Matthew F. Koff1, Suryanarayanan Kaushik2, Parina H. Shah1, Erin G. Argentieri1, and Hollis G. Potter1
1Hospital for Special Surgery, New York, NY, United States, 2General Electric Healthcare, Waukesha, WI, United States
Synopsis
Multi-spectral MRI
reduces susceptibility when imaging near total joint arthroplasty (TJA). MAVRIC SL acquires 24 bins of
off-resonance data to generate images, but most implants require fewer bins. This
study uses a calibration scan, to determine an adequate number of bins, coupled
with long echo trains and variable flip angles to permit an isotropic MAVRIC SL
acquisition in clinically feasible scan times. The isotropic MAVRIC SL
acquisition showed an improved detection of the implant-bone interface
surrounding TJA, while retaining image contrast and overall SNR. An isotropic
acquisition may improve the diagnostic capability of MAVRIC SL images.
Introduction
The inadequacy of conventional MR pulse sequences near
total joint arthroplasty (TJA) has been overcome using 3D Multi Spectral
Imaging (3D-MSI) sequences such as MAVRIC SL [1] and SEMAC [2]. MAVRIC SL
samples the broadened frequency distribution by acquiring multiple
spectral-bins at frequency offset from the Larmor frequency. These bins are
then combined to yield an artifact minimized image. Currently, MAVRIC SL acquires
24 spectral bins (±12 kHz) to overcome the most severe susceptibility of
stainless steel (~3000 ppm [1]). However, for implants that are smaller or have
a weaker susceptibility, an acquisition of 24 bins is excessive and can also
lead to visible ghosting. The number of bins needed for adequate spectral
coverage of an implant may be achieved by using a short spectral calibration
scan [3]. A reduced bin acquisition and the associated savings of scan time affords
the ability to acquire an isotropic acquisition of similar coverage. In this study,
we tested the feasibility of acquiring isotropic MAVRIC SL images in a
clinically viable scan time and further studied the clinical value added by the
acquisition. Methods
This study was IRB approved
with informed written consent. 6 TJA patients were enrolled: 4M/2F, 69 ±
12y.o., 4 total hip arthroplasties (THA)/2 total knee arthroplasties (TKA). Imaging
was performed using a clinical 1.5T MR450W scanner (GE Healthcare, Waukesha,
WI) with an 8 channel receive only cardiac coil or an 8 channel T/R phased
array knee coil (Invivo, Gainesville, FL). A spectral calibration scan was initially
run to determine the number of spectral bins needed [3] (Matrix: 128x32x24, TE/TR:
7.2ms/1.4s, slice thickness: 6mm, FOV: 36-40 cm, ETL: 16, scan time: ~35s). A coronal 3D
MAVRIC SL (MVSL) (1) series was
acquired for each THA (TE/TR: 7.2ms/4s, BW:±125kHz, FOV: 36-40cm,
Matrix:512x256, Slice Thickness: 3.5mm, ETL:24) and TKA (TE: 9ms, TR: 5s, BW: ±125kHz,
FOV: 22-26cm, Matrix:512x256, Slice Thickness: 3mm3, ETL:24), in
addition to an isotropic MAVRIC (MVISO)
series (THA: TE: 6.3, TR:2-4s, Matrix:288x288, voxel size: 1.3mm3,
ETL:48, TKA: TE: 6.3, TR:3.5-4s, Matrix:192x192, voxel size: 1.3mm3,
ETL:48).
MVSL
and MVISO images were compared side-by-side and evaluated qualitatively for
lesion conspicuity, tissue contrast, signal-to-noise ratio, blurring, and
artifact pile-up, using the ordinal scale of: 0 – MVSL much better than MVISO,
1– MVSL better than MVISO, 2– MVSL similar to MVISO, 3– MVISO better than MVSL,
4– MVISO much better than MVSL. In addition, the maximal fibrous membrane
thickness at the bone-implant interface and/or volume of osteolysis was
measured when present. Image evaluation was performed by a board certified
radiologist. Frequency
tables were created to assess distribution of ordinal variables, and a Sign test
was performed to assess display preference of either the MVSL or MVISO series
(SAS V9.3, Cary, NC). Differences of fibrous membrane thickness and osteolysis
were evaluated between MVSL and MVISO, when present. Significance was set at
p<0.05.
Results
The use of a spectral
calibration scan enabled MVISO imaging with scan times similar to the product sequence
(Table 1). Representative MVSL and MVISO images of a THA and TKA shown in Figure
1, with the region of fibrous membrane formation visible only in the MVISO
dataset indicated (Fig. 2). An example of metallic debris in the soft tissue
near a THA is shown clearly in a MVISO image as compared to MVSL (Fig. 3). MVISO
images tended to have less blurring than MVSL (Score>2 for 4/6 exams, Table
1) and also tended to have similar or better tissue contrast than MVSL images (All
exams had Score ≥ 2). The pile-up artifacts present were similar for both scans
(Score=2 for 5/6 scans), as was the overall image SNR (Score=2 for 5/6 exams). These
trends were not significant. A fibrous membrane was found in 4 subjects in
MVISO images but only in 3 subjects in the corresponding MVSL images. The thickness
differences between MVSL and MVISO images were ~0.1mm. Osteolysis was detected
in 1 subject, with the MVISO images showing a 0.69 cm3 larger volume
than the MVSL images.Discussion and Conclusions
This pilot cohort
study found that isotropic MAVRIC SL images displayed clinically relevant
features that were obscured in product MAVRIC SL images. The detection and
extent of osteolysis or fibrous membrane formation can be improved with thinner
slices. Future studies will utilize a MVSIO acquisition in conjunction a
calibration scan to benefit visualization of soft tissues near orthopaedic
hardware, including spinal instrumentation. Acknowledgements
HSS has an institutional research agreement in place
with GE Healthcare.References
1. Koch KM, et al. Magn Reson
Med. 2011 Jan;65(1):71-82.
2. Lu W, et al. Magn Reson
Med. 2009 Jul;62(1):66-76.
3. Kaushik SS, et al. Magn
Reson Med. 2016 Nov;76(5):1494-1503