[4,5]. Ten healthy volunteers were then imaged with
both dual-coil-array bilateral knee and single-coil-array single-knee
configurations(Table 1). Two
experienced musculoskeletal radiologists compared relative image quality
between blinded and randomized image pairs acquired with each configuration.
Differences in T2 relaxation time measurements[6] between dual-coil-array and single-coil-array
acquisitions were compared to the standard repeatability of single-coil-array
measurements using a Bland-Altman analysis.Results
The noise correlation
matrix for the 32 channels in the dual-coil-array configuration showed minimal cross-talk
between the two coil-arrays (channel numbers 1-16 vs 17-32) (Fig. 2a). Mean
g-factors ranged from 1.00 for an acceleration factor, Rz=2, to 1.38
for an acceleration factor Rz=6 (Fig. 2b). Morphologic images show
similar image quality between single-coil-array (Figs. 3a,3c) and dual-coil-array
acquisitions (Figs. 3b,3d). Image quality ratings of various joint tissues by
two blinded reviewers showed mainly no difference between single-coil-array and
dual-coil-array acquisitions. Overall, 89% of the ratings were "0" or
no difference between the image pairs (95% Confidence Interval (CI): [85%,93%]).
There were no ratings of "2"(dual-coil-array much better) or
"-2"(dual-coil-array much worse). Bilateral knee MRI acquisitions
with the dual-coil-array configuration also maintained quantitative accuracy of
measured T2 relaxation times compared with single-knee,
single-coil-array acquisitions (Fig. 3e,3f). The mean difference in T2
relaxation time measurements between the dual-coil-array configuration and the
single-coil-array configuration was 0.53±0.74 ms (95% limits of agreement: [-0.93
ms,1.99 ms]) with a concordance correlation coefficient of 0.97 (95% CI: [0.96,0.98]).
This was similar to the repeatability of the single-coil-array T2
relaxation time measurements, which showed a mean difference of -0.06±1.03 ms
(95% limits of agreement: [-2.07 ms,1.96 ms]) and a concordance correlation
coefficient of 0.97 (95% CI: [0.95,0.98]). Lastly, higher acceleration factors
can also be extended across both knees while maintaining SNR to acquire
high-resolution images, which can also be, reformatted to visual multiple scan
planes (Fig 4).Discussion
This work demonstrates that
a bilateral coil-array setup can image both knees simultaneously in similar
scan times to conventional unilateral knee scans, while maintaining image
quality and quantitative accuracy. There was nominal cross-talk between each
coil-array in the dual-coil-array configuration, which created minimal coil
noise amplification (g-factor). Image SNR is inversely proportional to the
g-factor as well as the square root of the acceleration factor, R. In 3D DESS
scans, data under-sampling by a factor of 3 was offset by a threefold increase
in the number of phase encodes in the z-direction to acquire both knees. Thus,
due to the low g-factor of the dual-coil-array configuration for RZ=3
(mean g-factor=1.01), this resulted in a negligible change in image SNR
compared to fully-sampled single knee acquisitions, with only a 6 second
increase in scan time to acquire the ARC (Autocalibrating Reconstruction for
Cartesian imaging – similar to GRAPPA) calibration region. Additionally, similar
image quality was observed between dual-coil-array and single-coil-array
images, as evaluated by blinded pairwise comparison of images from the two
configurations by experienced radiologists. Further, bilateral knee T2
relaxation time mapping with the dual-coil-array approach showed similar
quantitative T2 values compared to single-knee acquisitions with a
single coil-array. Simultaneous imaging of both knees can improve the value of
knee MRI in both clinical and research settings; providing an internal control
for clinical evaluation of pathology as well as helping reduce costs and
improve continuity in research studies.Conclusion
A bilateral coil-array setup can
image both knees simultaneous in similar scan times as conventional unilateral
knee scans with comparable image quality and quantitative accuracy. Bilateral
imaging has the potential to improve the value of MRI knee evaluations.Acknowledgements
This work was funded by GE
Healthcare and National Institute of Health (NIH) grants K99EB022634, R01EB002524, R01AR0063643, and K24AR062068.References
[1] Arden N, Nevitt MC. Osteoarthritis: epidemiology. Best
Practice & Research Clinical Rheumatology 2006;20(1):3-25.
[2] Metcalfe
AJ, Andersson ML, Goodfellow R, Thorstensson CA. Is knee osteoarthritis a
symmetrical disease? Analysis of a 12 year prospective cohort study. BMC
musculoskeletal disorders 2012;13:153.
[3] Jungmann
PM, Brucker PU, Baum T, Link TM, Foerschner F, Minzlaff P, Banke IJ, Saier T,
Imhoff AB, Rummeny EJ, Bauer JS. Bilateral cartilage T2 mapping 9 years after
Mega-OATS implantation at the knee: a quantitative 3T MRI study. Osteoarthritis
Cartilage 2015;23(12):2119-2128.
[4] Uecker
M, Lai P, Murphy MJ, Virtue P, Elad M, Pauly JM, Vasanawala SS, Lustig M.
ESPIRiT--an eigenvalue approach to autocalibrating parallel MRI: where SENSE
meets GRAPPA. Magn Reson Med 2014;71(3):990-1001.
[5] Pruessmann
KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast
MRI. Magnetic Resonance in Medicine 1999;42(5):952-962.
[6] Sveinsson
B, Chaudhari AS, Gold GE, Hargreaves BA. A simple analytic method for
estimating T2 in the knee from DESS. Magn Reson Imaging 2017;38:63-70.