Richard B Thompson1, Kelvin Chow2, and Justin Grenier1
1Department of Biomedical Engineering, University of Alberta, Edmonton, AB, Canada, 2Cardiovascular MR R&D, Siemens Medical Solutions USA, Inc., Chicago, IL, United States
Synopsis
Quantitative
multi-parametric MR imaging is an important component of diagnosis and
objective staging of diffuse liver disease, reducing the need for biopsy. The goal of the current study was to validate
a new simultaneous proton density fat fraction and water T1-mapping approach
(PDFF-T1) with low B1+ sensitivity in a short,
patient-friendly breath-hold acquisition. It is shown (simulations, phantoms,
in-vivo) that a time-varying flip angle excitation scheme leads to improved
accuracy of PDFF and that fat-water separation enables water-specific T1
mapping, all in a single patient-friendly breath-hold, with low sensitivity to
inhomogeneity in the transmitted radiofrequency (B1+) field.
INTRODUCTION
Quantitative
multi-parametric MR imaging is an important component of diagnosis and
objective staging of diffuse liver disease, cirrhosis and other precursor
diseases, reducing the need for biopsy. (1) Imaging
of fat content and tissue fibrosis (via T1 quantification) in particular
provide complimentary information. Currently,
fat fraction and T1 mapping require separate acquisitions, potentially with an
additional separate radio-frequency transmit (B1+)
field map to account for excitation inhomogeneity. The goal of the current study was to validate
a new simultaneous proton density fat fraction and water T1-mapping approach (PDFF-T1)
with low B1+ sensitivity in a short, patient-friendly breath-hold
acquisition.METHODS
To
achieve the simultaneous goals of fat fraction quantification, water T1
quantification and B1+ insensitivity, an investigational
prototype saturation-recovery multi-echo gradient-echo Dixon approach with
ramped excitation flip angles is proposed.
Non-saturation and saturation recovery images were acquired sequentially
(Fig. 1). T1 was evaluated from Dixon
water-separated images using signal ratios (lookup table approach based on
Bloch equation simulations, including actual slice profile). Bloch equation simulations were also used to evaluate
the dependence of calculated T1 on B1+
for comparison of conventional constant flip
angle and ramped-flip-angle approaches. Fat
fraction was evaluated from non-saturation images (conventional Dixon approach (2)). Pulse
Sequence Parameters: 3 interleaved slices, 8 mm
thickness, 192x144 acquisition matrix, 400 x 333 mm2 FOV, GRAPPA =
2, 1600 Hz/pixel receiver bandwidth, TR = 36ms, 6 equally spaced TEs 1.23 – 10.23
ms, flip angle = 20°, one non-sat image and four
saturation recovery images (TS=1302ms), 13 second acquisition time. Ramped flip
angles followed a sin(q) pattern
with increasing RF pulse index, with maximum value (q=p/2)
at the last RF pulse.(3) A custom
6-pulse saturation train was used to ensure >99% saturation efficiency at 3T.
(3) Phantom
and in-vivo experiments were completed on a 3T scanner (MAGNETOM Prisma,
Siemens Healthcare Erlangen, Germany). NiCl2-agarose
phantoms with a wide range of T1 and T2 values were used for
validation of the T1 mapping approach (comparison to saturation-recovery
spin-echo experiments). Fat fraction and
water T1 images from the proposed PDFF-T1 approach were acquired from a healthy
control and subject with mild fatty liver disease.RESULTS
Figure 2
compares the simulated signal yield of the constant and ramped-flip-angle
approach as a function of T1, for the target flip angle of 20°. The
constant-flip-angle readout yields ~50% signal variation over the range of T1
values (300-2000ms) while the ramped-flip-angle approach yields only ~7% signal
variation. For these same two excitation
schemes, lookup tables to convert signal ratios into T1 values are plotted for
a wide range of B1+
scaling factors (0.5-1.5) representing actual
flip angle of 10° to 30° (Figure 3).
The ramped-flip-angle approach has ~8x reduced sensitivity to errors in
B1+ as compared to the constant-flip-angle approach. Correlation and Bland-Altman plots illustrate
the relationship between the phantom PDFF-T1 T1
values and spin-echo experiments, showing excellent agreement (Figure 4). Fat fraction and T1 maps from the liver of a
healthy control and subject with mild fatty liver disease are compared (Figure
5).DISCUSSION
The
proposed PDFF-T1 sequence yielded fat fraction and water T1 maps in a single
patient-friendly breath-hold, with significant reduction of unwanted T1-weighting
and flip-angle (B1+)
dependence with the use of a ramped-flip-angle approach. Minimization of T1-weighing has the advantage
of: i) increase fat fraction accuracy and ii) reduced B1+ sensitivity
in T1 calculation (Fig. 3). The use of the robust Dixon fat-water separation
ensures minimal fat contamination of water-derived T1 values. The PDFF-T1 approach avoids the complex T1
signal response for mixed fat-water systems observed for bSSFP T1 mapping
methods. (4,5) The multi-echo acquisition also enables T2*
assessment. CONCLUSION
Fast
simultaneous quantification of PDFF (with minimal T1 effects) and water T1 is
possible with the proposed PDFF-T1 approach.
Ramped-flip-angles minimize sensitivity to B1+.Acknowledgements
No acknowledgement found.References
1. Taouli B, Ehman RL, Reeder SB.
Advanced MRI methods for assessment of chronic liver disease. AJR Am J
Roentgenol 2009;193(1):14-27.
2. Yu H, Shimakawa
A, McKenzie CA, Brodsky E, Brittain JH, Reeder SB. Multiecho water-fat
separation and simultaneous R2* estimation with multifrequency fat spectrum modeling.
Magn Reson Med 2008;60(5):1122-1134.
3. Chow K, Kellman
P, Spottiswoode BS, Nielles-Vallespin S, Arai AE, Salerno M, Thompson RB.
Saturation pulse design for quantitative myocardial T1 mapping. J Cardiovasc
Magn Reson 2015;17:84.
4. Larmour S, Chow
K, Kellman P, Thompson RB. Characterization of T1 bias in skeletal muscle from
fat in MOLLI and SASHA pulse sequences: Quantitative fat-fraction imaging with
T1 mapping. Magn Reson Med 2017;77(1):237-249.
5. Mozes FE,
Tunnicliffe EM, Pavlides M, Robson MD. Influence of fat on liver T1
measurements using modified Look-Locker inversion recovery (MOLLI) methods at
3T. J Magn Reson Imaging 2016;44(1):105-111.