xSPEN is a single-shot MRI approach whose timing and pre-acquisition hyperbolic phase, endow it with exceptional resilience to offsets. We recently introduced multi-scan, phase-encoded (PE) 3D xSPEN MRI which preserves this while increasing resolution along the PE (y) and slab-selection (z) dimensions. It is here shown that parallel receivers endow this 3D approach with unprecedented PE downsampling performances. This reflects xSPEN’s unusual kernel, whose hyperbolic phase couples the directly-sampled kz information with the y coil sensors. This mitigates the artifacts associated with a highly undersampled ky axis, as demonstrated by highly accelerated in vitro and human scans.
Results and Discussion
Experimental tests of these considerations were carried out on a 3T Siemens TrioTIM® using a 32-channels head coil on phantoms and on human volunteers; the latter following suitable written consent. Figure 3 compares the effects of downsampling in TSE4 vs PE xSPEN 3D acquisitions, using water and a human head as test cases. A slab thickness FOVz=12mm with 29 kz encodings was used, achieving an oversampling factor of 1.2 along this dimension. The echo train length in PE xSPEN was thus set 29, while the TSE turbo factor was also set to 29 in order to keep similar acquisition times for both sets of tests. The Figure clearly exemplifies what would happen if scan times were reduced in these experiments from 9.6 to 1.7 min, by increasing the ky downsampling. Notice the significant artifacts that appear in 3D TSE when downsampling exceeds ≈3. By contrast PE xSPEN remains nearly artifact-free till downsampling ca. 5x for the phantom, and brain scans remain artifact-free till reaching a downsampling factor of 7. Figure 4 exemplifies another aspect of PE xSPEN’s extreme downsampling ability, by comparing its g-maps5 with those of a 2D TSE acquisition for an acceleration of 6. Figure 4b clearly demonstrates the weaker artifacts appearing in the PE xSPEN (x,y) plane upon acceleration, while Fig. 4c rationalizes this by showing the much smaller g-factors of this new scheme. Figure 4d completes this picture by showing how these g-factor maps change for different kz acquisition points; although spatially- and kz-dependent, these g-factors always remain lower than those of TSE downsampled acquisitions.Conclusions
A new, highly accelerated approach to 3D MRI acquisitions based on phase encoded xSPEN was introduced, capable of delivering sub-millimeter resolutions in whole head 3D images within 2 min, with excellent SNR and free of aliasing artifacts. Further improvements including compressed sensing, SENSE-based reconstruction and multibanding, are currently in progress.1. Zhang Z., Seginer A ., Frydman L., Single-scan magnetic resonance imaging with exceptional resilience to field heterogeneities, Magn Reson Med 2017;77: 623–634
2. Zhang Z, Lustig M. Frydman L., High resolution imaging by phase encoded xSPEN MRI, Proceedings of 25th ISMRM, Honolulu, USA, April 22-27, 2017; Zhang Z, Lustig M. Frydman L., Phase-encoded xSPEN: A novel high-resolution volumetric alternative to RARE and spin-echo EPI, Magn Reson Med 2017 (in revision).
3. Griswold, M. A., P. M. Jakob, R. M. Heidemann, M. Nittka, V. Jellus, J. Wang, B. Kiefer and A. Haase. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 2002;47: 1202-1210.
4. Yuan C, Schmiedl UP, Weinberger E, Krueck WR, Rand SD. Three-dimensional fast spin-echo imaging: Pulse sequence and in vivo image evaluation. J Magn Reson Imaging 1993;3:894-899.
5. Robson P.M., Grant A.K., Madhuranthakam A.J., Lattanzi R., Sodickson D.K., McKenzie C.A., Comprehensive quantification of signal‐to‐noise ratio and g‐factor for image‐based and k‐space‐based parallel imaging reconstructions. Magn Reson Med, 2008. 60: 895-907.