0176

Measuring MRI Gradient Trajectory Dynamics using Simultaneous EEG-FMRI
Mark Chiew1, Jostein Holmgren1, Dean Fido1, Catherine E Warnaby1, and S Johanna Vannesjo1

1Wellcome Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom

Synopsis

Simultaneous EEG-FMRI data acquisition provides an opportunity for characterization of magnetic field dynamics during imaging, by leveraging the information contained in the EEG induced gradient artefacts. Using a simple effective loop model, we reduce the complex EEG electrode geometry to small effective loops located at off-isocentre positions, which we use to fit first order (linear in space) models of the magnetic field rate of change. From these, estimates of the actual field dynamics, and trajectory/encoding information can be derived at no cost. In this proof-of-principle, we demonstrate estimation of gradient dynamics during a conventional EPI acquisition using simultaneous EEG recordings.

Purpose

Image reconstruction benefits from knowing the actual magnetic field dynamics that are present during imaging. Existing methods for capturing this information use external hardware (e.g. field cameras1), system characterization (e.g. gradient impulse responses2), or intrinsic, data-driven corrections3. In simultaneous EEG-FMRI data acquisition, external hardware is present in the form of electrodes that are distributed across the head to measure scalp potentials during FMRI. While the EEG system is not designed to measure field characteristics, magnetic field fluctuations lead to large artefacts in the EEG recordings, generated by induction effects. Instead of discarding the gradient-induced signals as artefacts, we propose to use the induced gradient waveforms to infer the actual MRI gradient dynamics, which can be used to determine actual k-space trajectories for improved image reconstruction with virtually no additional cost.

Methods

Voltages recorded on the EEG channels are:

$$V_i=-\frac{\partial}{\partial t}{\int}B{\cdot}dA_i+V^{EEG}_{i}$$

where the first term is time-varying flux changes from field dynamics, and the second is scalp potential differences. We henceforth neglect scalp potentials, which are orders of magnitude lower than the gradient-induced voltages4, and are averaged over during model fitting.

As the electrodes do not form trivial closed loops, determining the surface over which to integrate is difficult, and induction effects are not attributable to point-locations in space. Here, we propose to model the complex electrode geometries using “effective” conducting loops of infinitesimal area, with offsets from isocentre.

This assumes $$$B=G_x(t)x+G_y(t)y$$$, or that the gradient fields are exactly linear (in space and superposition), ignoring higher order fields and cross-terms. We then simplify the integral as:

$$V_i=-G_{x}'\int x{\cdot}dA_i-G_{y}'{\int}y{\cdot}dA_i$$

$$V_i=-G_{x}'{\cdot}x_i-G_{y}'{\cdot}y_i$$

where $$$G'$$$ denotes the time-derivative. The time-independent integral terms are captured by constant, electrode specific coefficients $$$x_i$$$ and $$$y_i$$$. We can associate these coefficients with the locations of effective loops that integrate to unit area. With calibration, we can determine the subject and setup-specific $$$(x_i,y_i)$$$ for each electrode.

To calibrate, we must assume knowledge of a gradient slew rate, where errors can result in scaling uncertainties in the field estimates. We used a point in the middle of a relatively long gradient slew, to avoid transient effects, and assume nominal gradient values. Our calibration model used gradients at different orientations, to fit:

$$s_i=r_{i}sin(\theta_{gradient}+\phi_i)$$

where $$$r_i=\sqrt{(G'^{2}_{x}+G'^{2}_{y})(x^{2}_{i}+y^{2}_{i})}$$$, $$$\theta_{gradient}=tan^{-1}(G'_y/G'_x)$$$, and $$$\phi_i=tan^{-1}(y_i/x_i)$$$. Once $$$(x_i,y_i)$$$ are calibrated for each electrode, a first order spatial model is used to least-squares fit for $$$G'_x(t)$$$ and $$$G'_y(t)$$$, given $$$V_i(t)$$$. Because the EEG system used samples at 5 kHz, it is not sufficient to fully capture gradient dynamics. Therefore, rather than reconstructing trajectories, we limit our comparison here to a comparison of nominal and measured $$$G'_x$$$ and $$$G'_y$$$ as a proof of principle. The simultaneous EEG-fMRI recordings were performed at 3T, using 64 radial calibration pulses, and tested on a standard EPI acquisition with a readout bandwidth of 2170 Hz/px.

Results

Fig. 1 shows the results of fitting the effective loop models. The excellent correspondence to the model (31/32 electrodes with $$$r^2{\geq}0.99$$$) validates the effective loop approximation. Fig. 2a shows the nominal x-y positions of the electrodes, contrasted with the derived effective loop positions (Fig. 2b). A greater spread of the loop locations in y, compared to x, is possibly due to EEG cabling geometries.

Fig. 3 shows first order fits for the $$$G'_x(t)$$$ and $$$G'_y(t)$$$ coefficients at different points during the EPI readout. Fig. 3a shows the fit during a negative x-gradient slew, and Fig. 3b shows a positive x-slew. The fits range from $$${0.28}{\leq}r^2{\leq}0.96$$$ with a mean of 0.65. Fig. 4 show the nominal $$$G_x$$$, $$$G_y$$$ time-courses during an EPI readout, and associated nominal $$$-G'_x$$$ and $$$-G'_y$$$ time-courses. The EEG-derived $$$-G'_x$$$ and $$$-G'_y$$$ coefficients are plotted for comparison. Fig. 5 shows a zoom in on the $$$-G'_x$$$ estimates, highlighting the intrinsic limits of the 5 kHz EEG sampling. However, the correspondence to the nominal $$$-G'_x$$$ is unmistakable.

Discussion

We demonstrated the potential utility of using EEG-derived measures to characterize MRI gradient dynamics in a proof-of-principle experiment. These measures come at essentially no cost with EEG-FMRI data acquisition (aside from a calibration procedure), as they are derived directly the EEG gradient artefacts. While the characterizations may not be as precise as those derived from field probes, they may aid in more limited scopes, such as for EPI Nyquist ghost correction. Capturing the full bandwidth of gradient dynamics requires overcoming the inherent sampling limits of the EEG system. One potential solution may be to jitter the TE of the EPI readout in steps of the gradient raster time, so that a higher effective sampling bandwidth can be achieved by combining data across multiple measurements.

Acknowledgements

M.C. is supported by the UK Royal Academy of Engineering, S.J.V. is supported by a Marie Curie Fellowship

References

1. Barmet, C., Zanche, N. D., & Pruessmann, K. P. (2008). Spatiotemporal magnetic field monitoring for MR. Magnetic Resonance in Medicine, 60(1), 187–197.

2. Vannesjo, S. J., Haeberlin, M., Kasper, L., Pavan, M., Wilm, B. J., Barmet, C., & Pruessmann, K. P. (2013). Gradient system characterization by impulse response measurements with a dynamic field camera. Magnetic Resonance in Medicine, 69(2), 583–593.

3. Deshmane, A., Blaimer, M., Breuer, F., Jakob, P., Duerk, J., Seiberlich, N., & Griswold, M. (2016). Self-calibrated trajectory estimation and signal correction method for robust radial imaging using GRAPPA operator gridding. Magnetic Resonance in Medicine, 75(2), 883–896.

4. Lemieux, L., Allen, P. J., Franconi, F., Symms, M. R., & Fish, D. R. (1997). Recording of EEG during fMRI experiments: Patient safety. Magnetic Resonance in Medicine, 38(6), 943–952.

Figures

Figure 1 – Sinusoidal model fits of the effective loop model across all 32 electrodes (include ECG). All fits were r2 > 0.99 except for Oz, which was r2 = 0.96 (and is the electrode with effective distance closest to isocentre).

Figure 2 – (a) Nominal relative physical co-ordinates for the 31 (excluding ECG) electrodes. (b) The effective loop coordinates for the same electrodes, distributed in the x-y plane (with calibration-dependent scaling). Following the calibration procedure, we can simply consider the induced signals from each electrode as coming from a point-like pickup loop at its respective effective position in the x-y plane.

Figure 3 – First order (linear in space) field coefficient fits at two points during an EPI readout train. (a) During a negative x-gradient slew, (b) during a positive x-gradient slew.

Figure 4 – Comparison of estimated dGx/dt and dGy/dt coefficients over time during an EPI readout train. Rows 1-2: nominal Gx(t) trajectory (0.01 ms resolution), and the nominal -dGx/dt below (0.01 ms resolution). Row 3: the measured -dGx/dt coefficients (orange circles, 0.2ms resolution), showing a good correspondence to the nominal coefficients, but also showing some large transient effect at the beginning of the readout. Rows 4-6: the same as top, but for Gy and -dGy/dt

Figure 5 – A zoom in on the nominal and measured –dGx/dt, in a 5 ms window during the EPI readout. Here we can clearly see the limitation of the 5 kHz EEG sampling rate, but the correspondence of the EEG-derived estimates (orange circles) to the nominal waveform is apparent (blue line), where a sampled version of the nominal waveform is provided for reference (open blue circles). Here the measured data were scaled manually to provide the best visual correspondence.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
0176