Rosa Sanchez Panchuelo1, Robert Turner1,2,3, Olivier Mougin1, and Susan Francis1
1Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, United Kingdom, 2Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany, 3University of Amsterdam, Amsterdam, Netherlands
Synopsis
The present study uses an
efficient 2D, multi-shot, inversion-recovery EPI (MS-IR-EPI) acquisition that
combines separately excited k-space segments after each inversion pulse together
with steps in slice ordering to generate T1-maps with high SNR per
unit time. We show that although the inversion times systematically vary across
slices, consistent T1-maps can be generated across the whole brain. Such
T1-maps provide high spatial resolution and SNR, with little image
distortion and can be collected in a short acquisition time.
Introduction
Cortical parcellation for
unambiguous in-vivo identification of
distinct cortical areas has been performed using high resolution MRI that provides
some details of myeloarchitecture [1-3]. Quantitative T1 maps provide
evidence of heavily myelinated layers [4] and can easily be obtained with high
spatial resolution at 7T [5]. However, the low flip-angle 3D FLASH-type sequences
which are commonly used for T1 mapping have poorer SNR per unit time
compared with techniques such as EPI and GRASE that use high flip angles [6]. Further,
when inversion pulses are efficiently included in 3D sequences, the point
spread function (PSF) becomes broad, because the acquisition window is wide
compared with T1.
An alternative approach is to use
2D inversion-recovery EPI which has high SNR per unit time and a short echo
train length compared with T1 [7]. However, if each slice in a multislice
brain volume is required to have the same inversion time, the scan time becomes
long, especially if segmentation is added to improve spatial resolution,
because the longitudinal magnetization must recover after each inversion pulse.
Removing these requirements can greatly accelerate data acquisition, but
entails the computation of T1-maps for each slice using different
inversion times. Here we develop an efficient 2D, multi-shot,
inversion-recovery EPI (MS-IR-EPI) acquisition combining separately excited
k-space segments after each inversion pulse and shifted slice ordering to
provide high SNR per unit time.Methods
A 2D
MS-IR-EPI sequence 0.5x0.5mm2 in-plane resolution (TE=20ms, TR=3.2s,
EPI-factor=17, 12 shots, field-of-view (FOV)=180(AP)x154(RL), SENSE=1.5(RL), 38.4s acquisition time)
was evaluated at 7T for T1-mapping. Data sets consisting of 24 axial
slices (1.5mm slice thickness) were acquired with varying slice ordering schemes, thus
providing a number of equally spaced inversion times for each slice (Fig.1). Data
was acquired at a number of slice offsets=[0, 4, 5, 6, 8, 10, 12, 15, 16, 18,
20] and used to reconstruct T1-maps based on 2, 3, 4, 5 or 6 slice
offsets and thus range of inversion times (TI). Data sets were acquired with
and without fat suppression. In addition, a single slice dataset was acquired
using standard IR-EPI at 10 TIs=[100, 200, 400, 600, 900, 1100, 1300, 1600,
2100, 3100], TR=5s. To compute T1-maps, the polarity of the modulus
data was corrected (based on phase) and fitted for T1 and S0:
S(TI) = S0[1-2exp(−TI/T1)+exp(-TR/T1)].
For the MS-IR-EPI, Monte Carlo simulations were performed to evaluate the accuracy
and variance of the fitted T1s with different numbers of TIs and
different level of noise (1024 repeats). High spatial resolution (0.5 mm
isotropic) data were collected on two subjects using a MS-IR-EPI (48 slices, 4 TIs/slice
offsets each with 3 averages, 461s total acquisition time) and MP2RAGE sequence
(TR/TE=14/6.4ms, TI1/TI2=680/2080ms, TRshot=3.5s,
SENSE=2 (RL), 724s acquisition time) with matched FOV but 100 slices coverage. A
PSIR-reconstruction [8] of the MP2RAGE data was used to generate images
proportional to T1.Results and discussion
T1-maps
from the MS-IR-EPI sequence were homogeneous across slices, despite each slice
being fit with different inversion times (Fig.2A), and were independent of the
number of TIs when more than 2 inversion times were used in the fit (Fig.2B). Fitted
T1-values were shorter than those obtained with a single slice standard
IR-EPI sequence (Fig.2B). This difference was reduced when collecting data with
no fat suppression, suggesting this discrepancy is likely due to magnetization
transfer effects [9]. Future work will explore the use of different fat
suppression techniques while reducing the impact on fitted T1-values.
A Monte Carlo simulation showed that the accuracy of the fitted T1
is independent of the number of TIs used in the fit whilst the standard
deviation decreases as the number of TIs increases (Fig.3A), and as the SNR
increases (data not shown). Although the fitted T1 remains relative
constant across slices (Fig.2B) there is variation across slices when using 2 and
3 TIs for WM (Fig.3B) and 2 TIs for GM (Fig.3C). Figures 4 and 5 compare the 0.5mm
isotropic acquisition T1-maps generated from the MS-IR-EPI (no fat
suppression) with MP2RAGE processed images. Both methods yield similar SNR, but
note that while the MS-IR-EPI is based on a true T1-fit, MP2RAGE is
currently proportional to T1, and future work will generate a look-up
table to estimate T1 [5]. Figure 5 shows that the 2D MS-IR-EPI T1-maps have minimal image distortion (B) and are sharper than 3D MP2RAGE maps (Fig.5C). Conclusion
High quality T1-maps
can be generated using a high resolution 2D MS-IR-EPI sequence. Future work
will combine the MS-IR-EPI scheme with simultaneous multi-slice excitation (SMS) to
achieve whole brain coverage in short acquisition times.Acknowledgements
This work was funded by a MRC grant MR/M022722/1
and a Leverhulme fellowship.References
[1] Geyer S, Weiss M, Reimann et al. Microstructural Parcellation of
the Human Cerebral Cortex – From Brodmann's Post-Mortem Map to in vivo Mapping with High-Field
Magnetic Resonance Imaging Front Hum Neurosci. 2011;18:5-19. [2] Sánchez-Panchuelo
RM et al, Francis S, Schluppeck D & Bowtell R. Correspondence of human visual areas identified using functional and
anatomical MRI in vivo at 7 T. J
Magn Reson Imaging. 2012;35(2):287-99 [3] Glasser MF, Coalson TF, Robinson EC
et al. Nature. A multi-modal parcellation
of human cerebral cortex. 2016;536(7615):171-178. [4] Stüber C, Morawski M,
Schaeffer A et al. Myelin and iron
concentration in the human brain: A quantitative study of MRI contrast. Neuroimage.
2014;93(Pt 1):95-106 [5] Marques J, Kober T, Krueger G et al. MP2RAGE, a self
bias-field corrected sequence for improved segmentation and T1-mapping at high
field. Neuroimage. 2010;49(2):1271-81.
[6] Trampel R, Reimer E, Hueber L, et al. Anatomical
brain imaging at 7T using two-dimensional GRASE. Magn Reson Med. 2014;72(5):1291-301
[7] Mansfield P, Guilfoyle DN, Ordidge RJ and Coupland RE. Measurement of T1 by
echo-planar imaging and the construction of computer-generated images. Phys Med
Biol. 1986;31(2):113-24. [8] Mougin et al. Abdel-Fahim R, Denin R et al. Imaging grey matter with concomitant null
point imaging from the phase sensitive inversion recovery sequence. MRM.
2016;76(5):1512-1516. [9]. Wright ), Mougin O, Totman JJ, et al. Water proton
T1 measurements in brain tissue at 7, 3, and 1.5T using IR-EPI, IR-TSE, and
MPRAGE: results and optimization. MAGMA. 2008;21:121.