Ed Boskamp1, Victor Taracila1, Scott Lindsay2, Robert Stormont2, Reni Biswas3,4, Sheronda Statum3,4, Shane Aldas5, Cesar Barraza5, Fraser Robb1, Christine B Chung3,4, and Won Bae3,4
1G. E. Healthcare Technologies, Aurora, OH, United States, 2G. E. Healthcare Technologies, Waukesha, WI, United States, 3Radiology, University of California, San Diego, La Jolla, CA, United States, 4Radiology, VA San Diego Healthcare System, San Diego, CA, United States, 5Electrical Engineering, University of California, San Diego, La Jolla, CA, United States
Synopsis
MR
neurography of the Brachial Plexus (BP) is technically challenging, due in part
to complex topographical anatomy that is sub-optimally accommodated with
existing rigid receiver coil arrays. In this work we propose a wearable garment
with integrated flexible loop receiver coils that conforms to the contour of
the neck and shoulder, as well as variations in subject size. The loops can
flex in multiple dimensions without seriously affecting noise correlation.
Introduction
The
BP is comprised of the ventral rami of C5 through C8 and the first thoracic
nerve. BP (Figure 1) runs postero-laterally from the cervical spine with roots,
trunks, divisions, cords and terminal nerves that traverse the soft tissues
above and below the clavicle, and extend into the arm. This extended anatomic
region requires a large field of coverage for complete evaluation. Historically
MR imaging (commonly referred to as MR neurography or MRN) of BP has been done
with rigid head/neck/spine (HNS) receiver arrays but such array elements are
difficult to place close to the region of interest in a large percentage of
patients with varying shape and size and consequently do not provide a satisfactory
image quality (Figure 1). A new variable size and shape vest has been developed
with air coil1,2 elements sewn directly onto the vest. The elements
are light weight and flexible in all directions. Our purpose is to characterize
this new highly flexible RF coil design to the needs for BP imaging.Methods
Coil
Design: The BP was modeled based on several months of imaging. This model was
then imported into a finite element full wave solver (HFSS, Ansys Inc) and
merged with the average visible man model. In total twenty-four 11 cm diameter air
coils loops were positioned manually (Figure 2) by a musculoskeletal
radiologist familiar with BP anatomy. The air coil1,2 loops are made
from highly flexible 1.0 mm wire optimized for zero reactance at 127.7 MHz.
Loops were connected to a miniaturized (22 by 22 by 4.5mm) network containing
baluns, transmit blocking, and a low impedance preamplifier. An output balun
adds another 5 mm of build up height. Loops were sewn onto a vest made from TenCate Omniweave 600. This vest has a front plastic zipper and Velcro under the
arms and laterally on the left and right side, allowing for close fit on
patients of different size and shape. The posterior part of the coil has
foam padding to protect the circuits from being damaged by the weight of the
patient. Total weight of the
vest with all loops, cables, padding and cable baluns is less than 0.5 kg.
Decoupling during reception was achieved using the Roemer’s method3,
as well as electric field
containment and reduction methods. Coil-to-coil overlap can vary between
underlapped and 35% overlapped with minor impact on noise correlation. The
empty and loaded Q of the loops were measured on a network analyzer.
Coil Characterization in Volunteers: Three volunteers with varying weights
(98 lb female, 145 lb male, 170 lb male) and shapes were imaged. On each
volunteer, coronal spin echo imaging (TR=400 ms, TE=10 ms, ETL 4, FOV=30 cm,
matrix=320x256, slice=4 mm) with and without RF was performed to calculate noise
correlation3,4 (Figure 3) were 0 mean no
correlation and 1 means perfect correlation. In addition, clinical images (coronal T2
IDEAL water: TR=3100 ms, TE=82 ms, ETL=14, FOV=34 cm, matrix=320x224, slice=4
mm, 23 slices; maximum intensity projection through 8 mm thickness) were
obtained from a volunteer using both conventional Head-Neck-Spine/HNS (Figure 4A) and our BP coil (Figure 4B).
Results
The Q of the air coil elements was measured
to be 125 empty and 40 loaded. The noise correlation matrices of the different
size volunteers (Figure 3A, small; Figure 3B, large) were similar between
volunteers (mean standard deviation of 0.02; Figure 3C), and did not show any
appreciable coupling between elements evident by low correlation values (mean off-
diagonal value of 0.007). Comparison of images from HNS (Figure 4A) vs. BP
(Figure 4B) coils showed a marked qualitative improvement in signal-to-noise
ratio, contrast and resolution of the nerve bundles with the use of BP coil. For
RL acceleration g factors are <2 up to R=5 for 98lbs and up to R=4 for 140
and 170 lbs volunteers.Discussion
Presently the ongoing work is to optimize the
placement, number and size of air coil elements to obtain optimal coverage and
accelerated and non-accelerated SNR, using the finite element model of the BP inside
body model. Work also remains to
quantitatively determine performance improvements vs. conventional coils, in
terms of SNR and CNR in vivo, as well as diagnostics in injury patients.Conclusion
A dedicated coil for BP application has been
created and its noise characteristics has been described in volunteers of
varying sizes. Image quality, when compared to that from a conventional HNS
coil, showed marked improvements.Acknowledgements
Valuable support was received from Anthony
Stafford at GE Coils Aurora and Liang Xuan at GE Healthcare WaukeshaReferences
1. P. Rossman et al., Characterization of a
new ultra flexible low profile RF receive coil technology, Proceedings ISMRM
2017, 763.
2. S. Vasanawala et al., Development and
Clinical implementation of very light weight and highly flexible Air technology
arrays, Proceedings ISMRM 2017, 755.
3. P. Roemer et al., The NMR phased array,
MRM 1990 Nov;16(2):192-225
4. V. Taracila et al., Proceedings ISMRM 2015, 4893.