Synopsis
Several studies have reported different
findings on the correlations of the CW-T1ρ relaxation time in
articular cartilage with its different properties. Most studies agree on the
sensitivity of CW-T1ρ to cartilage proteoglycans,
although reports specifically against this also exist. Furthermore, CW-T1ρ has been connected to the collagen network properties and also
correlated with T2 relaxation time. Orientation dependence of CW-T1ρ has been reported, as well as its dependence on the spin-locking amplitude.
This study aims to combine all of these aspects in a single study.
Introduction
Continuous wave (CW)-T1ρ
relaxation time is one the candidates proposed for the quantitative assessment
of articular cartilage. Reports have demonstrated the sensitivity of CW-T1ρ to
proteoglycans, 1,2 to collagen network properties 3 and to the orientation of cartilage in the magnetic field. 3-6 Some reports have found that CW-T1ρ does not correlate with the
proteoglycan concentration, 7 and while similar with T2 relaxation time, it is not necessarily the
same. 8-10 Furthermore, orientation dependence has been reported for CW-T1ρ, and
ascribed to residual dipolar coupling 3,5 (the “magic angle effect”), reduced by increasing the spin-lock
amplitude. On the other hand, spin-lock amplitude has marked effects on the sensitivity
of CW-T1ρ to cartilage properties.
This study aimed to clarify
the correlations of CW-T1ρ with the properties of articular cartilage by
measuring the relaxation time over a broad range of spin-lock amplitudes and
over four different orientations of the tissue with respect to the main
magnetic field (B0).
Methods
Patellas were extracted from four
bovine knees and cylindrical cartilage-bone plugs (n=10) were harvested, immersed
in phosphate buffered saline and frozen at -22°C.
Biomechanical properties were measured using
a flat-ended indenter (dia.=1.0mm) and stress-relaxation and dynamic testing. 11,12 Young’s modulus at equilibrium, dynamic
modulus (@1.0Hz) and phase angle were calculated. 13,14 For MRI, the samples were immersed
in perflouropolyether (Galden HS 240, Solvay Solexis, Italy) in a holder
allowing re-orientation of the sample with respect to B0. MRI was done
at 9.4T using a 19-mm quadrature RF volume transceiver (RAPID Biomedical GmbH,
Rimpar, Germany) and a VnmrJ 3.1 Varian/Agilent DirectDrive console (Varian
Associates Inc., Pala Alto, CA, USA). Samples were scanned using an FSE readout
sequence with 1-mm slice, FOV=18x18mm, matrix (ROxPE)=256x32, ETL=8 and
effective TE=5ms, for a total time of 45 min per orientation. For each
orientation, inversion recovery T1 measurement (TR=7000ms, 6 TIs=0.2-3s); spin
echo T2 measurement (TR=5000ms, 7 TEs=10-128ms) and a CW-T1ρ scan
with 21 spin-lock amplitudes spanning the range 0-5000Hz with 6 spin-lock times
(TSL=0-128ms) was done. Furthermore, a 3-D dataset with GRE was acquired for
every orientation. The samples were measured at approximately 0°, 35°, 55° and 90° orientations. Pixel-wise relaxation time maps were calculated and ROIs were defined for
superficial, transitional and radial zones. Relaxation anisotropy was
calculated for deep cartilage. 3 Spearman correlation was used to compare the biomechanical
parameters with the relaxation times. Results
T2 and CW-T1ρ relaxation times were found sensitive to
orientation, while T1 relaxation time was not (Fig.1). Furthermore, an increase
in the T1ρ relaxation time with increasing
spin-lock amplitude was noted at all orientations (Fig.1).
The CW-
T1ρ relaxation anisotropy decreased and stabilized
after a spin-lock amplitude of approximately 1500Hz to a value of ~10%,
approaching the ~4% noted for T1 (Fig.2). Towards lower spin-lock amplitudes,
the anisotropy increased reaching ~60% at 150Hz spin-lock. Below this, an unexpected
decrease (although coincident with increasing artifacts) was noted. The T2
anisotropy was the highest at approximately 65% (Fig.2).
The average
Young’s modulus at equilibrium was 0.75±0.43MPa, dynamic modulus 4.26±1.61MPa and the phase angle 10.41°±2.70°. The
correlations between the CW-T1ρ and the biomechanical parameters had a significant
dependence on the specimen orientation, but especially on the spin-lock
amplitude (Fig.3). In
the superficial and transitional zones, equilibrium modulus and dynamic modulus
demonstrated a significant negative and the phase angle a significant positive
correlation with the CW-T1ρ relaxation time – for spin-lock amplitudes above 600-700Hz. Below 600Hz, the
correlation coefficients behaved more randomly (Fig.3). In the radial zone, the correlation coefficients
stabilized above spin-lock amplitude of ~1500Hz, and strongly depended on the sample
orientation (Fig.4).Discussion
The CW-T1ρ relaxation time appears to depend on the orientation of
cartilage with respect to B0, but also especially on the spin-lock
amplitude. From a relatively high anisotropy of ~60%, the CW-T1ρ relaxation anisotropy reached a stable level of ~10% above a
spin-lock amplitude of 1500Hz, indicating that at around this amplitude the
majority of residual dipolar coupling responsible for the orientation
sensitivity is overpowered by the spin-locking pulse, in line with the findings
of Akella 5 and Hänninen. 3 An interesting finding was also that the biomechanical
parameters correlated with the CW-T1ρ relaxation
time in the superficial and transitional zones – but consistently only above approximately
600-700Hz spin-locking amplitude. The correlation especially with Young’s
modulus at equilibrium appeared strong. All of these findings point both to the
potential of using CW-T1ρ
as a biomarker for cartilage properties, but also warrant caution with interpretation
of CW-T1ρ data. Especially attention to the used spin-locking amplitude should
be paid when interpreting CW-T1ρ results.Acknowledgements
Support from the Academy of Finland (grants
#285909, #293970 and #297033) is gratefully acknowledged.References
1. Wheaton
AJ, Dodge GR, Borthakur A, Kneeland JB, Schumacher HR, Reddy R. Detection of
changes in articular cartilage proteoglycan by T(1rho) magnetic resonance
imaging. J Orthop Res 2005;23(1):102-108.
2. Akella
SV, Regatte RR, Gougoutas AJ, Borthakur A, Shapiro EM, Kneeland JB, Leigh JS,
Reddy R. Proteoglycan-induced changes in T1r-relaxation of articular cartilage
at 4T. Magn Reson Med 2001;46(3):419-423.
3. Hanninen
N, Rautiainen J, Rieppo L, Saarakkala S, Nissi MJ. Orientation anisotropy of
quantitative MRI relaxation parameters in ordered tissue. Sci Rep
2017;7(1):9606. doi: 10.1038/s41598-017-10053-2.
4. Shao
H, Pauli C, Li S, Ma Y, Tadros AS, Kavanaugh A, Chang EY, Tang G, Du J. Magic
angle effect plays a major role in both T1rho and T2 relaxation in articular
cartilage. Osteoarthritis Cartilage 2017. doi: 10.1016/j.joca.2017.01.013.
5. Akella
SV, Regatte RR, Wheaton AJ, Borthakur A, Reddy R. Reduction of residual dipolar
interaction in cartilage by spin-lock technique. Magn Reson Med
2004;52(5):1103-1109.
6. Wang
N, Xia Y. Depth and orientational dependencies of MRI T(2) and T(1rho)
sensitivities towards trypsin degradation and Gd-DTPA(2-) presence in articular
cartilage at microscopic resolution. Magn Reson Imaging 2012;30(3):361-370.
doi: 10.1016/j.mri.2011.10.004.
7. van
Tiel J, Kotek G, Reijman M, Bos PK, Bron EE, Klein S, Nasserinejad K, van Osch
GJ, Verhaar JA, Krestin GP, Weinans H, Oei EH. Is T1rho Mapping an Alternative
to Delayed Gadolinium-enhanced MR Imaging of Cartilage in the Assessment of
Sulphated Glycosaminoglycan Content in Human Osteoarthritic Knees? An in Vivo
Validation Study. Radiology 2016;279(2):523-531. doi:
10.1148/radiol.2015150693.
8. Majumdar
S, Li X, Blumenkrantz G, Saldanha K, Ma CB, Kim H, Lozano J, Link T. MR imaging
and early cartilage degeneration and strategies for monitoring regeneration. J
Musculoskelet Neuronal Interact 2006;6(4):382-384.
9. Mlynárik
V, Trattnig S, Huber M, Zembsch A, Imhof H. The role of relaxation times in
monitoring proteoglycan depletion in articular cartilage. J Magn Reson Imaging
1999;10(4):497-502.
10. Menezes
NM, Gray ML, Hartke JR, Burstein D. T2 and T1rho MRI in articular cartilage
systems. Magn Reson Med 2004;51(3):503-509.
11. Li
LP, Herzog W, Korhonen RK, Jurvelin JS. The role of viscoelasticity of collagen
fibers in articular cartilage: axial tension versus compression. Med Eng Phys
2005;27(1):51-57. doi: 10.1016/j.medengphy.2004.08.009.
12. Makela
JT, Huttu MR, Korhonen RK. Structure-function relationships in osteoarthritic
human hip joint articular cartilage. Osteoarthritis Cartilage
2012;20(11):1268-1277. doi: 10.1016/j.joca.2012.07.016.
13. Hayes
WC, Keer LM, Herrmann G, Mockros LF. A mathematical analysis for indentation
tests of articular cartilage. J Biomech 1972;5(5):541-551.
14. Korhonen
RK, Laasanen MS, Töyräs J, Helminen HJ, Jurvelin JS. Comparison of the equilibrium response of articular
cartilage in unconfined compression, confined compression and indentation. J
Biomech 2002;35(7):903-909.