4D-Flow MRI is a powerful technique for simultaneously imaging vascular anatomy and hemodynamics. However, its clinical utility is limited by long (10-20 minute) scan times. This work aims to shorten scan times by using fast RF pulses and convex optimized gradient waveforms. The waveforms are optimized with arbitrary shapes, and are designed to go as fast possible without causing peripheral nerve stimulation by including an additional PNS constraint. The optimized sequence is implemented and tested in flow phantoms and a volunteer. The data acquired with the optimized waveforms is up to 33% faster, with no significant difference in measured data compared to a reference sequence.
RF Optimization – The slab-selective excitation profiles were compared for a standard windowed-sinc RF pulse and a minimum-phase SLR RF pulse [2]. VERSE optimization was also applied to these two pulses to further shorten the RF pulse duration with an equivalent excitation profile[3].
Gradient Optimization – The bipolar gradients were shortened with convex optimization[4]. This method finds the shortest possible gradient waveform of arbitrary shape that satisfies a set of given constraints: 1) hardware limits on gradient amplitude (80mT/m) and slew rate (200T/m/s); 2) target gradient zeroth moments (M0) for phase encoding, slab-selection rewinding, and readout pre-winding; 3) target gradient first moments (M1) for velocity encoding; and 4) PNS using the SAFE model for estimating the peripheral nerve response to applied gradients[5]. Additionally, the spoiler gradients were optimized using a similar procedure, but without any constraint on M1.
RF Pulse – RF and slab-selection gradients were tested independently of other optimizations on a static phantom to examine excitation profiles, and in a flow phantom with fluid pumped in and out of the slab. Flow values were compared to the standard vendor supplied windowed sinc pulse with Bland-Altman analysis.
RF+Gradient Waveform – The fully optimized sequence (using VERSE min-phase RF) was compared to a vendor supplied reference sequence that used conventional optimizations as in [1]. Optimized TE and TR were compared to the reference sequence over a range of Vencs and resolutions. The optimized sequence was implemented on a 3T scanner (Siemens,Prisma) with the following parameters: Venc=80cm/s, 1.0x1.0x2.0mm3 resolution, 220x164x64mm FOV, 8°flip, BW=550Hz/px. The optimized sequence was compared to the reference sequence in a flow phantom, in a static phantom for eddy current comparisons, and in vivo in the head of a healthy volunteer. Flow rates were compared with Bland-Altman analysis across multiple planes and ROIs.
RF Pulse – Figure 1 shows the results of the slab-selection comparison where excitation profiles of all the tested methods perform similarly. Flow values are not significantly changed with the use of any of the tested waveforms. The VERSE min-phase SLR pulse is the fastest (0.23ms vs reference 1.0ms).
RF+Gradient Waveform – Figure 2 shows substantial time savings for the optimized RF and gradient waveforms compared to the reference sequence for a given protocol (TR/TE reduction=5.62ms to 3.82ms, 3.11ms to 2.33ms respectively). Figure 3 shows timings from the optimized waveforms compared to the reference over a range of Vencs and resolutions. TE/TR were reduced by 13-38% and 24-33% respectively. Bland Altman analysis of flow-rates in the phantom experiment (Figure 4) shows good agreement. There was no significant difference in VNR between phantom experiments (19.5±0.6 vs 19.3±0.6). Figure 4b shows some increase in the background eddy current phase with the optimized method. Figure 5 shows the results of in vivo validation, where good qualitative or quantitative differences are seen. Due to the shorter TR, an additional segment could be used to reduce scan time by 25% (15:41 to 11:39), while also slightly improving temporal resolution (67ms to 61ms).
[1] Bernstein M a, Shimakawa a, Pelc NJ. Minimizing TE in moment-nulled or flow-encoded two- and three-dimensional gradient-echo imaging. J Magn Reson Imaging. 1992;2(5):583–8.
[2] Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging]. IEEE Trans Med Imaging. 1991;10(1):53–65.
[3] Hargreaves B a., Cunningham CH, Nishimura DG, Conolly SM. Variable-rate selective excitation for rapid MRI sequences. Magn Reson Med. 2004;52(3):590–7.
[4] Middione MJ, Wu HH, Ennis DB. Convex gradient optimization for increased spatiotemporal resolution and improved accuracy in phase contrast MRI. Magn Reson Med. 2014;72(6):1552–64.
[5] Franz X. Hebrank, Matthias Gebhardt. SAFE-Model - A New Method for Predicting Peripheral Nerve Stimulations in MRI. ISMRM. 2000.