0174

An empirical $$$B^-_1$$$ Non-uniformity Correction of Phased-Array Coil Images without Measuring Coil Sensitivity
Frederick C Damen1 and Kejia Cai1

1Radiology, University of Illinois at Chicago Medical Center, Chicago, IL, United States

Synopsis

Radio Frequency (RF) receiving coil arrays improve the signal-to-noise ratio (SNR), and enable partial parallel imaging. However, these benefits often come at the cost of image non-uniformity. $$$B^-_1$$$ non-uniformity correction techniques are confounded by signal that not only varies due to coil induced $$$B^-_1$$$ sensitivity, but also due to true signal variations in proton density, susceptibility, and relaxation rates. Herein, we propose an empirical method that produces a $$$B^-_1$$$ non-uniformity-corrected complex image from the phased-array coil images themselves using minimal assumptions and without measuring the coil sensitivities. This method is validated using MRI of the abdomen, brain, and a homogeneous phantom.

Introduction

Radio Frequency (RF) phased array receiving coils help to improve the signal-to-noise ratio (SNR), enabling partial parallel imaging[1-4]. These benefits come at the cost of image non-uniformity. $$$B^-_1$$$ non-uniformity correction techniques are confounded by signal that not only varies due to coil induced $$$B^-_1$$$ sensitivity, but also due to true signal variations in proton density, susceptibility, and relaxation rates[5-7]. Herein, we propose an empirical method that makes minimal assumptions and uses only the coil images themselves to produce a $$$B^-_1$$$ non-uniformity-corrected complex image. This method was validated with two anatomical imaging cases: (a) abdominal imaging that is challenging in both extreme variation in coil sensitivity and intrinsic signal, and (b) less challenging brain imaging.

Methods

The proposed correction method involves two main tasks: (a) estimating the spatial $$$B^-_1$$$ sensitivity for each coil element, and (b) correcting the coil images and combining them into a complex image.

(a)

Image voxels used in the estimation process were selected by thresholding a Z-score map (i.e., convolution using 5x5 kernel of mean/standard deviation) of the complex MRI coil image.

In the sensitivity model used, the recorded signal $$$r_c(x)$$$, for coils c, is the sum of the true signal $$$t(x)$$$ modulated by a smoothly varying coil sensitivity $$$b_c(x)$$$ and additive noise $$$n_c(x)$$$, i.e., $$$r_c(x) = t(x)b_c(x)+n_c(x)$$$. This model was transformed into two jointly additive models of magnitude ($$$m_c(x)=\ln \left|r_c(x)\right|$$$) and phase ($$$p_c(x)=\arg (r_c(x))$$$).

The coil sensitivity factor $$$\hat{b}_c(x)$$$ was estimated by performing $$$2^{nd}$$$ order 2D polynomial fit for $$$\hat{m}_c(x)$$$ and a $$$1^{st}$$$ order 2D polynomial fit for $$$\hat{p}_c(x)$$$, using singular value decomposition (SVD)[8,9,10], where $$$\hat{b}_c(x) = \exp(\hat{m}_c(x) + i\hat{p}_c(x))$$$.

Correction factor for each individual coil image $$$\alpha_c(x) = 1/\hat{b}_c(x)$$$.

(b)

Due to $$$\alpha_c(x)$$$ being determined on a per coil basis, $$$\alpha_c(x)$$$ is adjusted to normalize the corrections to a common reference; only Z-score selected voxels are used to determine this normalization. The true signal was estimated $$$\hat{t}(x)=\sum_c w_c(x) \alpha_c(x) r_c(x)$$$, where $$$w_c(x)$$$ is a Fermi filter used to remove heavily amplified voxels with very low SNR.

Improved corrections can be made by reiterating the aforementioned steps a and b, by first removing $$$\hat{t}(x)$$$ and $$$\hat{b}_c(x)$$$ from the processed signal, i.e., $$$r_c(x) = r_c(x)/\hat{t}(x)/\hat{b}_c(x)$$$.

Image Acquisition: Abdominal and brain MRI datasets were acquired on an AllTech EchoStar Spica 1.5T whole-body MRI system (AllTech Medical System, Solon, Ohio, USA). For abdominal MRI, a 15-channel body coil was used with a gradient echo sequence and parameters: TR/TE = 190/4.5 ms, FOV = 380x440 mm2, matrix size = 256x256, and GRAPPA factor = 2. For brain MRI, a 10-channel head coil was used with a fast spin echo sequence and parameters: TR/TE = 480/12 ms, FOV = 380x440 mm2, matrix size = 256x256, GRAPPA factor = 2. In addition, a spherical phantom (11.3 cm in diameter) with homogeneous water solution was scanned on a 3 Tesla GE MR750 (GE Healthcare, Waukesha, WI) MRI scanner using an 8-channel head coil and a spoiled gradient echo sequence with parameters TR/TE = 100/7.3 ms, FOV = 180x180 mm2, matrix size = 256x256.

For all three experiments, the conventional sum-of-squares (SoS) method for combining phase array coil images was used for comparison. In addition, Percent Image Uniformity (PIU) [11] was performed to quantitatively compare the results from phantom imaging.

Results

The Z-score map and a range of selection masks are shown in Figure 1. A threshold of 3.5 provides a balance between spatial coverage and anomalous voxel rejection.

For abdominal imaging, the original magnitude images, and the input/output images from each iteration of the algorithm are presented in Figure 2. We observed progressively improved image uniformity. Particularly, for the phase images, the first iteration successfully abated the branch cut issue as noted from coils 1, 3, 7, and 13. Our experience suggested that 2 to 3 iterations were adequate for correcting $$$B^-_1$$$ non-uniformity in both magnitude and phase images.

The corrected and combined images are presented in Figure 3a (magnitude) and Figure 3b (phase). For magnitude comparison, SoS image (Figure 3c) shows clear spatial non-uniformity.

The human brain images demonstrated similar results as shown in Figure 4.

For phantom imaging, Percent Image Uniformity of the spherical phantom using the proposed method was 96.4%, which was higher than 84.4% from SoS as shown in Figure 5.

Discussion and Conclusion

As demonstrated with abdominal, brain, and phantom MRI imaging, the empirical method we developed can correct $$$B^-_1$$$ non-uniformity from phased-array coil images, without measuring coil sensitivity, clearly evident when using the conventional SoS method. In addition, the proposed method also provides us additional and potentially useful phase images that can not be provided by SoS.

Acknowledgements

We acknowledge the NIH grant supports (R21 EB023516). We would like to thank the support from the Center for Magnetic Resonance Research at UIC.

References

1. Roemer P, Edelstein W, Hayes C, Souza S, Mueller O. The NMR phased array. Magn Reson Med 1990;16(2):192-225.

2. Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med 1997;38(4):591-603.

3. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 2002;47(6):1202-1210.

4. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med 1999;42(5):952-962.

5. Balafar M. Review of Intensity Inhomogeneity Correction Methods for Brain MRI Images. Issue;13:60-66.

6. Hou Z. A review on MR image intensity inhomogeneity correction. International Journal of Biomedical Imaging 2006;2006.

7. Vovk U, Pernus F, Likar B. A review of methods for correction of intensity inhomogeneity in MRI. Medical Imaging, IEEE Transactions on 2007;26(3):405-421.

8. Moore EH. On the reciprocal of the general algebraic matrix. Bulletin of the American Mathematical Society 1920;26(9):394–395.

9. Penrose R. A generalized inverse for matrices. Proceedings of the Cambridge Philosophical Society 1955;51:406–413.

10. Golub GH, Van Loan CF. Matrix computations: JHU Press; 2012.

11. Goerner FL, Duong T, Stafford RJ, Clarke GD. A comparison of five standard methods for evaluating image intensity uniformity in partially parallel imaging MRI. Med Phys 2013;40(8):082302.

Figures

Figure1: The demonstration of Z-score map with different thresholds for the selection of voxels of interest. a) computed Z-score map. Z-score selected voxels (black) at threshold of 1 (b), 5 (c), and 3.5 (d). A z-score threshold of 3.5 was empirically set in order to balance spatial coverage and anomalous voxels selected.

Figure 2: Abdominal coil images (1 to 15 rows) that depict the $$$B^-_1$$$ non-uniformity correction progress during correction iterations. Each row represents a single coil element. Column M is the original magnitude image. Columns M1 through M3 are the (natural) log-magnitude inputs to the next iteration of the model estimation; they are also coincidentally the outputs from the previous iteration. Similarly, columns P1 through P3 are the phase inputs to the next iteration of the model estimations. Each column is windowed/leveled the same. Blue color-coded voxels were masked out using Z-score maps.

Figure 3: Resultant combined output images of the abdomen: a) magnitude; b) phase. The traditional SoS image is shown in c).

Figure 4: Resultant combined output images of the head: a) magnitude; b) phase. The traditional SoS image is shown in (c).

Figure 5: Resultant combined magnitude images of a spherical phantom (a) and the traditional SoS image (b). The corresponding histograms (b, d) demonstrating better uniformity from the proposed method (b) than SoS (d).

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
0174