0065

Clinical Imaging Potential of FRONSAC
Nadine Luedicke Dispenza1, Sebastian Littin2, Maxim Zaitsev2, R. Todd Constable3,4, and Gigi Galiana3

1Department of Biomedical Engineering, Yale University, New Haven, CT, United States, 2Department of Diagnostic Radiology, Medical Physics, University Medical Center Freiburg, Freiburg, Germany, 3Department of Radiology and Biomedical Imaging, Yale University, New Haven, CT, United States, 4Department of Neurosurgery, Yale University, New Haven, CT, United States

Synopsis

Despite potential for more flexible and efficient encoding that better complements receiver geometry, the past decade of work with nonlinear gradients (NLGs) has shown relatively modest improvements on accelerated image quality. In this work we present the first experimental evidence that the previously introduced ROtary Nonlinear Spatial ACquisition (FRONSAC) can notably improve accelerated image quality, both in vitro and in humans. Furthermore, this work introduces and demonstrates a number of robust and flexible attributes of this method, which are crucial to reducing scan times in a clinical setting.

Introduction

In this work we demonstrate in vivo and in vitro that the application of sinusoidal NLGs improve undersampled Cartesian encoding (Figure 1a). Linear gradients sample k-space one point at a time, and receiver arrays sample a static distribution of k-space. In contrast, NLGs dynamically vary the size and shape of the sampling functions in k-space (Figure 1c), which provides more degrees of freedom to design an efficient trajectory to measure k-space. The encoding efficiency is increased when an undersampled linear acquisition is enhanced with FRONSAC encoding as measured with a previously introduced k-space metric (Figure 1d) (1). Previous work showed that FRONSAC reduces undersampling artifacts in accelerated imaging for a variety of linear gradient sequences (2). In the current work presented here, we show the first experimental Cartesian FRONSAC results, confirming the theoretical reduction of under sampling artifacts and we investigate features of FRONSAC critical to clinical imaging. In addition to providing experimental confirmation of the reduced undersampling artifact, we demonstrate that FRONSAC yields more benefit from the new generation reconstruction algorithms, now becoming commercially available (3-5). We also show that Cartesian FRONSAC yields benign artifacts in the presence of experimental imperfections or when the gradient fields deviate from the perfect spherical harmonics used in the original paper, thereby providing robust performance in a clinical setting. A single FRONSAC gradient is also shown to enhance undersampled image quality for nearly any imaging prescription, despite changes in the linear gradients associated with different image dimensions, slice orientation or spatial resolution. As such, the method does not require extensive field mapping. It is applicable to a number of different sequence types and is demonstrated here using both gradient echo and fast spin echo sequences.

Methods

In vivo and phantom experiments on a 3T MRI scanner (MAGNETOM Trio Tim, Siemens Healthcare, Erlangen, Germany) with an 8 channel RF head coil (Siemens) were performed at Yale University using a NLG insert (Tesla Engineering Ltd, Storrington, UK) (Figure 1b) that generates 3 spherical harmonic gradient fields: GC3 (X3-3XY2), GS3 (3YX2-Y3), and GZ2 (X2+Y2) with maximum strengths of 3254.8 mT/m3, 3155.4 mT/m3 and 475.08 mT/m2. Phantom experiments performed at the University Medical Center Freiburg used an 84 channel matrix gradient coil (6). A non-optimized cluster of elements approximated a C3 field with maximum strength of 452 mT/m3. Transverse slices at isocenter with a FOV of 250 mm were acquired with 1024 readouts with 128 excitations. GRE imaging parameters at Yale University were: TR=1000 ms/TR=600 ms; TE=18 ms; bandwidth=50 Hz/pixel; flip angle=30°/15°; slice thickness=3 mm, maximum C3/S3/Z3 strength=325.3 mT/m3, 316.7 mT/m3 and 41.6 mT/m2 with oscillation frequency of ω0/2pi=3.2 kHz. TSE imaging at Yale University parameters were: TR=3000 ms; turbo factor=8; echo spacing=24 ms; bandwidth=100 Hz/pixel; slice thickness=5 mm, maximum C3/S3/Z2 strength=390.7 mT/m3, 380.1 mT/m3 and 50.0 mT/m2 with oscillation frequency of ω0/2pi=4.8 kHz. University of Freiburg Medical Center parameters were: TR=700 ms; TE=11.2 ms; bandwidth=78.125 Hz/pixel; flip angle=20°; slice thickness=5 mm, maximum C3 strength=293.9 mT/m3 with oscillation frequency of ω0/2pi=5 kHz. Undersampling in the phase encoding direction was performed after acquisition during the image reconstruction by discarding echoes. All calculations were performed in MATLAB (MathWorks Inc, Natick, Massachusetts, USA). All reconstructions were performed via a conjugate gradient algorithm with 10 iterations using the GPU.

Results

Figure 2a shows improving image quality as function of increasing NLG waveforms added to a Cartesian trajectory. The FRONSAC encoding reduces the undersampling artifact, particularly in the presence of 2 or more waveforms, which provide more degrees of freedom to manipulate the sampling function. Figure 2b shows the addition of NLGs in FRONSAC2 better condition the data for a compressed sensing approach, while no benefit is seen in under sampled Cartesian data. FRONSAC is resilient to experimental imperfections including timing errors (Figure 3a) and off-resonance spins (Figure 3b), and the resulting artifacts are indistinguishable from well-known and generally benign Cartesian artifacts. Field purity is also not required with this approach since the reduction in undersampling artifacts using an approximate C3 shape is similar to comparable pure C3 FRONSAC1 (Figure 3c). Figure 4 shows that once a FRONSAC NLG waveform has been well characterized by spatial and temporal mapping, it can be applied to a variety of different Cartesian scan prescriptions and still produce profound improvements in accelerated image quality. Figure 5 demonstrates that FRONSAC is synergistic with other acceleration strategies such as turbo spin echo (TSE) imaging and FRONSAC gradients do not interfere with the standard contrast.

Conclusion

FRONSAC represents highly effective general tool for increased spatial encoding efficiency that can lead to significant further reductions in scan times.

Acknowledgements

This works was funded by the National Institutes of Health under 5R01EB012289-04, 4R01 EB016978-04, K01EB168977 and R01EB022030. We would like to thank Andrew Dewdney, Keith Heberlein, and Rodney Mick from Siemens for hardware support. We would like to thank Stefan Kroboth for his help imaging with the Freiburg nonlinear gradient setup and Dr. Ying-Hua (Eva) Chu for providing the field camera and for support with the corresponding measurements. We would also like to thank Terry Nixon and Scott McIntyre for support with the Yale nonlinear gradient hardware.

References

1. Dispenza NL, Tagare H, Galiana G, Constable RT. Trajectory Design of Optimized Repeating Linear and Nonlinear Gradient Encoding using a k-space Point Spread Function Metric.

2. Wang H, Tam LK, Constable RT, Galiana G. Fast rotary nonlinear spatial acquisition (FRONSAC) imaging. Magn Reson Med 2016;75(3):1154-1165.

3. Otazo R, Kim D, Axel L, Sodickson DK. Combination of compressed sensing and parallel imaging for highly accelerated first-pass cardiac perfusion MRI. Magn Reson Med 2010;64(3):767-776. 4. 4. Fessler JA, Sutton BP. Nonuniform fast Fourier transforms using min-max interpolation. IEEE Trans Signal Process 2003;51(2):560-574.

5. Lustig M, Donoho DL, Santos JM, Pauly JM. Compressed Sensing MRI. IEEE Signal Process Mag 2008;25(2):72-82.

6. Littin S, Jia F, Layton KJ, Kroboth S, Yu H, Hennig J, Zaitsev M. Development and implementation of an 84-channel matrix gradient coil. Magn Reson Med 2017:doi:10.1002/mrm.26700.

Figures

Fig . 1 | (a) FRONSAC adds small, sinusoidal, nonlinear gradients to the readout of a pulse sequence. (b) A coil in the magnet generates NLGs (c) Cartesian linear modulation in image space corresponds to a sampling point translated through k-space. FRONSAC image space modulation corresponds to a rotating and translating stamp sampling of k-space. (d) k-space coverage is calculated as the width of a PSF in k-space reconstructed from the encoding. An R=8 Cartesian sequence does not fully solve for points in the gaps of k-space, whereas the addition of a FRONSAC3 gradient greatly improves this sampling.

Fig.2 | (a) As the number of nonlinear gradient waveforms is increased from 1 to 3 (FRONSAC1 to FRONSAC3) the image quality improves. Improvements in image quality are especially apparent with the out of phase application of two or more FRONSAC waveforms (b) Compressed sensing does not help remove artifacts from the Cartesian case, but when applied to FRONSAC2 image reconstruction it further removes ghosting artifacts such as the tubes (blue arrow) and edge of the phantom (orange arrow).

Fig . 3 | Because FRONSAC gradients provide a small perturbation to the encoding function, artifacts remain Cartesian-like and benign. For example, (a) the magnitude image is not degraded by timing delays in the linear gradients, and (b) off-resonance spins cause a simple shift in the read direction, as seen in the shifted skull of this brain image. (c) For phantom images at R=6, applying either a pure or approximate C3 shape results in reduced ghosting artifacts inside the tubes (insets) and at the edge of the phantom (arrow), which distinguishes the method from nonlinear gradient methods such as PatLoc.

Fig . 4 | A single FRONSAC waveform improves many scan prescriptions. While routine changes in scan prescription (orientation, field of view, or resolution) alter the linear gradients, FRONSAC gradients provide marked improvements without requiring change. This means that the required field mapping can be performed for a single waveform and then used to improve a wide range of scans, which is a critical feature for clinical application. Note R=6 in this figure is an approximation signifying 12/64, 23/128, and 45/256 measurement lines.

Fig . 5 | FRONSAC is applied to a turbo spin echo sequence with an echo train length (ETL) of 8 and effective echo times shown above each column. To calculate the actual scan time reductions the ETL is multiplied with the undersampling factor, yielding scan accelerations of 32 and 64 for these images. FRONSAC images maintain identical contrast to the Cartesian images at each effective echo time.

Proc. Intl. Soc. Mag. Reson. Med. 26 (2018)
0065